In vitro and in vivo degradation of lactic acid-based interference screws used in cruciate ligament reconstruction

被引:80
作者
Schwach, G
Vert, M
机构
[1] PHUSIS Mat Bioresorbables, F-38330 Saint Ismier, France
[2] Univ Montpellier I, Fac Pharm, CRBA, F-34060 Montpellier, France
关键词
polylactide; poly(lactic acid); bioresorbable interference screw; hydrolytic degradation; surgical implants; biodegradable polymers; anterior cruciate ligament; biocompatibility;
D O I
10.1016/S0141-8130(99)00043-4
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Nowadays, many degradable polymers are being used under the form of interference screws to fix the bent-tendon-bone autograft in anterior cruciate ligament reconstruction. However, little is known about the post-implantation fate of these screws, especially about the formation of crystalline residues which seems to be a critical factor for the success of surgery with temporary implants based on lactic and glycolic acid derived polymers (PLAGA). In an attempt to bring in some new insights, various high molecular weight stereoregular poly(lactide)s (PLAX with X = percentage of L-lactyl units) obtained by ring-opening polymerization of lactides in the presence of zinc-metal (PLA98-Zn), zinc lactate (PLA98-Znlac) or stannous octoate (PLA100-Sn), were processed by injection-molding to make interference screws to be compared. In vivo data were collected from screws implanted in sheep knees with follow ups ranging from 6 months to 5 years. Histology confirmed the heterogeneous degradation mechanism introduced nearly 10 years ago from in vitro investigations of homemade implants having simpler geometry. The effects of the initiator system (zinc- or tin derivatives) used to polymerize the lactide monomer on the properties of injection molded interference screws was also investigated in vitro in a phosphate buffer solution at 37 degrees C. Major differences in terms of hydrophilicity, hydrolysis rate and loss of mechanical properties were observed between PLA-Zinc and PLA-Tin. Discussion of the behavior of interference screws of different compositions was made on the basis of the present understanding of PLAGA morphology and degradation characteristics. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:283 / 291
页数:9
相关论文
共 36 条
  • [1] PRELIMINARY-RESULTS OF AN ABSORBABLE INTERFERENCE SCREW
    BARBER, FA
    ELROD, BF
    MCGUIRE, DA
    PAULOS, LE
    [J]. ARTHROSCOPY, 1995, 11 (05): : 537 - 548
  • [2] LATE DEGRADATION TISSUE-RESPONSE TO POLY(L-LACTIDE) BONE PLATES AND SCREWS
    BERGSMA, JE
    DEBRUIJN, WC
    ROZEMA, FR
    BOS, RRM
    BOERING, G
    [J]. BIOMATERIALS, 1995, 16 (01) : 25 - 31
  • [3] BERO M, 1990, MAKROMOL CHEM, V191, P2287
  • [4] Osteoarthritis of the ankle after foreign-body reaction to absorbable pins and screws -: A three- to nine-year follow-up study
    Böstman, OM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02): : 333 - 338
  • [5] CONFIGURATIONAL STRUCTURES OF LACTIC-ACID STEREOCOPOLYMERS AS DETERMINED BY C-13-LABELED (H-1)-NMR
    CHABOT, F
    VERT, M
    CHAPELLE, S
    GRANGER, P
    [J]. POLYMER, 1983, 24 (01) : 53 - 59
  • [6] CHAMBAT P, 1995, P 26 ANN M E ORTH AS, P128
  • [7] FAYARD JP, KNEE SURG ARTHROSCOP, P216
  • [8] The effect of thermal treatment on sterility, molecular and mechanical properties of various polylactides .1. Poly(L-lactide)
    Gogolewski, S
    MainilVarlet, P
    [J]. BIOMATERIALS, 1996, 17 (05) : 523 - 528
  • [9] HYDROLYTIC DEGRADATION OF DEVICES BASED ON POLY(DL-LACTIC ACID) SIZE-DEPENDENCE
    GRIZZI, I
    GARREAU, H
    LI, S
    VERT, M
    [J]. BIOMATERIALS, 1995, 16 (04) : 305 - 311
  • [10] POLYLACTONES .23. POLYMERIZATION OF RACEMIC AND MESO D,L-LACTIDE WITH VARIOUS ORGANOTIN CATALYSTS STEREOCHEMICAL ASPECTS
    KRICHELDORF, HR
    BOETTCHER, C
    TONNES, KU
    [J]. POLYMER, 1992, 33 (13) : 2817 - 2824