Reprocessing filtering facepiece respirators in primary care using medical autoclave: prospective, bench-to-bedside, single-centre study

被引:15
作者
Harskamp, Ralf E. [1 ]
van Straten, Bart [2 ,3 ]
Bouman, Jonathan [1 ]
van Maltha-van Santvoort, Bernadette [4 ]
van den Dobbelsteen, John J. [2 ]
van der Sijp, Joost R. M. [5 ,6 ]
Horeman, Tim [2 ]
机构
[1] Amsterdam UMC, Locatie AMC, Dept Gen Practice, Amsterdam, Netherlands
[2] Delft Univ Technol, Fac Engn, Dept BioMed Engn, Delft, Netherlands
[3] Van Straten Med, De Meern, Netherlands
[4] Holendrecht Med Ctr, Amsterdam, Netherlands
[5] Med Ctr Haaglanden, The Hague, Zuid Holland, Netherlands
[6] GreenCycl, De Meem, Netherlands
关键词
primary care; respiratory infections; public health; INFLUENZA-VIRUS; AEROSOLS; COVID-19;
D O I
10.1136/bmjopen-2020-039454
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective There are widespread shortages of personal protective equipment as a result of the COVID-19 pandemic. Reprocessing filtering facepiece particle (FFP)-type respirators may provide an alternative solution in keeping healthcare professionals safe. Design Prospective, bench-to-bedside. Setting A primary care-based study using FFP-2 respirators without exhalation valve (3M Aura 1862+ (20 samples), Maco Pharma ZZM002 (14 samples)), FFP-2 respirators with valve (3M Aura 9322+ (six samples) and San Huei 2920V (16 samples)) and valved FFP type 3 respirators (Safe Worker 1016 (10 samples)). Interventions All masks were reprocessed using a medical autoclave (17 min at 121 degrees C with 34 min total cycle time) and subsequently tested up to three times whether these respirators retained their integrity (seal check and pressure drop) and ability to filter small particles (0.3-5.0 mu m) in the laboratory using a particle penetration test. Results We tested 33 respirators and 66 samples for filter capacity. All FFP-2 respirators retained their shape, whereas half of the decontaminated FFP-3 respirators showed deformities and failed the seal check. The filtering capacity of the 3M Aura 1862 was best retained after one, two and three decontamination cycles (0.3 mu m: 99.3%+/- 0.3% (new) vs 97.0 +/- 1.3, 94.2 +/- 1.3% or 94.4 +/- 1.6; p<0.001). Of the other FFP-2 respirators, the San Huei 2920 V had 95.5%+/- 0.7% at baseline vs 92.3%+/- 1.7% vs 90.0 +/- 0.7 after one-time and two-time decontaminations, respectively (p<0.001). The tested FFP-3 respirator (Safe Worker 1016) had a filter capacity of 96.5%+/- 0.7% at baseline and 60.3%+/- 5.7% after one-time decontamination (p<0.001). Breathing and pressure resistance tests indicated no relevant pressure changes between respirators that were used once, twice or thrice. Conclusion This small single-centre study shows that selected FFP-2 respirators may be reprocessed for use in primary care, as the tested masks retain their shape, ability to retain particles and breathing comfort after decontamination using a medical autoclave.
引用
收藏
页数:6
相关论文
共 22 条
[1]  
CDC, 2020, INF HEALTHC PROF COR
[2]   Sterilization of disposable face masks by means of standardized dry and steam sterilization processes; an alternative in the fight against mask shortages due to COVID-19 [J].
de Man, P. ;
van Straten, B. ;
van den Dobbelsteen, J. ;
van der Eijk, A. ;
Horeman, T. ;
Koeleman, H. .
JOURNAL OF HOSPITAL INFECTION, 2020, 105 (02) :356-357
[3]  
Fisher RJ, ASSESSMENT N95 RESPI, DOI 10.1101/2020.04.11.20062018v2.full.pdf
[4]   A pandemic influenza preparedness study: Use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets [J].
Heimbuch, Brian K. ;
Wallace, William H. ;
Kinney, Kimberly ;
Lumley, April E. ;
Wu, Chang-Yu ;
Woo, Myung-Heui ;
Wander, Joseph D. .
AMERICAN JOURNAL OF INFECTION CONTROL, 2011, 39 (01) :E1-E9
[5]  
Heneghan C., 2020, SARS COV 2 VIRAL LOA
[6]   Personal protective equipment and improving compliance among healthcare workers in high-risk settings [J].
Honda, Hitoshi ;
Iwata, Kentaro .
CURRENT OPINION IN INFECTIOUS DISEASES, 2016, 29 (04) :400-406
[7]  
Horton R, 2020, LANCET, V395, P1022, DOI 10.1016/S0140-6736(20)30727-3
[8]   Transmission of COVID-19 virus by droplets and aerosols: A critical review on the unresolved dichotomy [J].
Jayaweera, Mahesh ;
Perera, Hasini ;
Gunawardana, Buddhika ;
Manatunge, Jagath .
ENVIRONMENTAL RESEARCH, 2020, 188
[9]   Is the coronavirus airborne? Experts can't agree [J].
Lewis, Dyani .
NATURE, 2020, 580 (7802) :175-175
[10]   Filter quality of electret masks in filtering 14.6-594 nm aerosol particles: Effects of five decontamination methods [J].
Lin, Tzu-Hsien ;
Chen, Chih-Chieh ;
Huang, Sheng-Hsiu ;
Kuo, Chung-Wen ;
Lai, Chane-Yu ;
Lin, Wen-Yinn .
PLOS ONE, 2017, 12 (10)