Size Distribution of Magnetic Marker Estimated from AC Susceptibility in Solution for Biosensor Application

被引:27
作者
Enpuku, Keiji [1 ]
Tanaka, Tsuyoshi [1 ]
Tamai, Yuya [1 ]
Dang, Feng [2 ]
Enomoto, Naoya [2 ]
Hojo, Junichi [2 ]
Kanzaki, Hisao [3 ]
Usuki, Naoki [3 ]
机构
[1] Kyushu Univ, Res Inst Supercond Sci & Syst, Fukuoka 8190395, Japan
[2] Kyushu Univ, Dept Appl Chem, Fukuoka 8190395, Japan
[3] Hitachi Maxell Ltd, Div Res & Dev, Kyoto 6188525, Japan
基金
日本学术振兴会;
关键词
magnetic marker; magnetic nanoparticle; Brownian rotation; ac susceptibility; size distribution; singular value decomposition; biosensor;
D O I
10.1143/JJAP.47.7859
中图分类号
O59 [应用物理学];
学科分类号
摘要
The ac susceptibility of magnetic markers in solution was Studied for biosensor application, where the marker consisted of magnetic nanoparticles and a coating material. From the frequency dependence of the susceptibility caused by the Brownian rotation of the marker, we estimated the distribution of marker size, which is an important parameter for biosensor application. For this purpose, we analyzed the experimental data by the singular value decomposition (SVD) method. Using this method, we can directly estimate the size distribution without assuming any distribution function. The estimated distributions were also compared with those obtained from optical dynamic light scattering (DLS) measurements. It was shown that the size distribution estimated by magnetic measurement (SVD) slightly shifts to a size lower than that estimated by the optical measurement (DLS). It was also shown that the frequency dependence of the susceptibility can be better explained by the size distribution estimated by the SVD method than by the distribution estimated with DLS. The difference between the magnetic and optical measurement results was discussed in terms of aggregation of the markers. [DOI: 10.1143/JJAP.47.7859]
引用
收藏
页码:7859 / 7865
页数:7
相关论文
共 21 条
[11]   Wideband measurement and analysis techniques for the determination of the frequency-dependent, complex susceptibility of magnetic fluids [J].
Fannin, PC .
ADVANCES IN CHEMICAL PHYSICS, VOL 104, 1998, 104 :181-292
[12]   Brownian relaxation of magnetic colloids [J].
Fischer, B ;
Huke, B ;
Lücke, M ;
Hempelmann, R .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2005, 289 :74-77
[13]   Development of by time-dependent a liquid phase immunoassay measurements of the transient magneto-optical birefringence using functionalized magnetic nanoparticles [J].
Glöckl, G ;
Brinkmeier, V ;
Aurich, K ;
Romanus, E ;
Weber, P ;
Weitschies, W .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2005, 289 :480-483
[14]   Detection of bacteria in suspension by using a superconducting quantum interference device [J].
Grossman, HL ;
Myers, WR ;
Vreeland, VJ ;
Bruehl, R ;
Alper, MD ;
Bertozzi, CR ;
Clarke, J .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2004, 101 (01) :129-134
[15]   Magnetic susceptibility reduction method for magnetically labeled immunoassay [J].
Hong, Chin-Yih ;
Wu, C. C. ;
Chiu, Y. C. ;
Yang, S. Y. ;
Horng, H. E. ;
Yang, H. C. .
APPLIED PHYSICS LETTERS, 2006, 88 (21)
[16]   Magnetorelaxometry -: a new binding specific detection method based on magnetic nanoparticles [J].
Lange, J ;
Kötitz, R ;
Haller, A ;
Trahms, L ;
Semmler, W ;
Weitschies, W .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2002, 252 (1-3) :381-383
[17]   A combination of magnetic permeability detection with nanometer-scaled superparamagnetic tracer and its application for one-step detection of human urinary albumin in undiluted urine [J].
Lu, Min ;
Ibraimi, Filiz ;
Kriz, Dario ;
Kriz, Kirstin .
BIOSENSORS & BIOELECTRONICS, 2006, 21 (12) :2248-2254
[18]   Magnetorelaxometry of magnetic nanoparticles in magnetically unshielded environment utilizing a differential fluxgate arrangement -: art. no. 106102 [J].
Ludwig, F ;
Mäuselein, S ;
Heim, E ;
Schilling, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (10) :1-3
[19]   Magnetic biosensor for the detection of Yersinia pesos [J].
Meyer, Martin H. F. ;
Stehr, Matthias ;
Bhuju, Sabin ;
Krause, Hans-Joachim ;
Hartmann, Markus ;
Miethe, Peter ;
Singh, Mahavir ;
Keusgen, Michael .
JOURNAL OF MICROBIOLOGICAL METHODS, 2007, 68 (02) :218-224
[20]   Influence of particle size distribution on the initial susceptibility of magnetic fluids in the Brown relaxation range [J].
Payet, B ;
Vincent, D ;
Delaunay, L ;
Noyel, G .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1998, 186 (1-2) :168-174