Dosimetric impact of multileaf collimator leaf width according to sophisticated grade of technique in the IMRT and VMAT planning for pituitary adenoma lesion

被引:10
作者
Chae, Soo-Min [1 ]
Lee, Ki Woong [1 ]
Son, Seok Hyun [2 ]
机构
[1] Cheju Halla Gen Hosp, Dept Radiat Oncol, Jeju, South Korea
[2] Catholic Univ Korea, Coll Med, Incheon St Marys Hosp, Dept Radiat Oncol, Seoul, South Korea
关键词
multi-leaf collimator; radiosurgery; intensity-modulated radiotherapy; volumetric modulated arc therapy; INTENSITY-MODULATED RADIOTHERAPY; STEREOTACTIC BODY RADIOTHERAPY; PRIMARY BRAIN-TUMORS; RADIATION-THERAPY; GAMMA-KNIFE; ARC THERAPY; RADIOSURGICAL TREATMENT; TREATMENT PLANS; PROTOCOL; 90-05; CONFORMITY;
D O I
10.18632/oncotarget.12974
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
We analyzed the difference in the dosimetric effect between 5-mm and 2.5-mm multileaf collimator (MLC) leaf width according to the sophisticated grades of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). Nineteen patients with pituitary adenomas were selected for this study. The treatment plans were performed according to the size of the MLC (5-mm and 2.5-mm MLC), the type of technique (IMRT and VMAT), and the sophisticated grades of each technique (5-field, 9-field, 13-field, 17-field technique in IMRT and 1-arc and 2-arc techniques in VMAT). The downsizing effects of MLC leaf width were analyzed using target volume coverage (TVC), conformity index (CI), dose gradient index (GI), and normal tissue difference 70% isodose line and 50% isodose line. Upon replacing the 5-mm MLC with the 2.5-mm MLC, TVC and CI improved by 1.30% and 1.36%, respectively, in total plans. The TVC and CI improved by 1.68% and 1.67% in IMRT, respectively, and by 0.54% and 0.72% in VMAT, respectively. TVC improved by 2.53%, 1.82%, 1.34%, and 0.94%, and CI also improved by 2.70%, 1.81%, 1.24%, and 0.94%, in 5-field, 9-field, 13-field, and 17-field IMRT, respectively. TVC improved by 0.66% and 0.43%, and CI also improved by 0.93%, and 0.52% in 1-arc and 2-arc VMAT, respectively. Regarding the target coverage, there were dosimetric benefits of a smaller MLC leaf width. However, the downsizing effect of the MLC leaf width decreased with the use of a more precise RT technique and a more sophisticated grade of the same technique.
引用
收藏
页码:78119 / 78126
页数:8
相关论文
共 39 条
[11]   Stereotactic radiosurgery for pituitary adenomas: a review of the literature [J].
Laws, ER ;
Sheehan, JP ;
Sheehan, JM ;
Jagnathan, J ;
Jane, JA ;
Oskouian, R .
JOURNAL OF NEURO-ONCOLOGY, 2004, 69 (1-3) :257-272
[12]   DOSIMETRIC COMPARISON OF HELICAL TOMOTHERAPY AND DYNAMIC CONFORMAL ARC THERAPY IN STEREOTACTIC RADIOSURGERY FOR VESTIBULAR SCHWANNOMAS [J].
Lee, Tsair-Fwu ;
Chao, Pei-Ju ;
Wang, Chang-Yu ;
Lan, Jen-Hong ;
Huang, Yu-Je ;
Hsu, Hsuan-Chih ;
Sung, Chieh-Cheng ;
Su, Te-Jen ;
Lian, Shi-Long ;
Fang, Fu-Min .
MEDICAL DOSIMETRY, 2011, 36 (01) :62-70
[13]   Comparison of deliverable IMRT and VMAT for spine metastases using a simultaneous integrated boost [J].
Lee, Y. K. ;
Bedford, J. L. ;
McNair, H. A. ;
Hawkins, M. A. .
BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1022)
[14]  
Lindquist, 1995, Semin Radiat Oncol, V5, P197, DOI 10.1016/S1053-4296(05)80017-7
[15]   Quantifying the degree of conformity in radiosurgery treatment planning [J].
Lomax, NJ ;
Scheib, SG .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05) :1409-1419
[16]   Comparison of a micro-multileaf collimator with a 5-mm-leaf-width collimator for intracranial stereotactic radiotherapy [J].
Monk, JE ;
Perks, JR ;
Doughty, D ;
Plowman, PN .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (05) :1443-1449
[17]   Dose conformity of Gamma Knife radiosurgery and risk factors for complications [J].
Nakamura, JL ;
Verhey, LJ ;
Smith, V ;
Petti, PL ;
Lamborn, KR ;
Larson, DA ;
Wara, WM ;
McDermott, MW ;
Sneed, PK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (05) :1313-1319
[18]   A simple scoring ratio to index the conformity of radiosurgical treatment plans - Technical note [J].
Paddick, I .
JOURNAL OF NEUROSURGERY, 2000, 93 :219-222
[19]   A simple dose gradient measurement tool to complement the conformity index [J].
Paddick, Ian ;
Lippitz, Bodo .
JOURNAL OF NEUROSURGERY, 2006, 105 :194-201
[20]   The use of RapidArc volumetric-modulated arc therapy to deliver stereotactic radiosurgery and stereotactic body radiotherapy to intracranial and extracranial targets [J].
Roa, Dante E. ;
Schiffner, Daniel C. ;
Zhang, Juying ;
Dietrich, Salam N. ;
Kuo, Jeffrey V. ;
Wong, Jason ;
Ramsinghani, Nilam S. ;
Al-Ghazi, Muthana S. A. L. .
MEDICAL DOSIMETRY, 2012, 37 (03) :257-264