Development of a fixation device for robot assisted fracture reduction of femoral shaft fractures: A biomechanical study

被引:7
作者
Weber-Spickschen, T. S. [1 ]
Oszwald, M. [3 ]
Westphal, R. [2 ]
Krettek, C. [3 ]
Wahl, F. [2 ]
Gosling, T. [3 ]
机构
[1] Robert Koch Hosp Gehrden, Dept Orthopaed & Traumatol, D-30989 Gehrden, Germany
[2] Tech Univ Carolo Wilhelmina Braunschweig, Inst Robot & Proc Control, D-38106 Braunschweig, Germany
[3] Hannover Med Sch, Trauma Dept, Hannover, Germany
关键词
Robot; fracture reduction; femur shaft; interface; INTRAMEDULLARY; COMPUTER; FEMUR; REGISTRATION; DEFORMITY; NAIL;
D O I
10.3233/THC-2010-0583
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Robot assisted fracture reduction of femoral shaft fractures provides precise alignment while reducing the amount of intraoperative imaging. The connection between the robot and the fracture fragment should allow conventional intramedullary nailing, be minimally invasive and provide interim fracture stability. In our study we tested three different reduction tools: a conventional External Fixator, a Reposition-Plate and a Three-Point-Device with two variations (a 40 degrees and a 90 degrees version). We measured relative movements between the tools and the bone fragments in all translation and rotation planes. The Three-Point-Device 90 degrees showed the smallest average relative displacement and was the only device able to withstand the maximum applied load of 70 Nm without failure of any bone fragment. The Three-Point-Device 90 degrees complies with all the stipulated requirements and is a suitable interface for robot assisted fracture reduction of femoral shaft fractures.
引用
收藏
页码:207 / 216
页数:10
相关论文
共 50 条
  • [31] Risk factors for implant failure after fixation of proximal femoral fractures with fracture of the lateral femoral wall
    Gao, Zhechen
    Lv, Yang
    Zhou, Fang
    Ji, Hongquan
    Tian, Yun
    Zhang, Zhishan
    Guo, Yan
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2018, 49 (02): : 315 - 322
  • [32] Overgrowth of the Femur After Internal Fixation in Children With Femoral Shaft Fracture-A Multicenter Study
    Park, Kyeong-Hyeon
    Park, Byoung Kyu
    Oh, Chang-Wug
    Kim, Dae-Wook
    Park, Hoon
    Park, Kun-Bo
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2020, 34 (03) : E90 - E95
  • [33] A biomechanical comparison of the antegrade inserted universal femoral nail with the retrograde inserted universal tibial nail for use in femoral shaft fractures
    Frankle, M
    Cordey, J
    Sanders, RW
    Koval, K
    Perren, SM
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1999, 30 : S40 - S43
  • [34] Automated Robot Assisted Fracture Reduction
    Westphal, Ralf
    Winkelbach, Simon
    Goesling, Thomas
    Oszwald, Markus
    Huefner, Tobias
    Krettek, Christian
    Wahl, Friedrich M.
    ADVANCES IN ROBOTICS RESEARCH, 2009, : 251 - +
  • [35] The Biomechanical Analysis for DHS and Wire fixation Osteoporotic and Unstable Femoral Fracture
    Wu, H. F.
    Lai, K. A.
    Huang, M. T.
    Chen, H. S.
    Chung, K. C.
    Yang, F. S.
    2011 IEEE 37TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE (NEBEC), 2011,
  • [36] In Vitro Biomechanical Study of Femoral Neck Fracture Fixation with Two or Three Cannulated Screws in Dogs
    Caldeira, Fatima Maria C.
    Rahal, Sheila C.
    Cassanego, Guilherme R.
    Ribeiro, Celso R.
    Serrato, Sergio A. A.
    Mele, Carlos E.
    Masseran, Paulo
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2025, 38 (01) : 18 - 24
  • [37] Automatic Bone Segmentation and Ultrasound-CT Registration for Robotic Assisted Femoral Shaft Fracture Reduction
    Wang, Tianmiao
    Wang, Lifeng
    Tang, Peifu
    Hu, Lei
    Zhang, Lihai
    Liu, Hongpeng
    Hao, Ming
    Han, Zhonghao
    Zhao, Yanpeng
    Guo, Na
    Li, Changsheng
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2018, 8 (01) : 151 - 156
  • [38] Cement augmentation of an angular stable plate osteosynthesis for supracondylar femoral fractures-biomechanical investigation of a new fixation device
    Baeumlein, Martin
    Klasan, Antonio
    Kloetzer, Christine
    Bockmann, Benjamin
    Eschbach, Daphne
    Knobe, Matthias
    Buecking, Benjamin
    Ruchholtz, Steffen
    Bliemel, Christopher
    BMC MUSCULOSKELETAL DISORDERS, 2020, 21 (01)
  • [39] Preoperative virtual reduction reduces femoral malrotation in the treatment of bilateral femoral shaft fractures
    Omar, Mohamed
    Suero, Eduardo M.
    Hawi, Nael
    Decker, Sebastian
    Krettek, Christian
    Citak, Musa
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2015, 135 (10) : 1385 - 1389
  • [40] Ipsilateral proximal and shaft femoral fractures treated with bridge-link type combined fixation system
    Kang, Liangqi
    Liu, Hui
    Ding, Zhenqi
    Ding, Yiqiang
    Hu, Wei
    Wu, Jin
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2020, 15 (01)