Magnetic Resonance Imaging in Patients with Recently Implanted Pacemakers

被引:36
作者
Friedman, Hannah L. [1 ]
Acker, Nancy [1 ]
Dalzell, Connie [1 ]
Shen, Win K. [2 ]
Asirvatham, Samuel J. [1 ]
Cha, Yong-Mei [1 ]
Hodge, David [3 ]
Felmlee, Joel [4 ]
Watson, Robert [4 ]
Friedman, Paul A. [1 ]
机构
[1] Mayo Clin, Heart Rhythm Serv, Rochester, MN 55905 USA
[2] Mayo Clin, Serv Cardiol, Scottsdale, AZ USA
[3] Mayo Clin, Serv Biostat, Rochester, MN 55905 USA
[4] Mayo Clin, Serv Radiol, Rochester, MN 55905 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2013年 / 36卷 / 09期
关键词
cardiac pacemaker; magnetic resonance imaging; INDUSTRY VIEWPOINT; CARDIOVERTER-DEFIBRILLATORS; MRI; ICDS; SAFETY;
D O I
10.1111/pace.12213
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Recent studies have shown that magnetic resonance imaging (MRI) of patients with pacemakers can be safely performed under careful monitoring, but they excluded patients with recently implanted devices. Patients with recent implants may be at a greater risk for complications during MRI imaging due to lack of lead and wound maturity. Methods We implemented a clinical protocol for MRI imaging of patients with implanted cardiac devices, and prospectively collected data. For this study, we retrospectively analyzed two groups of patients: those with recently implanted (42 days) and nonrecently implanted (>42 days) leads at the time of MRI scanning. All devices were interrogated before and after scanning, and were reprogrammed during the scan as per protocol. Results Of the 219 scans (in 171 patients), eight included patients with recently implanted (range: 7-36 days) and 211 with only nonrecently implanted pacemaker leads. During the scan, there were no complications in the early or late group. In one patient imaged 79 days postimplant, frequent premature ventricular complexes were noted during the scan, requiring no action. No patient reported pain during or immediately after the procedure. No clinically significant changes in function were seen at subsequent follow up (average 104 days post-MRI). Compared to patients with nonrecently implanted leads, there was no difference in any parameter between the two groups. Conclusions With a strong clinical indication and with careful monitoring, MRI imaging is feasible in patients with recently implanted pacemakers, although experience is limited.
引用
收藏
页码:1090 / 1095
页数:6
相关论文
共 15 条
[1]   Safety of magnetic resonance imaging in patients with permanent pacemakers: a collaborative clinical approach [J].
Boilson, Barry Anthony ;
Wokhlu, Anita ;
Acker, Nancy G. ;
Felmlee, Joel P. ;
Watson, Robert E., Jr. ;
Julsrud, Paul R. ;
Friedman, Paul A. ;
Cha, Yong-Mei ;
Rea, Robert F. ;
Hayes, David L. ;
Shen, Win-Kuang .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2012, 33 (01) :59-67
[2]  
Fuertes Beatriz, 2003, Indian Pacing Electrophysiol J, V3, P231
[3]   Successful Cervical MR Scan in a Patient Several Hours after Pacemaker Implantation [J].
Goldsher, Dorith ;
Jahshan, Shady ;
Roguin, Ariel .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (10) :1355-1356
[4]   Current clinical issues for MRI scanning of pacemaker and defibrillator patients [J].
Kalin, R ;
Stanton, MS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04) :326-328
[5]   Industry viewpoint: St. Jude Medical: Pacemakers, ICDs and MRI [J].
Levine, PA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04) :266-267
[6]   In vivo heating of pacemaker leads during magnetic resonance imaging [J].
Luechinger, R ;
Zeijlemaker, VA ;
Pedersen, EM ;
Mortensen, P ;
Falk, E ;
Duru, F ;
Candinas, R ;
Boesiger, P .
EUROPEAN HEART JOURNAL, 2005, 26 (04) :376-383
[7]   Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions [J].
Mollerus, Michael ;
Albin, Glenn ;
Lipinski, Margaret ;
Lucca, Jill .
EUROPACE, 2010, 12 (07) :947-951
[8]   A Prospective Evaluation of a Protocol for Magnetic Resonance Imaging of Patients With Implanted Cardiac Devices [J].
Nazarian, Saman ;
Hansford, Rozann ;
Roguin, Ariel ;
Goldsher, Dorith ;
Zviman, Menekhem M. ;
Lardo, Albert C. ;
Caffo, Brian S. ;
Frick, Kevin D. ;
Kraut, Michael A. ;
Kamel, Ihab R. ;
Calkins, Hugh ;
Berger, Ronald D. ;
Bluemke, David A. ;
Halperin, Henry R. .
ANNALS OF INTERNAL MEDICINE, 2011, 155 (07) :415-U39
[9]   Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices [J].
Roguin, Ariel ;
Schwitter, Juerg ;
Vahlhaus, Christian ;
Lombardi, Massimo ;
Brugada, Josep ;
Vardas, Panos ;
Auricchio, Angelo ;
Priori, Silvia ;
Sommer, Torsten .
EUROPACE, 2008, 10 (03) :334-344
[10]   Magnetic Resonance Imaging in Patients With Implantable Cardioverter-Defibrillators and Pacemakers [J].
Roguin, Ariel .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (06) :556-557