Fluoroscopic Radiation Exposure Are We Protecting Ourselves Adequately?

被引:29
作者
Hoffler, C. Edward [1 ]
Ilyas, Asif M. [1 ]
机构
[1] Thomas Jefferson Univ, Rothman Inst, Philadelphia, PA 19107 USA
关键词
C-ARM FLUOROSCOPY; IONIZING-RADIATION; MINI; SURGEONS; STANDARD; HANDS; RISK;
D O I
10.2106/JBJS.N.00839
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: While traditional intraoperative fluoroscopy protection relies on thyroid shields and aprons, recent data suggest that the surgeon's eyes and hands receive more exposure than previously appreciated. Using a distal radial fracture surgery model, we examined (1) radiation exposure to the eyes, thyroid, chest, groin, and hands of a surgeon mannequin; (2) the degree to which shielding equipment can decrease exposure; and (3) how exposure varies with fluoroscopy unit size. Methods: An anthropomorphic model was fit with radiation-attenuating glasses, a thyroid shield, an apron, and gloves. "Exposed" thermoluminescent dosimeters overlaid the protective equipment at the eyes, thyroid, chest, groin, and index finger while "shielded" dosimeters were placed beneath the protective equipment. Fluoroscopy position and settings were standardized. The mini-c-arm milliampere-seconds were fixed based on the selection of the kilovolt peak (kVp). Three mini and three standard c-arms scanned a model of the patient's wrist continuously for fifteen minutes each. Ten dosimeter exposures were recorded for each c-arm. Results: Hand exposure averaged 31 mu Sv/min (range, 22 to 48 mSv/min), which was 13.0 times higher than the other recorded exposures. Eye exposure averaged 4 mu Sv/min, 2.2 times higher than the mean thyroid, chest, and groin exposure. Gloves reduced hand exposure by 69.4%. Glasses decreased eye exposure by 65.6%. There was no significant difference in exposure between mini and standard fluoroscopy. Conclusions: Surgeons' hands receive the most radiation exposure during distal radial plate fixation under fluoroscopy. There was a small but insignificant difference in mean exposure between standard fluoroscopy and mini-fluoroscopy, but some standard units resulted in lower exposure than some mini-units. On the basis of these findings, we recommend routine protective equipment to mitigate exposure to surgeons' hands and eyes, in addition to the thyroid, chest, and groin, during fluoroscopy procedures.
引用
收藏
页码:721 / 725
页数:5
相关论文
共 19 条
[1]   Radiation exposure in hand surgery: Mini versus standard C-arm [J].
Athwal, GS ;
Bueno, RA ;
Wolfe, SW .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2005, 30A (06) :1310-1316
[2]  
Burns S, 2013, J BONE JOINT SURG AM, V95A, P1307, DOI [10.2106/JBJS.L.00893, 10.2106/JBJS.L00893]
[3]  
Bushberg J.T., 2012, ESSENTIAL PHYS MED I, V3
[4]   Risk of cataract after exposure to low doses of ionizing radiation: A 20-year prospective cohort study among US radiologic technologists [J].
Chodick, Gabriel ;
Bekiroglu, Nural ;
Hauptmann, Michael ;
Alexander, Bruce H. ;
Freedman, D. Michal ;
Doody, Michele Morin ;
Cheung, Li C. ;
Simon, Steven L. ;
Weinstock, Robert M. ;
Bouville, Andre ;
Sigurdson, Alice J. .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 2008, 168 (06) :620-631
[5]   Exposure to direct and scatter radiation with use of mini-c-arm fluoroscopy [J].
Giordano, Brian D. ;
Ryder, Steven ;
Baumhauer, Judith F. ;
DiGiovanni, Benedict F. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (05) :948-952
[6]   Patient and Surgeon Radiation Exposure: Comparison of Standard and Mini-C-Arm Fluoroscopy [J].
Giordano, Brian D. ;
Baumhauer, Judith F. ;
Morgan, Thomas L. ;
Rechtine, Glenn R., II .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (02) :297-304
[7]   Interventional cardiologists and risk of radiation-induced cataract: Results of a French multicenter observational study [J].
Jacob, Sophie ;
Boveda, Serge ;
Bar, Olivier ;
Brezin, Antoine ;
Maccia, Carlo ;
Laurier, Dominique ;
Bernier, Marie-Odile .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 167 (05) :1843-1847
[8]  
Kothan Suchart, 2011, American Journal of Applied Sciences, V8, P923, DOI 10.3844/ajassp.2011.923.926
[9]   Fractures of the distal part of the radius [J].
Koval, Kenneth J. ;
Harrast, John J. ;
Anglen, Jeffrey O. ;
Weinstein, James N. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (09) :1855-1861
[10]  
Lo N. N., 1996, Singapore Medical Journal, V37, P69