Pulpal blood flow recorded from exposed dentine with a laser Doppler flow meter using red or infrared light

被引:2
|
作者
Kijsamanmith, Kanittha [1 ]
Vongsavan, Noppakun [1 ]
Matthews, Bruce [2 ]
机构
[1] Mahidol Univ, Dept Oral Biol, Fac Dent, Yothi St, Bangkok 10400, Thailand
[2] Univ Bristol, Sch Physiol Pharmacol & Neurosci, Bristol BS8 1TD, Avon, England
关键词
laser doppler flow meter; Pulpal blood flow; Human premolar teeth; Exposed dentine; HUMAN TEETH; FLOWMETRY; CAVITY;
D O I
10.1016/j.archoralbio.2017.12.009
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To determine the percentage of the blood flow signal that is derived from dental pulp when recording from exposed dentine in a human premolar. Design: Recordings were made from 7 healthy teeth in 5 subjects (aged 22-33 yr.) with a laser Doppler flow meter (Periflux 4001) using either a red (635 nm) or an infrared (780 nm) laser. After exposing dentine above the buccal pulpal horn (cavity diam. 1.6 mm, depth 3 mm) and isolating the crown with opaque rubber dam, blood flow was recorded alternately with infrared or red light from the exposed dentine under four conditions: before and after injecting local anaesthetic (3% Mepivacaine without vasoconstrictor) (LA) over the apex of the root of the tooth; after exposing the pulp by cutting a buccal, class V cavity in the tooth; and after sectioning the coronal pulp transversely through the exposure. Results: There was no significant change in mean blood flow recorded with either light source when the tooth was anaesthetized or when the pulp was exposed. After the pulp had been sectioned, the blood flow recorded with infrared light fell by 67.8% and with red light, by 68.4%. The difference between these effects was not significant. Conclusions: When recording blood flow from exposed corona! dentine with either infrared or red light in a tooth isolated with opaque rubber dam, about 68% to the signal was contributed by the pulp. The signal:noise ratio was better with infrared than red light, and when recording from dentine than enamel.
引用
收藏
页码:163 / 167
页数:5
相关论文
共 27 条
  • [21] Possible application of transmitted laser light for the assessment of human pulp vitality. Part 2. Increased laser power for enhanced detection of pulpal blood flow
    Sasano, T
    Onodera, D
    Hashimoto, K
    Iikubo, M
    Satoh-Kuriwada, S
    Shoji, N
    Miyahara, T
    DENTAL TRAUMATOLOGY, 2005, 21 (01) : 37 - 41
  • [22] Orthostatic response of cephalic blood flow using a mini laser Doppler blood flowmeter and hemodynamics of a new active standing test
    Goma, Masaki
    Kimura, Yoshinori
    Shimura, Hiroki
    Kaneshige, Masahiro
    Kobayashi, Tetsuro
    Kikuchi, Masashi
    Terada, Nobuyuki
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2015, 115 (10) : 2167 - 2176
  • [23] Laser doppler flowmetry to detect pulp vitality, clinical reference range and coincidence rate for pulpal blood flow in permanent maxillary incisors in Chinese children: a clinical study
    Yang, Kuan
    Guo, Feifei
    Zhou, Zhifei
    Hui, Zeming
    Wang, Zirui
    Wang, Junhui
    Chen, Yujiang
    Ge, Xin
    Huang, Ruizhe
    Wang, Xiaojing
    BMC ORAL HEALTH, 2023, 23 (01)
  • [24] Orthostatic response of cephalic blood flow using a mini laser Doppler blood flowmeter and hemodynamics of a new active standing test
    Masaki Goma
    Yoshinori Kimura
    Hiroki Shimura
    Masahiro Kaneshige
    Tetsuro Kobayashi
    Masashi Kikuchi
    Nobuyuki Terada
    European Journal of Applied Physiology, 2015, 115 : 2167 - 2176
  • [25] Laser doppler flowmetry to detect pulp vitality, clinical reference range and coincidence rate for pulpal blood flow in permanent maxillary incisors in Chinese children: a clinical study
    Kuan Yang
    Feifei Guo
    Zhifei Zhou
    Zeming Hui
    Zirui Wang
    Junhui Wang
    Yujiang Chen
    Xin Ge
    Ruizhe Huang
    Xiaojing Wang
    BMC Oral Health, 23
  • [26] Nonstaining Blood Flow Imaging Using Optical Interference Due to Doppler Shift and Near-Infrared Imaging of Molecular Distribution in Developing Fish Egg Embryos
    Ishigaki, Mika
    Puangchit, Paralee
    Yasui, Yui
    Ishida, Akane
    Hayashi, Hiroki
    Nakayama, Yoshihiko
    Taniguchi, Hideya
    Ishimaru, Ichiro
    Ozaki, Yukihiro
    ANALYTICAL CHEMISTRY, 2018, 90 (08) : 5217 - 5223
  • [27] Assessment of abnormal blood flow and efficacy of treatment in patients with systemic sclerosis using a newly developed microwireless laser Doppler flowmeter and arm-raising test
    Kido, M.
    Takeuchi, S.
    Hayashida, S.
    Urabe, K.
    Sawada, R.
    Furue, M.
    BRITISH JOURNAL OF DERMATOLOGY, 2007, 157 (04) : 690 - 697