Micromotions at the taper interface between stem and neck adapter of a bimodular hip prosthesis during activities of daily living

被引:36
作者
Jauch, Sabrina Yvonne [1 ]
Huber, Gerd [1 ]
Sellenschloh, Kay [1 ]
Haschke, Henning [1 ]
Baxmann, Marc [2 ]
Grupp, Thomas M. [2 ,3 ]
Morlock, Michael M. [1 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, D-21073 Hamburg, Germany
[2] Aesculap AG Res & Dev, Tuttlingen, Germany
[3] Univ Munich, Clin Orthoped Surg, Munich, Germany
关键词
modular prosthesis; micromotion; fretting; stem-neck interface; taper; MODULAR FEMORAL-NECK; CREVICE CORROSION; FAILURE; ARTHROPLASTY; REPLACEMENT; RETRIEVAL; COMPONENT; FRACTURE;
D O I
10.1002/jor.22354
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The stem-neck taper interface of bimodular hip endoprostheses bears the risk of micromotions that can result in ongoing corrosion due to removal of the passive layer and ultimately cause implant fracture. We investigated the extent of micromotions at the stem-neck interface and the seating behavior of necks of one design made from different alloys during daily activities. Modular hip prostheses (n=36, Metha (R), Aesculap AG, Germany) with neck adapters (CoCr29Mo6 or Ti6Al4V) were embedded in PMMA (ISO 7206-4) and exposed to cyclic loading with peak loads ranging from walking (Fmax=2.3kN) to stumbling (Fmax=5.3kN). Translational and rotational micromotions at the taper interface and seating characteristics during assembly and loading were determined using four eddy-current sensors. Seating during loading after implant assembly was dependent on load magnitude but not on material coupling. Micromotions in the stem-neck interface correlated positively with load levels (CoCr: 2.6-6.3 mu m, Ti: 4.6-13.8 mu m; p<0.001) with Ti neck adapters exhibiting significantly larger micromotions than CoCr (p<0.001). These findings explain why high body weights and activities related to higher loads could increase the risk of fretting-induced implant failures in clinical application, especially for Ti-Ti combinations. Still, the role of taper seating is not clearly understood. (c) 2013 Orthopaedic Research Society Published by Wiley Periodicals, Inc. J Orthop Res 31:1165-1171, 2013
引用
收藏
页码:1165 / 1171
页数:7
相关论文
共 35 条
[1]  
Abdullah K., 2010, MODELING SIMULATION
[2]   Corrosion-Induced Fracture of a Double-Modular Hip Prosthesis A Case Report [J].
Atwood, Sara A. ;
Patten, Eli W. ;
Bozic, Kevin J. ;
Pruitt, Lisa A. ;
Ries, Michael D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (06) :1522-1525
[3]  
Australian Orthopaedic Association, 2011, AUSTR JOINT REPL REG
[4]   Realistic loads for testing hip implants [J].
Bergmann, G. ;
Graichen, F. ;
Rohlmann, A. ;
Bender, A. ;
Heinlein, B. ;
Duda, G. N. ;
Heller, M. O. ;
Morlock, M. M. .
BIO-MEDICAL MATERIALS AND ENGINEERING, 2010, 20 (02) :65-75
[5]   Two-Year Results of a Modular Short Hip Stem Prosthesis - A Prospective Study [J].
Braun, A. ;
Sabah, A. .
ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2009, 147 (06) :700-706
[6]   FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS [J].
BROWN, SA ;
FLEMMING, CAC ;
KAWALEC, JS ;
PLACKO, HE ;
VASSAUX, C ;
MERRITT, K ;
PAYER, JH ;
KRAAY, MJ .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :19-26
[7]   CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[8]  
Dennis DA, 2005, ORTHOPEDICS, V28, pS1049
[9]   Environmental effect on fretting of metallic materials for orthopaedic implants [J].
Duisabeau, L ;
Combrade, P ;
Forest, B .
WEAR, 2004, 256 (7-8) :805-816
[10]   Results of a Hydroxyapatite-Coated Modular Femoral Stem in Primary Total Hip Arthroplasty. A Minimum 5-Year Follow-Up [J].
Garcia-Rey, Eduardo ;
Munoz, Teresa ;
Montejo, Jorge ;
Martinez, Javier .
JOURNAL OF ARTHROPLASTY, 2008, 23 (08) :1132-1139