Fast design of local N-gram-specific absorption rate-optimized radiofrequency pulses for parallel transmit systems

被引:36
|
作者
Sbrizzi, Alessandro [1 ]
Hoogduin, Hans [1 ]
Lagendijk, Jan J. [1 ]
Luijten, Peter [1 ]
Sleijpen, Gerard L. G. [2 ]
van den Berg, Cornelis A. T. [1 ]
机构
[1] Univ Med Ctr Utrecht, Imaging Div, Utrecht, Netherlands
[2] Univ Utrecht, Dept Math, Utrecht, Netherlands
关键词
parallel transmission; multidimensional RF pulse design; transmit array; ultra-high-field MRI; local SAR reduction; RF PULSES; FIELD INHOMOGENEITY; EXCITATION; COIL; SENSE; ARRAY; POWER; MRI; SAR; 7T;
D O I
10.1002/mrm.23049
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Designing multidimensional radiofrequency pulses for clinical application must take into account the local specific absorption rate (SAR) as controlling the global SAR does not guarantee suppression of hot spots. The maximum peak SAR, averaged over an N grams cube (local NgSAR), must be kept under certain safety limits. Computing the SAR over a three-dimensional domain can require several minutes and implementing this computation in a radiofrequency pulse design algorithm could slow down prohibitively the numerical process. In this article, a fast optimization algorithm is designed acting on a limited number of control points, which are strategically selected locations from the entire domain. The selection is performed by comparing the largest eigenvalues and the corresponding eigenvectors of the matrices which locally describe the tissue's amount of heating. The computation complexity is dramatically reduced. An additional critical step to accelerate the computations is to apply a multi shift conjugate gradient algorithm. Two transmit array setups are studied: a two channel 3 T birdcage body coil and a 12-channel 7 T transverse electromagnetic (TEM) head coil. In comparison with minimum power radiofrequency pulses, it is shown that a reduction of 36.5% and 35%, respectively, in the local NgSAR can be achieved within short, clinically feasible, computation times. Magn Reson Med, 2012. (c) 2011 Wiley Periodicals, Inc.
引用
收藏
页码:824 / 834
页数:11
相关论文
共 9 条
  • [1] IMPULSE: A scalable algorithm for design of minimum specific absorption rate parallel transmit RF pulses
    Pendse, Mihir
    Stara, Riccardo
    Khalighi, Mohammad Mehdi
    Rutt, Brian
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (04) : 2808 - 2822
  • [2] Parallel Transmission Pulse Design with Explicit Control for the Specific Absorption Rate in the Presence of Radiofrequency Errors
    Martin, Adrian
    Schiavi, Emanuele
    Eryaman, Yigitcan
    Herraiz, Joaquin L.
    Gagoski, Borjan
    Adalsteinsson, Elfar
    Wald, Lawrence L.
    Guerin, Bastien
    MAGNETIC RESONANCE IN MEDICINE, 2016, 75 (06) : 2493 - 2504
  • [3] Specific Absorption Rate Reduction in Parallel Transmission by k-Space Adaptive Radiofrequency Pulse Design
    Homann, Hanno
    Graesslin, Ingmar
    Nehrke, Kay
    Findeklee, Christian
    Doessel, Olaf
    Boernert, Peter
    MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (02) : 350 - 357
  • [4] Design of Parallel Transmission Pulses for Simultaneous Multislice with Explicit Control for Peak Power and Local Specific Absorption Rate
    Guerin, Bastien
    Setsompop, Kawin
    Ye, Huihui
    Poser, Benedikt A.
    Stenger, Andrew V.
    Wald, Lawrence L.
    MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (05) : 1946 - 1953
  • [5] Global and peak local specific absorption rate control on parallel transmit systems usingk-means SAR compression model
    Mao, Xianglun
    Bari, Sumra
    Love, David J.
    Rispoli, Joseph V.
    MAGNETIC RESONANCE IN MEDICINE, 2021, 85 (02) : 1093 - 1103
  • [6] Comparison of tight-fitting 7T parallel-transmit head array designs using excitation uniformity and local specific absorption rate metrics
    Kazemivalipour, Ehsan
    Wald, Lawrence L.
    Guerin, Bastien
    MAGNETIC RESONANCE IN MEDICINE, 2024, 91 (03) : 1209 - 1224
  • [7] Local Specific Absorption Rate (SAR), Global SAR, Transmitter Power, and Excitation Accuracy Trade-Offs in Low Flip-Angle Parallel Transmit Pulse Design
    Guerin, Bastien
    Gebhardt, Matthias
    Cauley, Steven
    Adalsteinsson, Elfar
    Wald, Lawrence L.
    MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (04) : 1446 - 1457
  • [8] Fast Low-Specific Absorption Rate B0-Mapping along Projections at High Field Using Two-Dimensional Radiofrequency Pulses
    Reynaud, Olivier
    Gallichan, Daniel
    Schaller, Benoit
    Gruetter, Rolf
    MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (03) : 901 - 908
  • [9] Parallel-transmission spatial spectral pulse design with local specific absorption rate control: Demonstration for robust uniform water-selective excitation in the human brain at 7 T
    Shao, Xin
    Zhang, Zhe
    Ma, Xiaodong
    Liu, Fan
    Guo, Hua
    Ugurbil, Kamil
    Wu, Xiaoping
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (03) : 1238 - 1255