Radiation Dose Reduction via Sinogram Affirmed Iterative Reconstruction and Automatic Tube Voltage Modulation (CARE kV) in Abdominal CT

被引:27
作者
Shin, Hyun Joo [1 ]
Chung, Yong Eun [1 ]
Lee, Young Han [1 ]
Choi, Jin-Young [1 ]
Park, Mi-Suk [1 ]
Kim, Myeong-Jin [1 ]
Kim, Ki Whang [1 ]
机构
[1] Yonsei Univ, Severance Hosp, Coll Med, Res Inst Radiol Sci,Dept Radiol, Seoul 120752, South Korea
关键词
Computed tomography; Radiation dose reduction; Iterative reconstruction; Tube voltage modulation; IMAGE SPACE; COMPUTED-TOMOGRAPHY; CHEST CT; OPTIMIZATION; QUALITY; ALGORITHM; SELECTION; MDCT;
D O I
10.3348/kjr.2013.14.6.886
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate the feasibility of sinogram-affirmed iterative reconstruction (SAFIRE) and automated kV modulation (CARE kV) in reducing radiation dose without increasing image noise for abdominal CT examination. Materials and Methods: This retrospective study included 77 patients who received CT imaging with an application of CARE kV with or without SAFIRE and who had comparable previous CT images obtained without CARE kV or SAFIRE,. using the standard dose (i.e., reference mAs of 240) on an identical CT scanner and reconstructed with filtered back projection (FBP) within 1 year. Patients were divided into two groups: group A (33 patients, CT scanned with CARE kV); and group B (44 patients, scanned after reducing the reference mAs from 240 to 170 and applying both CARE kV and SAFIRE). CT number, image noise for four organs and radiation dose were compared among the two groups. Results: Image noise increased after CARE kV application (p < 0.001) and significantly decreased as SAFIRE strength increased (p < 0.001). Image noise with reduced-mAs scan (170 mAs) in group B became similar to that of standard-dose FBP images after applying CARE kV and SAFIRE strengths of 3 or 4 when measured in the aorta, liver or muscle (p >= 0.108). Effective doses decreased by 19.4% and 41.3% for groups A and B, respectively (all, p < 0.001) after application of CARE kV with or without SAFIRE. Conclusion: Combining CARE kV, reduction of mAs from 240 to 170 mAs and noise reduction by applying SAFIRE strength 3 or 4 reduced the radiation dose by 41.3% without increasing image noise compared with the standard-dose FBP images.
引用
收藏
页码:886 / 893
页数:8
相关论文
共 24 条
[1]  
Barrett H.H., 1981, RADIOLOGICAL IMAGING
[2]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[3]   Computed tomography-old ideas and new technology [J].
Fleischmann, Dominik ;
Boas, F. Edward .
EUROPEAN RADIOLOGY, 2011, 21 (03) :510-517
[4]   CT iterative reconstruction in image space: A phantom study [J].
Ghetti, C. ;
Ortenzia, O. ;
Serreli, G. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :161-165
[5]   CT Radiation Dose Optimization and Estimation: an Update for Radiologists [J].
Goo, Hyun Woo .
KOREAN JOURNAL OF RADIOLOGY, 2012, 13 (01) :1-11
[6]   Recent Technologic Advances in Multi-Detector Row Cardiac CT [J].
Halliburton, Sandra Simon .
CARDIOLOGY CLINICS, 2009, 27 (04) :655-+
[7]  
Hart D., 2002, NRPBW4
[8]  
Hur S, 2012, KOREAN J RADIOL, V13, P152, DOI [10.3348/kjr.2012.13.2.152, 10.7874/kja.2012.16.3.152]
[9]   Radiation Dose Reduction of Chest CT with Iterative Reconstruction in Image Space - Part I: Studies on Image Quality Using Dual Source CT [J].
Hwang, Hye Jeon ;
Seo, Joon Beom ;
Lee, Jin Seong ;
Song, Jae-Woo ;
Kim, Song Soo ;
Lee, Hyun Joo ;
Lim, Chae Hun .
KOREAN JOURNAL OF RADIOLOGY, 2012, 13 (06) :711-719
[10]   Radiation Dose Reduction of Chest CT with Iterative Reconstruction in Image Space - Part II: Assessment of Radiologists' Preferences Using Dual Source CT [J].
Hwang, Hye Jeon ;
Seo, Joon Beom ;
Lee, Jin Seong ;
Song, Jae-Woo ;
Kim, Song Soo ;
Lee, Hyun Joo ;
Lim, Chae Hun .
KOREAN JOURNAL OF RADIOLOGY, 2012, 13 (06) :720-727