Analysis on the selection of optimal ADC parameters in indirect measurement

被引:0
|
作者
Zhu, Xiaoming [1 ,2 ]
Wang, Xiaodong [1 ]
Xu, Zhijun [1 ]
Li, Bingyu [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Opt Fine Mech & Phys, Changchun 130033, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
关键词
CRDS; Signal progressing; ADC; RING-DOWN SPECTROSCOPY; ABSORPTION-MEASUREMENTS; OPTICAL CAVITY; SPECTROMETER; AMMONIA; TIME;
D O I
10.1016/j.infrared.2015.04.013
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
How to select optimal ADC parameters in indirect measurement including resolution, sampling rate and sampling error of ADC, and system error is proposed. Fitting of the photo time constant in cavity ring-down spectrum is taken as an example. The effects of resolution, sampling rate and sampling error (with/without system error) of ADC on fitting accuracy are simulated sequentially. The influence of mirror reflectivity on the parameters of fitting functions is analyzed and simulated. Parameters of ADC to distinguish concentration of methane of 1 ppbv at different mirror reflectivity are given. (C) 2015 Published by Elsevier B.V.
引用
收藏
页码:501 / 505
页数:5
相关论文
共 50 条
  • [1] Optimal Selection of Switch Model Parameters for ADC-Based Power Converters
    Alsarayreh, Saif
    Suto, Zoltan
    ENERGIES, 2024, 17 (01)
  • [2] Selection of measurement frequencies for optimal extraction of tissue impedance model parameters
    Kun, S.
    Peura, R.A.
    Medical and Biological Engineering and Computing, 1999, 37 (06): : 699 - 703
  • [3] Selection of measurement frequencies for optimal extraction of tissue impedance model parameters
    S. Kun
    R. A. Peura
    Medical & Biological Engineering & Computing, 1999, 37 : 699 - 703
  • [4] Selection of measurement frequencies for optimal extraction of tissue impedance model parameters
    Kun, S
    Peura, RA
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (06) : 699 - 703
  • [5] The analysis of optimal selection on the parameters of the ant colony algorithm
    Zhang, Yi
    Liang, Yanchun
    DYNAMICS OF CONTINUOUS DISCRETE AND IMPULSIVE SYSTEMS-SERIES B-APPLICATIONS & ALGORITHMS, 2007, 14 : 130 - 133
  • [6] Optimal Selection of ADC Frequencies in Multirate Sampling Receivers
    Pinarcik, A. Burak
    Korucu, A. Bugra
    Alp, Y. Kemal
    Gok, Gokhan
    Durak-Ata, Lutfiye
    Arikan, Orhan
    2020 28TH SIGNAL PROCESSING AND COMMUNICATIONS APPLICATIONS CONFERENCE (SIU), 2020,
  • [7] OPTIMAL SELECTION OF FLASKS FOR INDIRECT METHOD OF MANOMETRY
    REES, ED
    ROSS, A
    ANALYTICAL BIOCHEMISTRY, 1964, 7 (03) : 275 - &
  • [8] Genetic parameters, path analysis and indirect selection of agronomic traits of cassava germplasm
    Diniz, Rafael P.
    De Oliveira, Eder J.
    ANAIS DA ACADEMIA BRASILEIRA DE CIENCIAS, 2019, 91 (03):
  • [9] Joint Measurement of Signal Parameters and ADC Transition Levels
    Carbone, Paolo
    Schoukens, Johan
    Moschitta, Antonio
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2018, 67 (12) : 2752 - 2760
  • [10] Determining the optimal allocation of parameters for multivariate measurement system analysis
    Pan, Jeh-Nan
    Li, Chung-I
    Ou, Szu-Chen
    EXPERT SYSTEMS WITH APPLICATIONS, 2015, 42 (20) : 7036 - 7045