Comparison of the use of new handheld tonometers and established applanation tonometers in dogs

被引:62
作者
Görig, C
Coenen, RTI
Stades, FC
Djajadiningrat-Laanen, SC
Boevé, MH
机构
[1] Univ Utrecht, Fac Med Vet, Dept Clin Sci Compan Anim, NL-3508 TD Utrecht, Netherlands
[2] Anim Clin Groenlo Lichtenvoorde, NL-7131 LB Lichtenvoorde, Netherlands
关键词
D O I
10.2460/ajvr.67.1.134
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective-To examine the practical aspects, accuracy, and reproducibility of 2 new automatic handheld tonometers in dogs and compare them with results for 2 established applanation tonometers. Animals-15 freshly enucleated canine eyes for manometric evaluation and 20 conscious research dogs, 20 client-owned dogs, and 12 dogs with acute glaucoma for clinical tonometry. Procedure-Calibration curves were determined for all 4 tonometers on 15 enucleated canine eyes. Intraocular pressure (IOP) was measured with each tonometer consecutively in conscious dogs, with the MacKay-Marg applanation tonometer as the reference device. Measurements were repeated in 20 sedated dogs. An induction impact tonometer was evaluated clinically on dogs with acute glaucoma. Additionally, measurements obtained by an experienced and an inexperienced examiner and with or without use of topical anesthesia were compared. Results-The portable pneumatonometer was cumbersome and time-consuming. Compared with results for the reference applanation tonometer, and confirmed by manometry, the portable pneumatonometer increasingly underestimated actual IOP values with increasing IOP The induction-impact tonometer provided accurate and reproducible measurement values. There was a significant strong correlation between the IOP values obtained by the 2 examiners (r(2), 0.82) and also with or without topical anesthesia (r(2), 0.86). In dogs with glaucoma, the fitted line comparing values for the reference applanation tonometer and induction-impact tonometer closely resembled an ideal 1:1 relationship. Conclusions and Clinical Relevance-Use of the portable pneumatonometer in dogs appears to have disadvantages. The induction-impact tonometer appears to provide a promising alternative to the use of applanation tonometers in dogs.
引用
收藏
页码:134 / 144
页数:11
相关论文
共 37 条
[1]   EVALUATION OF THE TONO-PEN AND THE PULSAIR TONOMETERS [J].
ARMSTRONG, TA .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1990, 109 (06) :716-720
[2]   INFLUENCE OF TOPICAL ANESTHESIA ON TONOMETRIC VALUES OF INTRAOCULAR-PRESSURE [J].
BAUDOUIN, C ;
GASTAUD, P .
OPHTHALMOLOGICA, 1994, 208 (06) :309-313
[3]   THE TONO-PEN - A MANOMETRIC AND CLINICAL-STUDY [J].
BOOTHE, WA ;
LEE, DA ;
PANEK, WC ;
PETTIT, TH .
ARCHIVES OF OPHTHALMOLOGY, 1988, 106 (09) :1214-1217
[4]  
BRUBAKER RF, 1993, DUANES CLIN OPHTHALM, V3, P1
[5]  
DZIEZYC J, 1992, J AM VET MED ASSOC, V201, P430
[6]  
GELATT KN, 1981, INVEST OPHTH VIS SCI, V20, P683
[7]  
GELATT KN, 1981, AM J VET RES, V42, P1778
[8]  
GELATT KN, 1977, INVEST OPHTH VIS SCI, V16, P963
[9]   Non-invasive determination of intraocular pressure in the rat eye. Comparison of an electronic tonometer (TonoPen), and a rebound (impact probe) tonometer [J].
Goldblum, D ;
Kontiola, AI ;
Mittag, T ;
Chen, B ;
Danias, J .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2002, 240 (11) :942-946
[10]  
GOLDMAN H, 1957, S GLAUC T 2 C NEW YO, P186