Polarization-sensitive optical coherence tomography based on polarization-maintaining fibers and frequency multiplexing

被引:55
作者
Al-Qaisi, Muhammad K. [1 ]
Akkin, Taner [1 ]
机构
[1] Univ Minnesota, Dept Biomed Engn, Minneapolis, MN 55455 USA
基金
美国国家卫生研究院;
关键词
D O I
10.1364/OE.16.013032
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report a novel polarization-maintaining fiber based optical coherence tomography for single detector imaging of tissue reflectivity and birefringence. A single depth scan yields quantitative birefringence information along the A-line accurately. Since the orthogonal polarization channels are frequency multiplexed, the polarization information is extracted by using digital band-pass filters. Here, we introduce the optical system and present the reflectivity and birefringence images of biological tissues with an axial resolution of 7.9 mu m and SNR of 30 dB. (C) 2008 Optical Society of America.
引用
收藏
页码:13032 / 13041
页数:10
相关论文
共 29 条
[1]   Single camera based spectral domain polarization sensitive optical coherence tomography [J].
Baumann, Bernhard ;
Goetzinger, Erich ;
Pircher, Michael ;
Hitzenberger, Christoph K. .
OPTICS EXPRESS, 2007, 15 (03) :1054-1063
[2]   Polarization-sensitive spectral-domain optical coherence tomography using a single line scan camera [J].
Cense, Barry ;
Mujat, Mircea ;
Chen, Teresa C. ;
Park, B. Hyle ;
de Boer, Johannes F. .
OPTICS EXPRESS, 2007, 15 (05) :2421-2431
[3]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[4]   Polarization-maintaining fiber-based optical low-coherence reflectometer for characterization and ranging of birefringence [J].
Davé, DP ;
Akkin, T ;
Milner, TE .
OPTICS LETTERS, 2003, 28 (19) :1775-1777
[5]   Optical low-coherence reflectometer for differential phase measurement [J].
Davé, DP ;
Milner, TE .
OPTICS LETTERS, 2000, 25 (04) :227-229
[6]   Phase-sensitive frequency-multiplexed optical low-coherence reflectometery [J].
Davé, DP ;
Akkin, T ;
Milner, TE ;
Rylander, HG .
OPTICS COMMUNICATIONS, 2001, 193 (1-6) :39-43
[7]   Polarization effects in optical coherence tomography of various biological tissues [J].
de Boer, JF ;
Srinivas, SM ;
Park, BH ;
Pham, TH ;
Chen, ZP ;
Milner, TE ;
Nelson, JS .
IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 1999, 5 (04) :1200-1204
[8]   Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography [J].
de Boer, JF ;
Cense, B ;
Park, BH ;
Pierce, MC ;
Tearney, GJ ;
Bouma, BE .
OPTICS LETTERS, 2003, 28 (21) :2067-2069
[9]   Determination of the depth-resolved Stokes parameters of light backscattered from turbid media by use of polarization-sensitive optical coherence tomography [J].
de Boer, JF ;
Milner, TE ;
Nelson, JS .
OPTICS LETTERS, 1999, 24 (05) :300-302
[10]  
Drexler W, 2001, J RHEUMATOL, V28, P1311