Effects of motion, ambient light, and hypoperfusion on pulse oximeter function

被引:84
作者
Trivedi, NS
Ghouri, AF
Shah, NK
Lai, E
Barker, SJ
机构
[1] Department of Anesthesiology, University of California at Irvine, Orange, CA
[2] Department of Anesthesiology, University of Arizon at Tuczon, Tucson, AZ
[3] Department of Anesthesiology, University of California at Irvine, Orange, CA 92668, 101 City Drive South
关键词
oxygen saturation; desaturation; pulse oximeter; failure of;
D O I
10.1016/S0952-8180(97)00039-1
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Study Objective: To compare the performance of five pulse oximeters during hypopfusion, probe motion, and exposure to ambient light interference. Design: Prospective study. Setting: Laboratory facility at a university medical center. Patients: 8 unanesthetized, ASA physical status I volunteers. Interventions: We evaluated five common pulse oximeters with respect to three scenarios: (1) an operating room light was shone on oximeter probes, (2) a motion generator was used to generate 2 Hz and 4 Hz hand motion, and (3) a pneumatic compression device overlying the brachial artery was used to simulate hypoperfusion. Electrocardiographic (EGG) and arterial blood gas values were considered gold standards for heart rate (HR) and oxygen saturation. (SpO(2)) respectively. SpO(2) nondisplay and values greater than 4% from simultaneous arterial SaO(2)-oximeter values were defined as errors. Nondisplay of HR, or HR greater than 5 % from ECG values, were also considered errors. Measurements and Main Results: The Ohmeda and Nellcor N200 with finger probe had the highest total failure rates with respect to both SpO(2) and HR due to ambient light interference (p < 0.05). The Nellcor N200 with finger probe and N200 with C lodz were the most accurate with regard to SpO(2) during 2 Hz and 4 Hz motion (p < 0.05). However, all oximeters failed dramatically during 4 Hz motion when measuring HR lit the hypoperfusion model, the Nellcor N200 with finger probe and the Nellcor C Lock oximeters performed significantly better than all others in term; of both HR and SpO, (p < 0.05), while the Criticare oximeter failed 100% of the time. Conclusion: There are significant differences in the accuracy of commercially available pulse oximeters during nonideal circumstances, with failure rates varying from approximately 5% to 50% depending on the oximeter and source of interference Furthermore, no single oximeter performed the best under all conditions. (C) 1997 by Elsevier Science Inc.
引用
收藏
页码:179 / 183
页数:5
相关论文
共 10 条
  • [1] Robust Design of Pulse Oximeter Using Dynamic Control and Motion Artifact Detection Algorithms
    Cho, Jung Hyun
    Kim, Jong Cheol
    Yoon, Gil Won
    JOURNAL OF ELECTRICAL ENGINEERING & TECHNOLOGY, 2014, 9 (05) : 1780 - 1787
  • [2] A Low-Noise Analog Front End with Interstage Systematically Ambient Interference Cancellation for Pulse Oximeter
    Shen, Jiulong
    Lin, Zhi
    Huang, Linqing
    Ruan, Xiukai
    Tang, Fang
    ELECTRONICS, 2022, 11 (06)
  • [3] Development of a Wearable Reflection-Type Pulse Oximeter System to Acquire Clean PPG Signals and Measure Pulse Rate and SpO2 with and without Finger Motion
    Banik, Partha Pratim
    Hossain, Shifat
    Kwon, Tae-Ho
    Kim, Hyoungkeun
    Kim, Ki-Doo
    ELECTRONICS, 2020, 9 (11) : 1 - 26
  • [4] Light-guided dynamic phantom to mimic microvasculature for biomedical applications: an exploration for pulse oximeter
    Ma, Hui
    Angelone, Dario
    Guadagno, Claudia Nunzia
    Andersson-Engels, Stefan
    Sekar, Sanathana Konugolu Venkata
    JOURNAL OF BIOMEDICAL OPTICS, 2024, 29
  • [5] Effects of motion on the performance of pulse oximeters in volunteers
    Barker, SJ
    Shah, NK
    ANESTHESIOLOGY, 1996, 85 (04) : 774 - 781
  • [6] Interference of coronal enamel and dentin thickness and ambient light on pulse oximetry interpretation
    Silva, Julio Almeida
    Goncalves de Alencar, Ana Helena
    Sestari, Larissa Emanuelle
    Barbosa, Higor Almeida
    de Siqueira, Patricia Correia
    Decurcio, Daniel Almeida
    Barletta, Fernando Branco
    Estrela, Carlos
    BRAZILIAN ORAL RESEARCH, 2020, 34
  • [7] Monitoring of haemoglobin oxygen saturation in healthy infants using a new generation pulse oximeter which takes motion artifacts into account
    Meyts, I
    Van Reempts, P
    De Boeck, K
    EUROPEAN JOURNAL OF PEDIATRICS, 2002, 161 (12) : 653 - 655
  • [8] Pulse oximeter using a gain-modulated avalanche photodiode operated in a pseudo-lock-in light detection mode
    Miyata, T
    Iwata, T
    Araki, T
    ICO20: BIOMEDICAL OPTICS, 2006, 6026
  • [9] The effects of motion on the performance of pulse oximeters in volunteers (revised publication)
    Barker, SJ
    Shah, NK
    ANESTHESIOLOGY, 1997, 86 (01) : 101 - 108
  • [10] Posture effects on the calibratability of remote pulse oximetry in visible light
    Moco, A.
    Stuijk, S.
    de Haan, G.
    PHYSIOLOGICAL MEASUREMENT, 2019, 40 (03)