Development and testing of a silicone in vitro model of descending aortic dissection

被引:14
作者
Birjiniuk, Joav [1 ]
Ruddy, Jean Marie [2 ]
Iffrig, Elizabeth [1 ]
Henry, Travis S. [3 ]
Leshnower, Bradley G. [4 ]
Oshinski, John N. [1 ,3 ]
Ku, David N. [5 ]
Veeraswamy, Ravi K. [6 ]
机构
[1] Georgia Inst Technol, Wallace H Coulter Dept Biomed Engn, Atlanta, GA 30332 USA
[2] Med Univ S Carolina, Dept Surg, Div Vasc Surg, Charleston, SC 29425 USA
[3] Emory Univ, Sch Med, Dept Radiol & Imaging Sci, Atlanta, GA USA
[4] Emory Univ, Sch Med, Joseph B Whitehead Dept Surg, Div Cardiothorac Surg, Atlanta, GA USA
[5] Georgia Inst Technol, Coll Engn, George W Woodruff Sch Mech Engn, Atlanta, GA 30332 USA
[6] Emory Univ, Sch Med, Div Vasc Surg, Joseph B Whitehead Dept Surg, Atlanta, GA USA
关键词
Aortic dissection; Flow model; Translational research; 4D flow MRI; TRUE-LUMEN; ENTRY TEARS; PART II; THERAPY; SIZE;
D O I
10.1016/j.jss.2015.03.024
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Stanford type B dissection of the descending aorta is a potentially fatal condition that is poorly understood. Limited scientific understanding of the role of current interventional techniques, as well as heterogeneity in the condition, contributes to lack of consensus as to the most effective treatment strategy. This study introduces an anatomically accurate model for investigating aortic dissection in a laboratory setting. Materials and methods: A silicone model was fabricated and filled with fluid to mimic human blood. Flow was established, and the model was scanned using a four-dimensional flow magnetic resonance imaging protocol. On analysis, luminal flow rates were quantified by multiplying local velocity by included area. Results: The upstream total flow was compared with the sum of the flow in the true and false lumens. The two values were within the margin of error. Furthermore, flow rates matched with the relative areas of each compartment. Conclusions: These results validate our model as a novel and unique system that mimics a type B aortic dissection and will allow for more sophisticated analysis of dissection physiology in future studies. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:502 / 507
页数:6
相关论文
共 26 条
[1]   Nanoindentation of histological specimens: Mapping the elastic properties of soft tissues [J].
Akhtar, R. ;
Schwarzer, N. ;
Sherratt, M. J. ;
Watson, R. E. B. ;
Graham, H. K. ;
Trafford, A. W. ;
Mummery, P. M. ;
Derby, B. .
JOURNAL OF MATERIALS RESEARCH, 2009, 24 (03) :638-646
[2]  
[Anonymous], CHAR PROP SIL RUBB C
[3]   True-lumen collapse in aortic dissection - Part II. Evaluation of treatment methods in phantoms with pulsatile flow [J].
Chung, JW ;
Elkins, C ;
Sakai, T ;
Kato, N ;
Vestring, T ;
Semba, CP ;
Slonim, SM ;
Dake, MD .
RADIOLOGY, 2000, 214 (01) :99-106
[4]   A new imaging method for assessment of aortic dissection using four-dimensional phase contrast magnetic resonance imaging [J].
Clough, Rachel E. ;
Waltham, Matthew ;
Giese, Daniel ;
Taylor, Peter R. ;
Schaeffter, Tobias .
JOURNAL OF VASCULAR SURGERY, 2012, 55 (04) :914-923
[5]   A new method for quantification of false lumen thrombosis in aortic dissection using magnetic resonance imaging and a blood pool contrast agent [J].
Clough, Rachel E. ;
Hussain, Tarique ;
Uribe, Sergio ;
Greil, Gerald F. ;
Razavi, Reza ;
Taylor, Peter R. ;
Schaeffter, Tobias ;
Waltham, Matthew .
JOURNAL OF VASCULAR SURGERY, 2011, 54 (05) :1251-1258
[6]   Long-Term Outcome of Aortic Dissection With Patent False Lumen Predictive Role of Entry Tear Size and Location [J].
Evangelista, Artur ;
Salas, Armando ;
Ribera, Aida ;
Ferreira-Gonzalez, Ignacio ;
Cuellar, Hug ;
Pineda, Victor ;
Gonzalez-Alujas, Teresa ;
Bijnens, Bart ;
Permanyer-Miralda, Gaieta ;
Garcia-Dorado, David .
CIRCULATION, 2012, 125 (25) :3133-+
[7]   Microstructural and biomechanical alterations of the human aorta as a function of age and location [J].
Haskett, Darren ;
Johnson, Gregory ;
Zhou, Aifang ;
Utzinger, Urs ;
Vande Geest, Jonathan .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2010, 9 (06) :725-736
[8]   The results of stent graft versus medication therapy for chronic type B dissection [J].
Jia, Xin ;
Guo, Wei ;
Li, Tian-xiao ;
Guan, Sheng ;
Yang, Rui-min ;
Liu, Xiao-ping ;
Zhang, Min-hong ;
Xiong, Jiang .
JOURNAL OF VASCULAR SURGERY, 2013, 57 (02) :406-414
[9]   Aortic dissection: CT features that distinguish true lumen from false lumen [J].
LePage, MA ;
Quint, LE ;
Sonnad, SS ;
Deeb, GM ;
Williams, DM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (01) :207-211
[10]   4D flow MRI [J].
Markl, Michael ;
Frydrychowicz, Alex ;
Kozerke, Sebastian ;
Hope, Mike ;
Wieben, Oliver .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (05) :1015-1036