Evaluation of Elekta 4D cone beam CT-based automatic image registration for radiation treatment of lung cancer

被引:12
作者
Li, Jun [1 ]
Harrison, Amy [1 ]
Yu, Yan [1 ]
Xiao, Ying [1 ]
Werner-Wasik, Maria [1 ]
Lu, Bo [1 ]
机构
[1] Thomas Jefferson Univ, Dept Radiat Oncol, Philadelphia, PA 19107 USA
关键词
STEREOTACTIC RADIOTHERAPY; INTRAPULMONARY TUMORS; GUIDED RADIOTHERAPY; GUIDANCE; LANDMARKS; LOCALIZATION; ACCURACY; CARINA; ERRORS; SPINE;
D O I
10.1259/bjr.20140620
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The study was aimed to evaluate the precision of Elekta four-dimensional (4D) cone beam CT (CBCT)-based automatic dual-image registrations using different landmarks for clipbox for radiation treatment of lung cancer. Methods: 30 4D CBCT scans from 15 patients were studied. 4D CBCT images were registered with reference CT images using dual-image registration: a clipbox registration and a mask registration. The image registrations performed in clinic using a physician-defined clipbox, were reviewed by physicians, and were taken as the standard. Studies were conducted to evaluate the automatic dual registrations using three kinds of landmarks for clipbox: spine, spine plus internal target volume (ITV) and lung (including as much of the lung as possible). Translational table shifts calculated from the automatic registrations were compared with those of the standard. Results: The mean of the table shift differences in the lateral direction were 0.03, 0.03 and 0.03 cm, for clipboxes based on spine, spine plus ITV and lung, respectively. The mean of the shift differences in the longitudinal direction were 0.08, 0.08 and 0.08 cm, respectively. The mean of the shift differences in the vertical direction were 0.03, 0.03 and 0.03 cm, respectively. Conclusion: The automatic registrations using three different landmarks for clipbox showed similar results. One can use any of the three landmarks in 4D CBCT dual-image registration. Advance in knowledge: The study provides knowledge and recommendations for application of Elekta 4D CBCT image registration in radiation therapy of lung cancer.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Quantitative evaluation of 4D Cone beam CT scans with reduced scan time in lung cancer patients
    Bryce-Atkinson, Abigail
    Marchant, Thomas
    Rodgers, John
    Budgell, Geoff
    McWilliam, Alan
    Faivre-Finn, Corinne
    Whitfield, Gillian
    van Herk, Marcel
    RADIOTHERAPY AND ONCOLOGY, 2019, 136 : 64 - 70
  • [2] Analysis of the setup errors of medical image registration-based cone-beam CT for lung cancer
    Li, Jun
    Tang, Xiao-Bin
    Zhang, Xi-Zhi
    Zhang, Xian-Wen
    Ge, Yun
    Chen, Da
    Chai, Lei
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (04) : 521 - 530
  • [3] Iterative volume of interest based 4D cone-beam CT
    Martin, Rachael
    Ahmad, Moiz
    Hugo, Geoffrey
    Pan, Tinsu
    MEDICAL PHYSICS, 2017, 44 (12) : 6515 - 6528
  • [4] CONE-BEAM COMPUTED TOMOGRAPHIC IMAGE GUIDANCE FOR LUNG CANCER RADIATION THERAPY
    Bissonnette, Jean-Pierre
    Purdie, Thomas G.
    Higgins, Jane A.
    Li, Winnie
    Bezjak, Andrea
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (03): : 927 - 934
  • [5] Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumors
    Sweeney, Reinhart A.
    Seubert, Benedikt
    Stark, Silke
    Homann, Vanessa
    Mueller, Gerd
    Flentje, Michael
    Guckenberger, Matthias
    RADIATION ONCOLOGY, 2012, 7
  • [6] DOSIMETRIC IMPACT OF ONLINE CORRECTION VIA CONE-BEAM CT-BASED IMAGE GUIDANCE FOR STEREOTACTIC LUNG RADIOTHERAPY
    Galerani, Ana Paula
    Grills, Inga
    Hugo, Geoffrey
    Kestin, Larry
    Mohammed, Nasiruddin
    Chao, K. Kenneth
    Suen, Andrew
    Martinez, Alvaro
    Yan, Di
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (05): : 1571 - 1578
  • [7] Cone beam CT verification for oesophageal cancer - impact of volume selected for image registration
    Hawkins, Maria A.
    Aitken, Alexandra
    Hansen, Vibeke N.
    McNair, Helen A.
    Tait, Diana M.
    ACTA ONCOLOGICA, 2011, 50 (08) : 1183 - 1190
  • [8] Evaluation of the cone beam CT for internal target volume localization in lung stereotactic radiotherapy in comparison with 4D MIP images
    Wang, Lu
    Chen, Xiaoming
    Lin, Mu-Han
    Xue, Jun
    Lin, Teh
    Fan, Jiajin
    Jin, Lihui
    Ma, Charlie M.
    MEDICAL PHYSICS, 2013, 40 (11)
  • [9] Actively triggered 4d cone-beam CT acquisition
    Fast, Martin F.
    Wisotzky, Eric
    Oelfke, Uwe
    Nill, Simeon
    MEDICAL PHYSICS, 2013, 40 (09)
  • [10] Evaluation of Deformable Image Registration-Based Contour Propagation From Planning CT to Cone-Beam CT
    Woerner, Andrew J.
    Choi, Mehee
    Harkenrider, Matthew M.
    Roeske, John C.
    Surucu, Murat
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (06) : 801 - 810