A calculation method based on isotope ratios for the determination of dead time and its uncertainty in ICP-MS and application of the method to investigating some features of a continuous dynode multiplier

被引:48
作者
Held, A [1 ]
Taylor, PDP [1 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
关键词
D O I
10.1039/a809098j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method for the determination of dead time (and its uncertainty) of the ion-counting detection system of ICP-MS instruments is presented. It is based on isotope ratio measurements, which are more precise than ion current measurements and will result in values for the dead time with a measurable uncertainty. The suggested method allows us to estimate the uncertainty associated with the dead time. Typical relative uncertainties of the dead time values were found to be in the range of 10-20%. This method has been applied to monitoring the dead time for a detection system over the lifetime of a particular electron multiplier. The dead time was found to be an important criterion for replacing an old electron multiplier when optimum isotope ratio measurements are required. The dependence of the dead time on the mass of the measured isotopes has been investigated. No mass dependence of the dead time could be concluded within the measurement uncertainty on this particular instrumental set-up.
引用
收藏
页码:1075 / 1079
页数:5
相关论文
共 25 条
[1]   PROPAGATION OF UNCERTAINTIES IN ISOTOPE-DILUTION MASS-SPECTROMETRY USING PULSE COUNTING DETECTION [J].
ADRIAENS, AG ;
KELLY, WR ;
ADAMS, FC .
ANALYTICAL CHEMISTRY, 1993, 65 (05) :660-663
[2]   DETERMINATION OF FISSION-PRODUCTS AND ACTINIDES IN SPENT NUCLEAR-FUELS BY ISOTOPE-DILUTION ION CHROMATOGRAPHY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
ALONSO, JIG ;
SENA, F ;
ARBORE, P ;
BETTI, M ;
KOCH, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (05) :381-393
[3]   Characterisation and correction of instrumental bias in inductively coupled plasma quadrupole mass spectrometry for accurate measurement of lead isotope ratios [J].
Begley, IS ;
Sharp, BL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (04) :395-402
[4]  
Bevington P., 1969, DATA REDUCTION ERROR, P98
[5]  
DOERFFEL K, 1990, STAT ANAL CHEM, P159
[6]  
*EURACHEM WORK GRO, 1995, QUANT UNC AN MEAS
[7]   Measurements of dead time and characterization of ion counting systems for mass spectrometry [J].
Fahey, AJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (03) :1282-1288
[8]  
*GAL EL CORP, 1991, CHANN EL MULT HDB MA
[9]   EFFECTIVE DEADTIME OF PULSE-COUNTING DETECTOR SYSTEMS [J].
HAYES, JM ;
MATTHEWS, DE ;
SCHOELLER, DA .
ANALYTICAL CHEMISTRY, 1978, 50 (01) :25-32
[10]   HIGH PRECISION PULSE COUNTING - LIMITATIONS AND OPTIMAL CONDITIONS [J].
HAYES, JM ;
SCHOELLER, DA .
ANALYTICAL CHEMISTRY, 1977, 49 (02) :306-311