Collimator Design for a Brain SPECT/MRI Insert

被引:14
作者
Salvado, Debora [1 ]
Erlandsson, Kjell [1 ]
Bousse, Alexandre [1 ]
Occhipinti, Michele [2 ]
Busca, Paolo [2 ]
Fiorini, Carlo [2 ]
Hutton, Brian F. [1 ,3 ]
机构
[1] UCL, Inst Nucl Med, London W1E 6BT, England
[2] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, I-20133 Milan, Italy
[3] Univ Wollongong, Ctr Med Radiat Phys, Sydney, NSW, Australia
关键词
Collimator design; multi-modality; pinhole; slit-slat; SPECT insert; stationary system; SLIT-SLAT; EFFECTIVE DIAMETER; SMALL ANIMALS; RESOLUTION; SYSTEM; SIMULATION; PERFORMANCE; IMAGER;
D O I
10.1109/TNS.2015.2450017
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging, with a high-sensitivity collimator and high-resolution detectors. We have compared eight collimator designs, four multi-pinhole and four multi-slit slit-slat configurations. The collimation was designed for a system with 2 rings of 25 5 x 5 cm detectors. We introduce the concept of 1/2-pinhole and 1/2-slit, which are transaxially shared between two adjacent detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a range of intrinsic detector resolutions R-i and target resolutions R-t at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, 0.8 mm R-i and 10 mm R-t FWHM, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and transaxial resolution, while the latter gives better axial resolution. Although the 2 x 2-pin-hole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilisation of the detector FOV. The best options are therefore the 5+2 1/2-pinhole and the 1+2 1/2-slit systems, with sensitivities of 1.8 x 10(-4) and 3.2 x 10(-4), respectively. Noiseless brain phantom reconstructions with the multi-pinhole collimator are slightly superior as compared to slit-slat, in terms of symmetry and accuracy of the activity distribution, but the same is not true when noise is included. DOI information reduces artefacts and improves uniformity in geometric phantoms. Further evaluation is needed with prototype collimators.
引用
收藏
页码:1716 / 1724
页数:9
相关论文
共 50 条
  • [41] Simulations of a Multipinhole SPECT Collimator for Clinical Dopamine Transporter (DAT) Imaging
    Konik, Arda
    De Beenhouwer, Jan
    Mukherjee, Joyeeta M.
    Kalluri, Kesava
    Banerjee, Soumyanil
    Zeraatkar, Navid
    Fromme, Timothy
    King, Michael A.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2018, 2 (05) : 444 - 451
  • [42] Design and evaluation of two multi-pinhole collimators for brain SPECT
    Chen, Ling
    Tsui, Benjamin M. W.
    Mok, Greta S. P.
    ANNALS OF NUCLEAR MEDICINE, 2017, 31 (08) : 636 - 648
  • [43] A Vector Uniform Cramer-Rao Bound for SPECT System Design
    Meng, L. J.
    Li, Nan
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3329 - 3339
  • [44] A Monte Carlo study on the performance evaluation of a parallel hole collimator for a HiReSPECT: A dedicated small-animal SPECT
    Abbaspour, Samira
    Tanha, Kaveh
    Mahmoudian, Babak
    Assadi, Majid
    Islamian, Jalil Pirayesh
    APPLIED RADIATION AND ISOTOPES, 2018, 139 : 53 - 60
  • [45] Development and Initial Results of a Brain PET Insert for Simultaneous 7-Tesla PET/MRI Using an FPGA-Only Signal Digitization Method
    Won, Jun Yeon
    Park, Haewook
    Lee, Seungeun
    Son, Jeong-Whan
    Chung, Yina
    Ko, Guen Bae
    Kim, Kyeong Yun
    Song, Junghyun
    Seo, Seongho
    Ryu, Yeunchul
    Chung, Jun-Young
    Lee, Jae Sung
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2021, 40 (06) : 1579 - 1590
  • [46] Slit-Slat Collimator Equipped Gamma Camera for Whole-Mouse SPECT-CT Imaging
    Cao, Liji
    Peter, Joerg
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (03) : 530 - 536
  • [47] Optimization of a depth of interaction encoding PET block detector fora PET/MRI insert
    Selfridge, Aaron R.
    Cherry, Simon R.
    Judenhofer, Martin S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (23)
  • [48] MRI compatibility study of a prototype radiofrequency penetrable oval PET insert at 3 T
    Akram, Md Shahadat Hossain
    Nishikido, Fumihiko
    Levin, Craig S.
    Takyu, Sodai
    Obata, Takayuki
    Yamaya, Taiga
    JAPANESE JOURNAL OF RADIOLOGY, 2023, 42 (4) : 382 - 390
  • [49] Segmented slant hole collimator for stationary cardiac SPECT: Monte Carlo simulations
    Mao, Yanfei
    Yu, Zhicong
    Zeng, Gengsheng L.
    MEDICAL PHYSICS, 2015, 42 (09) : 5426 - 5434
  • [50] Evaluation of pinhole collimator materials for micron-resolution ex vivo SPECT
    Minh Phuong Nguyen
    Goorden, Marlies C.
    Kamphuis, Chris
    Beekman, Freek J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (10)