摘要
This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging, with a high-sensitivity collimator and high-resolution detectors. We have compared eight collimator designs, four multi-pinhole and four multi-slit slit-slat configurations. The collimation was designed for a system with 2 rings of 25 5 x 5 cm detectors. We introduce the concept of 1/2-pinhole and 1/2-slit, which are transaxially shared between two adjacent detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a range of intrinsic detector resolutions R-i and target resolutions R-t at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, 0.8 mm R-i and 10 mm R-t FWHM, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and transaxial resolution, while the latter gives better axial resolution. Although the 2 x 2-pin-hole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilisation of the detector FOV. The best options are therefore the 5+2 1/2-pinhole and the 1+2 1/2-slit systems, with sensitivities of 1.8 x 10(-4) and 3.2 x 10(-4), respectively. Noiseless brain phantom reconstructions with the multi-pinhole collimator are slightly superior as compared to slit-slat, in terms of symmetry and accuracy of the activity distribution, but the same is not true when noise is included. DOI information reduces artefacts and improves uniformity in geometric phantoms. Further evaluation is needed with prototype collimators.
引用
收藏
页码:1716 / 1724
页数:9
相关论文
共 50 条
  • [31] Development of a Ring PET Insert for MRI
    Bauer, Chris
    Stolin, Alexander
    Proffitt, James
    Martone, Peter
    Brefczynski-Lewis, Julie
    Lewis, James
    Hankiewicz, Janusz
    Raylman, Ray
    Majewski, Stan
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [32] Multi-pinhole collimator design for small-object imaging with SiliSPECT: a high-resolution SPECT
    Shokouhi, S.
    Metzler, S. D.
    Wilson, D. W.
    Peterson, T. E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (02) : 207 - 225
  • [33] A Skew-Slit Collimator for Small-Animal SPECT
    Zeng, Gengsheng L.
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2008, 36 (04) : 207 - 212
  • [34] Evaluation of Stationary and Semi-stationary Acquisitions from Dual-head Multi-pinhole Collimator for Myocardial Perfusion SPECT
    Yan, Pengyu
    Chen, Ling
    Tsui, Benjamin M. W.
    Mok, Greta S. P.
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2016, 36 (05) : 675 - 685
  • [35] Conceptual design of a large pixelated CZT detector with four-hole collimator matched pixel detector for SPECT imaging: a Monte Carlo simulation study
    Boutaghane, N.
    Bouzid, B.
    Zaidi, H.
    JOURNAL OF INSTRUMENTATION, 2019, 14 (02)
  • [36] Optimization and Calibration of Slat Position for a SPECT With Slit-Slat Collimator and Pixelated Detector Crystals
    Deng, Xiao
    Ma, Tianyu
    Lecomte, Roger
    Yao, Rutao
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (05) : 2234 - 2243
  • [37] Development of a Collimator Representation in the TITAN Transport Code for SPECT Simulation
    Royston, Katherine Keller
    Haghighat, Alireza
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2647 - 2650
  • [38] High-sensitivity cardiac SPECT system design with collimator-less interspaced mosaic-patterned scintillators
    Wang, Rui
    Zhang, Debin
    Hu, Yifan
    Lyu, Zhenlei
    Ma, Tianyu
    FRONTIERS IN MEDICINE, 2023, 10
  • [39] FlexiSPECT: A SPECT System Consisting of a Compact High-Resolution Scintillation Detector (SPECTatress) and a Lofthole Collimator
    Deprez, Karel
    Vandenberghe, Stefaan
    Vandeghinste, Bert
    Van Holen, Roel
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (01) : 53 - 64
  • [40] A Monte Carlo Simulation Study of Optimization for Collimator in a Pixelated SPECT Camera
    Telikani, Zeinab
    Sadremomtaz, Alireza
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2019, 50 (01) : 163 - 170