Verification of high dose rate brachytherapy dose distributions with EBT3 Gafchromic film quality control techniques

被引:44
作者
Palmer, Antony L. [1 ,2 ]
Nisbet, Andrew [1 ,3 ]
Bradley, David [1 ]
机构
[1] Univ Surrey, Fac Engn & Phys Sci, Dept Phys, Surrey GU2 7JP, England
[2] Portsmouth Hosp NHS Trust, Queen Alexandra Hosp, Dept Med Phys, Portsmouth PO6 3LY, Hants, England
[3] Royal Surrey Cty Hosp NHS Fdn Trust, Dept Med Phys, Guildford GU2 7XX, Surrey, England
关键词
ENERGY-DEPENDENCE; DOSIMETRY; PHOTON; CODE; AAPM; CALIBRATION;
D O I
10.1088/0031-9155/58/3/497
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is essential that quality control (QC) techniques are developed to keep pace with modern high dose rate (HDR) brachytherapy. Current QC methods may be insufficient to fully assure the accuracy of 3D-optimized dose delivery. This work presents an evaluation of Gafchromic EBT3 film, with multi-channel analysis, in HDR dose environments for advanced QC and commissioning. 'Film-array in water' and 'three-channel Solid Water block' purpose-designed phantoms are utilized. Dose and dose-rate dependency and practical film usage has been evaluated. EBT3 measurements of dose with radial distance from a HDR source are compared to Monte Carlo data. Semi-3D dose distributions around clinical HDR applicators are compared to treatment plans. The measurement of delivery accuracy for inverse-planned pseudo-clinical test cases, with correct delivery and simulated treatment errors, has also been investigated. Local gamma criteria of 3%, 3 mm is recommended with passing rates of at least 96% typically achieved. The system is sensitive to simulated errors in HDR delivery, with significant reductions of passing rate. It has been demonstrated that EBT3 Gafchromic film, in combination with multi-channel analysis, is appropriate for applicator, treatment unit and planning system commissioning measurements as well as practical routine QC to confirm agreement of planned and delivered complex HDR brachytherapy dose distributions.
引用
收藏
页码:497 / 511
页数:15
相关论文
共 34 条
[1]   Radiochromic film dosimetry of HDR 192Ir source radiation fields [J].
Aldelaijan, Saad ;
Mohammed, Huriyyah ;
Tomic, Nada ;
Liang, Li-Heng ;
DeBlois, Francois ;
Sarfehnia, Arman ;
Abdel-Rahman, Wamied ;
Seuntjens, Jan ;
Devic, Slobodan .
MEDICAL PHYSICS, 2011, 38 (11) :6074-6083
[2]   Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies [J].
Arjomandy, Bijan ;
Tailor, Ramesh ;
Anand, Aman ;
Sahoo, Narayan ;
Gillin, Michael ;
Prado, Karl ;
Vicic, Milos .
MEDICAL PHYSICS, 2010, 37 (05) :1942-1947
[3]   Statistical variability and confidence intervals for planar dose QA pass rates [J].
Bailey, Daniel W. ;
Nelms, Benjamin E. ;
Attwood, Kristopher ;
Kumaraswamy, Lalith ;
Podgorsak, Matthew B. .
MEDICAL PHYSICS, 2011, 38 (11) :6053-6064
[4]   The IPEM code of practice for determination of the reference air kerma rate for HDR 192Ir brachytherapy sources based on the NPL air kerma standard [J].
Bidmead, A. M. ;
Sander, T. ;
Locks, S. M. ;
Lee, C. D. ;
Aird, E. G. A. ;
Nutbrown, R. F. ;
Flynn, A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (11) :3145-3159
[5]  
Bonnet J, 2008, BRACHYTHERAPY, V7, P174
[6]   Energy dependence of radiochromic dosimetry films for use in radiotherapy verification [J].
Chelminski, Krzysztof ;
Bulski, Wojciech ;
Georg, Dietmar ;
Bodzak, Dominika ;
Maniakowski, Zbigniew ;
Oborska, Dominika ;
Rostkowska, Joanna ;
Kania, Malgorzata .
REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2010, 15 (02) :40-46
[7]  
Chiu-Tsao S-T, 2006, BRACHYTHERAPY, V5, P97
[8]  
Daigh J D, 2010, 0902815143 DEP VET A
[9]  
Detorie NA, 2010, ACR TECHNICAL STANDA
[10]   Technical note:: Dosimetric study of a. new Co-60 source used in brachytherapy [J].
Granero, D. ;
Perez-Calatayud, J. ;
Ballester, F. .
MEDICAL PHYSICS, 2007, 34 (09) :3485-3488