The reliability of computed tomography (CT) values and dimensional measurements of the oropharyngeal region using cone beam CT: comparison with multidetector CT

被引:89
|
作者
Yamashina, A. [1 ]
Tanimoto, K. [1 ]
Sutthiprapaporn, P. [1 ,2 ]
Hayakawa, Y. [3 ]
机构
[1] Hiroshima Univ, Grad Sch Biomed Sci, Dept Oral & Maxillofacial Surg, Hiroshima 730, Japan
[2] Khon Kaen Univ, Fac Dent, Dept Oral Diag, Khon Kaen, Thailand
[3] Kitami Inst Technol, Fac Engn, Dept Comp Sci, Kitami, Hokkaido, Japan
关键词
computed tomography; cone beam; computed tomography value; oropharynx;
D O I
10.1259/dmfr/45926904
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To evaluate the reliability of cone beam CT (CBCT) values and dimensional measurements of oropharyngeal air spaces as compared with those of multidetector row CT (MDCT). Methods: A soft tissue equivalent phantom with different sized holes was used. The phantom was scanned using CBCT and MDCT. In addition, a volunteer was examined using both methods for clinical comparison. The CT data were retrieved to a personal computer and treated using image processing software for three-dimensional rendering and dimensional measurements. The CT values of air, water and soft tissues were measured experimentally and the CT values of air, fat and muscle were measured in a clinical case. The diameter of each hole drilled in the phantom was measured on CBCT and MDCT images using the software tool and digimatic callipers. Results: The CBCT values were quite different from the Hounsfield units obtained with a typical MDCT system. The standard deviations were almost ten times larger with CBCT. In vivo assessment showed that the CBCT values for fat had a wide range that partially overlapped the values for muscle. The measurement of distances greater than 4 mm was consistent for all methods. The difference for holes was less than 0.2 mm. Conclusions: The phantom study showed that CBCT provides limited quantitative CT values for each pixel on sliced images for differentiating air, water and soft tissues. However, the measurement of air spaces with CBCT was quite accurate.
引用
收藏
页码:245 / 251
页数:7
相关论文
共 50 条
  • [11] Comparison between cone-beam CT and multidetector-row CT by ROC analysis regarding diagnostic accuracy for artificial alveolar bone defects in the mandibular molar region
    Minami, Seiji
    Ohnishi, Takashi
    Sano, Tomoaki
    Sugiura, Kazutaka
    Nakayama, Eiji
    ORAL RADIOLOGY, 2015, 31 (02) : 97 - 104
  • [12] Pulmonary Artery Imaging in Patients with Chronic Thromboembolic Pulmonary Hypertension: Comparison of Cone-Beam CT and 64-Row. Multidetector CT
    Hinrichs, Jan B.
    von Falck, Christian
    Hoeper, Marius M.
    Olsson, Karen M.
    Wacker, Frank K.
    Meyer, Bernhard C.
    Renne, Julius
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2016, 27 (03) : 361 - 368
  • [13] Comparison of linear and volumetric measurements obtained from periodontal defects by using cone beam-CT and micro-CT: an in vitro study
    Tayman, M. Ayse
    Kamburoglu, Kivanc
    Kucuk, Ozlem
    Ates, Funda S. O.
    Gunhan, Meral
    CLINICAL ORAL INVESTIGATIONS, 2019, 23 (05) : 2235 - 2244
  • [14] Comparing rib cortical thickness measurements from computed tomography (CT) and Micro-CT
    Hostetler, Zachary S.
    Stitzel, Joel D.
    Weaver, Ashley A.
    COMPUTERS IN BIOLOGY AND MEDICINE, 2019, 111
  • [15] Quality of Temporal Bone CT Images: A Comparison of Flat Panel Cone Beam CT and Multi-slice CT
    Komori, Masahiro
    Yanagihara, Naoaki
    Hinohira, Yasuyuki
    Kashiba, Keiko
    Miuchi, Shinya
    Kishida, Yoshiomi
    JOURNAL OF INTERNATIONAL ADVANCED OTOLOGY, 2012, 8 (01): : 57 - 62
  • [16] Conventional Multi-Slice Computed Tomography (CT) and Cone-Beam CT (CBCT) for Computer-Aided Implant Placement. Part II: Reliability of Mucosa-Supported Stereolithographic Guides
    Arisan, Volkan
    Karabuda, Zihni Cuneyt
    Piskin, Buelent
    Ozdemir, Tayfun
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2013, 15 (06) : 907 - 917
  • [17] Comparison of Supervised and Unsupervised Approaches for the Generation of Synthetic CT from Cone-Beam CT
    Rossi, Matteo
    Cerveri, Pietro
    DIAGNOSTICS, 2021, 11 (08)
  • [18] Cone-Beam Computed Tomography (CBCT) Versus CT in Lung Ablation Procedure: Which is Faster?
    Cazzato, Roberto Luigi
    Battistuzzi, Jean-Benoit
    Catena, Vittorio
    Grasso, Rosario Francesco
    Zobel, Bruno Beomonte
    Schena, Emiliano
    Buy, Xavier
    Palussiere, Jean
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2015, 38 (05) : 1231 - 1236
  • [19] Cone-Beam Computed Tomography (CBCT) Versus CT in Lung Ablation Procedure: Which is Faster?
    Roberto Luigi Cazzato
    Jean-Benoit Battistuzzi
    Vittorio Catena
    Rosario Francesco Grasso
    Bruno Beomonte Zobel
    Emiliano Schena
    Xavier Buy
    Jean Palussiere
    CardioVascular and Interventional Radiology, 2015, 38 : 1231 - 1236
  • [20] Asymmetry assessment using cone beam CT A Class I and Class II patient comparison
    Sievers, Matthew M.
    Larson, Brent E.
    Gaillard, Philippe R.
    Wey, Andrew
    ANGLE ORTHODONTIST, 2012, 82 (03) : 410 - 417