Hand-Held Bone Cutting Tool With Autonomous Penetration Detection for Spinal Surgery

被引:35
作者
Osa, Takayuki [1 ]
Abawi, Christian Farid [2 ]
Sugita, Naohiko [1 ]
Chikuda, Hirotaka [3 ]
Sugita, Shurei [3 ]
Tanaka, Takeyuki [3 ]
Oshima, Hirofumi [3 ]
Moro, Toru [3 ]
Tanaka, Sakae [3 ]
Mitsuishi, Mamoru [1 ]
机构
[1] Univ Tokyo, Dept Mech Engn, Tokyo 1138656, Japan
[2] Fraunhofer Inst Mfg Engn & Automat, D-70569 Stuttgart, Germany
[3] Univ Tokyo, Grad Sch Med, Tokyo 1138654, Japan
关键词
Bone cutting tool; penetration detection; spinal surgery; INTERVENTIONAL ROBOTICS; TECHNOLOGY; SYSTEM;
D O I
10.1109/TMECH.2015.2410287
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In spinal surgery, a surgeon often needs to remove some parts of the spine to relieve the pressure on the spinal cord or other nerves. In this procedure, the surgeon needs to cut and drill some holes in the spine. This operation is very risky because there are some nerves beneath the target bones, and this procedure therefore requires a skilled and experienced surgeon. However, if the cutting tool could detect penetration of the bone autonomously, the safety of the procedure would be improved drastically. This study presents a hand-held bone cutting tool system that detects the penetration of the workpiece. The system learns the cutting states and motion states from demonstrations by a surgeon, and it autonomously detects the penetration of the workpiece and stops the actuation of the cutting tool immediately before total penetration. The proposed scheme for penetration detection does not require knowledge of the shape and the position of the workpiece and, therefore, it does not require any costly systems, such as robotic arms and position sensor systems. In addition, the proposed scheme can be easily applied to various shapes of the cutting tool, such as drills and saws. The developed system was evaluated through experiments. The results showed that the performance of the developed system was satisfactory in both a motorized and a hand-held setup.
引用
收藏
页码:3018 / 3027
页数:10
相关论文
共 24 条
  • [1] Study on a mechatronic tool for drilling in the osteosynthesis of long bones: Tool/bone interaction, modeling and experiments
    Allotta, B
    Belmonte, F
    Bosio, L
    Dario, P
    [J]. MECHATRONICS, 1996, 6 (04) : 447 - 459
  • [2] [Anonymous], 2008, LearningOpenCV: ComputervisionwiththeOpenCVlibrary
  • [3] Bennett K.P., 1992, Optimization methods and software, V1, P23, DOI DOI 10.1080/10556789208805504
  • [4] Estimation of Soft Tissue Mechanical Parameters From Robotic Manipulation Data
    Boonvisut, Pasu
    Cavusoglu, M. Cenk
    [J]. IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2013, 18 (05) : 1602 - 1611
  • [5] Brisson G, 2004, LECT NOTES COMPUT SC, V3217, P105
  • [6] CORTES C, 1995, MACH LEARN, V20, P273, DOI 10.1023/A:1022627411411
  • [7] Dioiaiti N., 2004, 2004 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (IEEE Cat. No.04CH37566), P2538
  • [8] Surgical and interventional robotics: Part II - Surgical CAD-CAM systems
    Fichtinger, Gabor
    Kazanzides, Peter
    Okamura, Allison M.
    Hager, Gregory D.
    Whitcomb, Louis L.
    Taylor, Russell H.
    [J]. IEEE ROBOTICS & AUTOMATION MAGAZINE, 2008, 15 (03) : 94 - 102
  • [9] Real-Time Identification of Hunt-Crossley Dynamic Models of Contact Environments
    Haddadi, Amir
    Hashtrudi-Zaad, Keyvan
    [J]. IEEE TRANSACTIONS ON ROBOTICS, 2012, 28 (03) : 555 - 566
  • [10] Hara M, 2009, IEEE INT CONF ROBOT, P4238