Characterization of 3-Dimensional Printing and Casting Materials for use in Magnetic Resonance Imaging Phantoms at 3 T

被引:6
作者
Yunker, B. E. [1 ,2 ]
Stupic, K. F. [1 ]
Wagner, J. L. [2 ]
Huddle, S. [2 ]
Shandas, R. [2 ]
Weir, R. F. [2 ]
Russek, S. E. [1 ]
Keenan, K. E. [1 ]
机构
[1] NIST, Phys Measurement Lab, Boulder, CO 80305 USA
[2] Univ Colorado Denver Anschutz, Aurora, CO 80045 USA
关键词
3D printing; medical imaging; MRI; phantom; polymer; RELAXATION-TIMES; BREAST-TISSUE; IN-VIVO; MRI; T-1; MODELS; SEQUENCES; RESECTION;
D O I
10.6028/jres.125.028
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Imaging phantoms are used to calibrate and validate the performance of magnetic resonance imaging (MRI) systems. Many new materials have been developed for additive manufacturing (three-dimensional [3D] printing) processes that may be useful in the direct printing or casting of dimensionally accurate, anatomically accurate, patient-specific, and/or biomimetic MRI phantoms. The T-1, T-2, and T-2* spin relaxation times of polymer samples were tested to discover materials for use as tissue mimics and structures in MRI phantoms. This study included a cohort of polymer compounds that was tested in cured form. The cohort consisted of 101 standardized polymer samples fabricated from: two-part silicones and polyurethanes used in commercial casting processes; one-part optically cured polyurethanes used in 3D printing; and fused deposition thermoplastics used in 3D printing. The testing was performed at 3 T using inversion recovery, spin echo, and gradient echo sequences for T-1, T-2, and T-2*, respectively. T-1, T-2, and T-2* values were plotted with error bars to allow the reader to assess how well a polymer matches a tissue for a specific application. A correlation was performed between T-1, T-2, T-2* values and material density, elongation, tensile strength, and hardness. Two silicones, SI_XP-643 and SI_P-45, may be usable mimics for reported liver values; one silicone, SI_XP-643, may be a useful mimic for muscle; one silicone, SI_XP-738, may be a useful mimic for white matter; and four silicones, SI_P-15, SI_GI-1000, SI_GI-1040, and SI_GI-1110, may be usable mimics for spinal cord. Elongation correlated to T-2 (p = 0.0007), tensile strength correlated to T-1 (p = 0.002), T-2 (p = 0.0003), and T-2* (p = 0.003). The 80 samples not providing measurable signal with T-1, T-2, T-2* relaxation values too short to measure with the standard sequences, may be useful for MRI-invisible fixturing and medical devices at 3 T.
引用
收藏
页数:24
相关论文
共 43 条
  • [1] Use of three-dimensional models to assist in the resection of malignant cardiac tumors
    Al Jabbari, Odeaa
    Abu Saleh, Walid K.
    Patel, Avni P.
    Igo, Stephen R.
    Reardon, Michael J.
    [J]. JOURNAL OF CARDIAC SURGERY, 2016, 31 (09) : 581 - 583
  • [2] Quantitative Ultrashort Echo Time (UTE) MRI of Human Cortical Bone: Correlation With Porosity and Biomechanical Properties
    Bae, Won C.
    Chen, Peter C.
    Chung, Christine B.
    Masuda, Koichi
    D'Lima, Darryl
    Du, Jiang
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 2012, 27 (04) : 848 - 857
  • [3] A Robust Methodology for In Vivo T1 Mapping
    Barral, Joelle K.
    Gudmundson, Erik
    Stikov, Nikola
    Etezadi-Amoli, Maryam
    Stoica, Petre
    Nishimura, Dwight G.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (04) : 1057 - 1067
  • [4] Rapid prototyping compliant arterial phantoms for in-vitro studies and device testing
    Biglino, Giovanni
    Verschueren, Peter
    Zegels, Raf
    Taylor, Andrew M.
    Schievano, Silvia
    [J]. JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2013, 15
  • [5] What are normal relaxation times of tissues at 3 T?
    Bojorquez, Jorge Zavala
    Bricq, Stephanie
    Acquitter, Clement
    Brunotte, Francois
    Walker, Paul M.
    Lalande, Alain
    [J]. MAGNETIC RESONANCE IMAGING, 2017, 35 : 69 - 80
  • [6] Bushberg J, 2012, ESSENTIAL PHYS MED I, DOI [10.2147/OPTH.S123640, DOI 10.2147/OPTH.S123640]
  • [7] MR Fingerprinting for Rapid Quantitative Abdominal Imaging
    Chen, Yong
    Jiang, Yun
    Pahwa, Shivani
    Ma, Dan
    Lu, Lan
    Twieg, Michael D.
    Wright, Katherine L.
    Seiberlich, Nicole
    Griswold, Mark A.
    Gulani, Vikas
    [J]. RADIOLOGY, 2016, 279 (01) : 278 - 286
  • [8] 3D Myocardial T1 Mapping at 3T Using Variable Flip Angle Method: Pilot Study
    Clique, Helene
    Cheng, Hai-Ling Margaret
    Marie, Pierre-Yves
    Felblinger, Jacques
    Beaumont, Marine
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (02) : 823 - 829
  • [9] MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: Preliminary results
    de Bazelaire, CMJ
    Duhamel, GD
    Rofsky, NM
    Alsop, DC
    [J]. RADIOLOGY, 2004, 230 (03) : 652 - 659
  • [10] Effect of physiological Heart Rate variability on quantitative T2 measurement with ECG-gated Fast Spin Echo (FSE) sequence and its retrospective correction
    de Roquefeuil, Marion
    Vuissoz, Pierre-Andre
    Escanye, Jean-Marie
    Felblinger, Jacques
    [J]. MAGNETIC RESONANCE IMAGING, 2013, 31 (09) : 1559 - 1566