A validated, stability-indicating HPLC method for the determination of dexamethasone related substances on dexamethasone-coated drug-eluting stents

被引:56
作者
Chen, Quan [1 ]
Zielinsk, Dorota [1 ]
Chen, Jenny [1 ]
Koski, Andrew [1 ]
Werst, David [1 ]
Nowak, Steve [1 ]
机构
[1] Abbott Labs, Global Analyt Res & Dev, Abbott Pk, IL 60064 USA
关键词
Dexamethasone; Degradation products; HPLC; LC/MS; Impurity identification; Drug-eluting stents;
D O I
10.1016/j.jpba.2008.07.010
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An HPLC method was developed and validated to determine trace amounts of dexamethasone related substances on dexamethasone-coated drug-eluting stents. Separation of dexamethasone from its major process impurities and degradation products was achieved on a Zorbax Eclipse XDB C8 column using gradient elution and UV detection at 239 nm. The method was validated according to ICH guideline requirements. In addition, stent extraction efficiency, solution stability and method robustness were evaluated. The method was determined to be linear in the range of 0.01-0.30 mu g ml(-1) for the quantitation of major dexamethasone related substances. Method accuracy was assessed by spiking dexamethasone acetate at three levels over the range of 0.025-0.175 mu g ml(-1). The dexamethasone acetate recovery ranged from 89.6 to 105.8%. The intermediate precision over the three levels was less than 6% (n = 9). The method was also shown to be repeatable at concentration levels of 0.025, 0.125 and 0.175 mu g ml(-1) dexamethasone with relative standard deviation values of 4.1, 1.7 and 1.6%, respectively. The method was found to be specific, stability-indicating, and sensitive with a detection limit of 0.008 mu g ml(-1) and a quantitation limit of 0.025 mu g ml(-1) dexamethasone. Finally, the method was demonstrated to be robust, resistant to small variations of chromatographic variables such as pH, mobile phase organic/aqueous composition and column temperature. Identifying unknown dexamethasone degradation products in dexamethasone-coated drug-eluting stents was of critical interest to ensure product quality, since degradants have a significant impact on safety, efficacy, and product storage and handling. The developed chromatographic method was designed to be compatible with mass spectrometric detection. This paper also discusses using this chromatographic method coupled to an ion-trap LCQ mass spectrometer to elucidate proposed structures for four major dexamethasone degradants. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:732 / 738
页数:7
相关论文
共 18 条
  • [1] [Anonymous], JAPANESE PHARMACOPOE
  • [2] [Anonymous], 2017, The European Pharmacopoeia (Ph Eur)
  • [3] Analysis of betamethasone, dexamethasone and related compounds by liquid chromatography/electrospray mass spectrometry
    Arthur, KE
    Wolff, JC
    Carrier, DJ
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (06) : 678 - 684
  • [4] Determination of dexamethasone and two excipients (creatinine and propylparaben) in injections by using UV-spectroscopy and multivariate calibrations
    Collado, MS
    Robles, JC
    De Zan, M
    Cámara, MS
    Mantovani, VE
    Goicoechea, HC
    [J]. INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2001, 229 (1-2) : 205 - 211
  • [5] Gentile DM, 1996, J PHARMACOL EXP THER, V277, P105
  • [6] Goodman-Hilman A, 1996, PHARM BASIS THERAPEU
  • [7] STUDIES ON THE STABILITY OF CORTICOSTEROIDS .5. THE DEGRADATION PATTERN OF HYDROCORTISONE IN AQUEOUS-SOLUTION
    HANSEN, J
    BUNDGAARD, H
    [J]. INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1980, 6 (3-4) : 307 - 319
  • [8] Evaluation of a high-dose Dexamethasone-eluting stent
    Hoffmann, R
    Langenberg, R
    Radke, P
    Franke, A
    Blindt, R
    Ortlepp, J
    Popma, JJ
    Weber, C
    Hanrath, P
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2004, 94 (02) : 193 - 195
  • [9] *ICH TECHN REQ REG, VAL AN PROC TEXT MET
  • [10] MECHANISM OF PORTER-SILBER REACTION .I. REARRANGEMENT OF DIHYDROXYACETONE GROUP OF STEROIDS
    LEWBART, ML
    MATTOX, VR
    [J]. JOURNAL OF ORGANIC CHEMISTRY, 1964, 29 (03) : 513 - &