Timing resolution improvement using DOI information in a four-layer scintillation detector for TOF-PET

被引:32
|
作者
Shibuya, Kengo [1 ]
Nishikido, Fumihiko [1 ]
Tsuda, Tomoaki [2 ]
Kobayashi, Tetsuya [3 ]
Lam, Chihfung [1 ]
Yamaya, Taiga [1 ]
Yoshida, Eiji [1 ]
Inadama, Naoko [1 ]
Murayama, Hideo [1 ]
机构
[1] Natl Inst Radiol Sci, Mol Imaging Ctr, JPET Project Team, Inage Ku, Chiba 2630024, Japan
[2] Shimadzu Co Ltd, Technol Res Lab, Seika, Kyoto 6190237, Japan
[3] Chiba Univ, Grad Sch Engn, Dept Med Syst Engn, Inage Ku, Chiba 2638522, Japan
关键词
positron emission tomography (PET); time-of-flight (TOF); depth-of-interaction (DOI); digital sampling; timing resolution; constant fraction discrimination (CFD); TOF shift correction;
D O I
10.1016/j.nima.2008.05.020
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Depth-of-interaction (DOI) detectors are considered to be advantageous for time-of-flight positron emission tomography (TOF-PET) because they can correct timing errors arising in the scintillation crystals due to a propagation speed difference between annihilation radiation and scintillation photons. We experimentally measured this timing error, using our four-layer DOI encoding method. The upper layers exhibited the larger timing delays due to the longer path lengths after conversion from annihilation radiation into scintillation photons that traveled by zigzag paths at a speed decreased by a factor of the refractive index (n). The maximum timing delay between the uppermost and the lowermost layers was evaluated as 164 ps when n = 1.47. A TOF error correction was demonstrated to improve the timing resolution of the four-layer DOI detector by 10.3%, which would increase the effective sensitivity of the scanner by about 12% comparison with a non-DOI TOF-PET scanner. This is the first step towards combining these two important fields in PET instrumentation, namely DOI and TOF, for the purpose of achieving a higher sensitivity as well as a more uniform spatial resolution. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:572 / 577
页数:6
相关论文
共 50 条
  • [21] Improvement of crystal identification performance for a four-layer DOI detector composed of crystals segmented by laser processing
    Mohammadi, Akram
    Inadama, Naoko
    Yoshida, Eiji
    Nishikido, Fumihiko
    Shimizu, Keiji
    Yamaya, Taiga
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 866 : 29 - 35
  • [22] Optimizing the timing resolution of SiPM sensors for use in TOF-PET detectors
    Vinke, R.
    Lohner, H.
    Schaart, D. R.
    van Dam, H. T.
    Seifert, S.
    Beekman, F. J.
    Dendooven, P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 610 (01): : 188 - 191
  • [23] Development of a 4-Layer DOI TOF-PET detector module with a 6 mm-pitch MPPC array
    Takyu, Sodai
    Yoshida, Eiji
    Nishikido, Fumihiko
    Nitta, Munetaka
    Akamatsu, Go
    Yamaya, Taiga
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [24] A CNN-based four-layer DOI encoding detector using LYSO and BGO scintillators for small animal PET imaging
    He, Wen
    Zhao, Yangyang
    Zhao, Xin
    Huang, Wenjie
    Zhang, Lei
    Prout, David L.
    Chatziioannou, Arion F.
    Ren, Qiushi
    Gu, Zheng
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (09):
  • [25] Energy Estimation Technique Utilizing Timing Information for TOF-PET Application
    Mandai, Shingo
    Venialgo, Esteban
    Charbon, Edoardo
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [26] Energy spectra and their calibration of four-layer DOI detector for brain PET scanner: jPET-D4
    Yoshida, E
    Kitamura, K
    Tsuda, T
    Inadama, N
    Murayama, H
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4172 - 4176
  • [27] Performance evaluation of a four-layer LSO detector for a small animal DOI PET scanner: jPET-RD
    Tsuda, T
    Murayama, H
    Kitamura, K
    Inadama, N
    Yamaya, T
    Yoshida, E
    Nishikido, F
    Hamamoto, M
    Kawai, H
    Ono, Y
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3321 - 3324
  • [28] Energy spectra analysis of the four-layer DOI detector for the brain PET scanner: jPET-D4
    Yoshida, E
    Kitamura, K
    Tsuda, T
    Shibuya, K
    Yamaya, T
    Inadama, N
    Hasegawa, T
    Murayama, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 557 (02): : 664 - 669
  • [29] Computer simulation of four-layer depth of interaction detector for PET
    Sato, M
    Haneishi, H
    Inadama, N
    Murayama, H
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1913 - 1917
  • [30] Performance evaluation of a subset of a four-layer LSO detector for a small animal DOI PET scanner: jPET-RD
    Tsuda, T
    Murayama, H
    Kitamura, K
    Inadama, N
    Yamaya, T
    Yoshida, E
    Nishikido, F
    Hamamoto, M
    Kawai, H
    Ono, Y
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 35 - 39