Balanced detection for spectral domain optical coherence tomography

被引:39
作者
Kuo, Wen-Chuan [1 ,2 ]
Lai, Chih-Ming [3 ]
Huang, Yi-Shiang [1 ]
Chang, Cheng-Yi [4 ]
Kuo, Yue-Ming [1 ]
机构
[1] Natl Yang Ming Univ, Inst Biophoton, Taipei 112, Taiwan
[2] Natl Yang Ming Univ, BMIRC, Taipei 112, Taiwan
[3] Ming Chuan Univ, Dept Elect Engn, Tao Yuan 333, Taiwan
[4] Natl Taiwan Normal Univ, Inst Electroopt Sci & Technol, Taipei 116, Taiwan
关键词
NOISE; CAPILLAROSCOPY; REFLECTOMETRY; SENSITIVITY;
D O I
10.1364/OE.21.019280
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The use and advantages of applying balanced-detection (BD) operation method to high speed spectral domain optical coherence tomography (SDOCT) are presented in this study, which we believe is the first such demonstration. Compared to conventional SDOCT, BD-SDOCT provides two unique advantages. First, the method can suppress background noise and autocorrelation artifacts in biological tissues. Second, it is a power-efficient method which is particularly helpful for high speed SDOCT to eliminate random intensity noise and to achieve shot noise limited detection. This performance allows in vivo three-dimensional tissue visualization with high imaging quality and high speed. (c) 2013 Optical Society of America
引用
收藏
页码:19280 / 19291
页数:12
相关论文
共 27 条
[1]   A DUAL-DETECTOR OPTICAL HETERODYNE RECEIVER FOR LOCAL OSCILLATOR NOISE SUPPRESSION [J].
ABBAS, GL ;
CHAN, VWS ;
YEE, TK .
JOURNAL OF LIGHTWAVE TECHNOLOGY, 1985, 3 (05) :1110-1122
[2]   Continuum generation in a novel photonic crystal fiber for ultrahigh resolution optical coherence tomography at 800 nm and 1300 nm [J].
Aguirre, AD ;
Nishizawa, N ;
Fujimoto, JG ;
Seitz, W ;
Lederer, M ;
Kopf, D .
OPTICS EXPRESS, 2006, 14 (03) :1145-1160
[3]   Self-phase-modulated Kerr-lens mode-locked Cr:forsterite laser source for optical coherence tomography [J].
Bouma, BE ;
Tearney, GJ ;
Bilinsky, IP ;
Golubovic, B ;
Fujimoto, JG .
OPTICS LETTERS, 1996, 21 (22) :1839-1841
[4]   Optical coherence tomography using a frequency-tunable optical source [J].
Chinn, SR ;
Swanson, EA ;
Fujimoto, JG .
OPTICS LETTERS, 1997, 22 (05) :340-342
[5]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[6]   Simultaneous dual-band optical coherence tomography in the spectral domain for high resolution in vivo imaging [J].
Cimalla, Peter ;
Walther, Julia ;
Mehner, Mirko ;
Cuevas, Maximiliano ;
Koch, Edmund .
OPTICS EXPRESS, 2009, 17 (22) :19486-19500
[7]   Nailfold capillaroscopy is useful for the diagnosis and follow-up of autoimmune rheumatic diseases. A future tool for the analysis of microvascular heart involvement? [J].
Cutolo, M. ;
Sulli, A. ;
Secchi, M. E. ;
Paolino, S. ;
Pizzorni, C. .
RHEUMATOLOGY, 2006, 45 :43-46
[8]   In vivo ultrahigh-resolution optical coherence tomography [J].
Drexler, W ;
Morgner, U ;
Kärtner, FX ;
Pitris, C ;
Boppart, SA ;
Li, XD ;
Ippen, EP ;
Fujimoto, JG .
OPTICS LETTERS, 1999, 24 (17) :1221-1223
[9]   Ultrahigh-resolution optical coherence tomography [J].
Drexler, W .
JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (01) :47-74
[10]   Role of beat noise in limiting the sensitivity of optical coherence tomography [J].
Haskell, Richard C. ;
Liao, David ;
Pivonka, Adam E. ;
Bell, Tera L. ;
Haberle, Brendan R. ;
Hoeling, Barbara M. ;
Petersen, Daniel C. .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2006, 23 (11) :2747-2755