Frequency selectivity of contralateral residual acoustic hearing in bimodal cochlear implant users, and limitations on the ability to match the pitch of electric and acoustic stimuli

被引:17
作者
Green, Tim [1 ]
Faulkner, Andrew [1 ]
Rosen, Stuart [1 ]
机构
[1] UCL, London WC1N 1PF, England
关键词
Cochlear implant; bimodal hearing; frequency selectivity; SPEECH RECOGNITION; FUNDAMENTAL-FREQUENCY; OPPOSITE EARS; PERCEPTION; BENEFIT; ADULTS; PLACE; NOISE; AIDS; ADAPTATION;
D O I
10.3109/14992027.2011.642010
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: To assess the reliability of across-ear, acoustic-electric pitch/timbre comparisons for determining effective characteristic frequencies of cochlear implant electrodes. Study sample: Nine CI users with contralateral residual acoustic hearing. Design: Absolute acoustic thresholds in the unimplanted ear were measured and frequency selectivity was assessed via psychophysical tuning curves. An adjustment method was used to match the percepts elicited by pulse trains on individual electrodes with various acoustic signals (pure tones, narrow-band noises, and bandpass filtered pulse trains). The starting frequency of the acoustic signal was roved and matches were obtained at different loudness levels. Results: Acoustic frequency selectivity varied widely. Two subjects showed clear evidence of frequency selectivity extending above 500 Hz. Only these subjects produced consistent pitch matches over repeated measurements. For other subjects, the acoustic frequency eventually selected tended to correlate with the initially presented frequency. There was limited evidence of level effects and these were inconsistent across subjects and electrodes. Conclusions: Across-modality pitch/timbre matching appears unlikely to provide a generally applicable method for determining the effective characteristic frequencies of cochlear implant electrodes. Frequency selectivity above 500 Hz may be necessary for consistent pitch/timbre matches.
引用
收藏
页码:389 / 398
页数:10
相关论文
共 50 条
  • [41] Electric to acoustic pitch matching: a possible way to improve individual cochlear implant fitting
    Di Nardo, Walter
    Cantore, Italo
    Marchese, Maria Raffaella
    Cianfrone, Francesca
    Scorpecci, Alessandro
    Giannantonio, Sara
    Paludetti, Gaetano
    [J]. EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2008, 265 (11) : 1321 - 1328
  • [42] Electric to acoustic pitch matching: a possible way to improve individual cochlear implant fitting
    Walter Di Nardo
    Italo Cantore
    Maria Raffaella Marchese
    Francesca Cianfrone
    Alessandro Scorpecci
    Sara Giannantonio
    Gaetano Paludetti
    [J]. European Archives of Oto-Rhino-Laryngology, 2008, 265 : 1321 - 1328
  • [43] Impact of Reverberation on Speech Perception in Noise in Bimodal/Bilateral Cochlear Implant Users with and without Residual Hearing
    Koenig, Clara
    Baumann, Uwe
    Stoever, Timo
    Weissgerber, Tobias
    [J]. JOURNAL OF CLINICAL MEDICINE, 2024, 13 (17)
  • [44] Natural history of contralateral residual hearing in unilateral cochlear implant users - long-term findings
    Yehudai, Noam
    Shpak, Talma
    Most, Tova
    Luntz, Michal
    [J]. ACTA OTO-LARYNGOLOGICA, 2012, 132 (10) : 1073 - 1076
  • [45] Cortical Processing of Frequency Changes Reflected by the Acoustic Change Complex in Adult Cochlear Implant Users
    Liang, Chun
    Houston, Lisa M.
    Samy, Ravi N.
    Abedelrehim, Lamiaa Mohamed Ibrahim
    Zhang, Fawen
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2018, 23 (03) : 152 - 164
  • [46] Speech perception with electric-acoustic stimulation. Comparison with bilateral cochlear implant users in different noise conditions
    Rader, T.
    [J]. HNO, 2015, 63 (02) : 85 - 93
  • [47] The Acoustic Change Complex Compared to Hearing Performance in Unilaterally and Bilaterally Deaf Cochlear Implant Users
    van Heteren, Jan A. A.
    Vonck, Bernard M. D.
    Stokroos, Robert J.
    Versnel, Huib
    Lammers, Marc J. W.
    [J]. EAR AND HEARING, 2022, 43 (06) : 1783 - 1799
  • [48] Towards decoding selective attention through cochlear implant electrodes as sensors in subjects with contralateral acoustic hearing
    Aldag, Nina
    Buechner, Andreas
    Lenarz, Thomas
    Nogueira, Waldo
    [J]. JOURNAL OF NEURAL ENGINEERING, 2022, 19 (01)
  • [49] Pitch Matching in Cochlear Implant Users With Single-Sided Deafness: Effects of Electrode Position and Acoustic Stimulus Type
    Adel, Youssef
    Nagel, Sharon
    Weissgerber, Tobias
    Baumann, Uwe
    Macherey, Olivier
    [J]. FRONTIERS IN NEUROSCIENCE, 2019, 13
  • [50] Hearing Asymmetry Biases Spatial Hearing in Bimodal Cochlear-Implant Users Despite Bilateral Low-Frequency Hearing Preservation
    Sharma, Snandan
    Mens, Lucas H. M.
    Snik, Ad F. M.
    van Opstal, A. John
    van Wanrooij, Marc M. M.
    [J]. TRENDS IN HEARING, 2023, 27