共 21 条
Development and in-house validation of a sensitive LC-MS/MS method for simultaneous quantification of gelsemine, koumine and humantenmine in porcine plasma
被引:18
作者:
Yang, Kun
[1
]
Long, Xue-Ming
[2
]
Liu, Yan-Chun
[1
]
Chen, Fu-Hua
[2
]
Liu, Xiao-Feng
[1
]
Sun, Zhi-Liang
[1
,3
]
Liu, Zhao-Ying
[1
,3
,4
]
机构:
[1] Hunan Agr Univ, Coll Vet Med, Hunan Engn Res Ctr Vet Drug, Changsha 410128, Hunan, Peoples R China
[2] Inst Hunan Prov Vet Drugs & Feed Supervis, Changsha 410006, Hunan, Peoples R China
[3] Hunan Agr Univ, Coll Vet Med, Hunan Coinnovat Ctr Utilizat Bot Funct Ingredient, Changsha 410128, Hunan, Peoples R China
[4] Hunan Agr Univ, Natl & Local Union Engn Res Ctr Vet Herbal Med Re, Changsha 410128, Hunan, Peoples R China
来源:
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
|
2018年
/
1076卷
基金:
国家重点研发计划;
关键词:
Indole alkaloid;
Gelsemine;
Gelsemium elegans;
Plasma;
Koumine;
Humantenmine;
TANDEM MASS-SPECTROMETRY;
TIME-OF-FLIGHT;
RAT PLASMA;
CHROMATOGRAPHY;
ALKALOIDS;
QUADRUPOLE;
COMPONENTS;
ELEGANS;
D O I:
10.1016/j.jchromb.2018.01.019
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
Three monomers of G. elegans indole alkaloids (gelsemine, koumine and humantenmine) were simultaneously detected in porcine plasma for the first time with the development and validation of a sensitive and reliable LC-ESI-MS/MS method. Using a gradient mobile phase at a constant flow rate of 0.2 mL/min via electrospray ionization (positive ion mode) in a multiple reaction monitoring (MRM) scan, gelsemine, koumine and humantenmine were eluted, separated and detected at an appropriate retention time. The porcine plasma was prepared using protein precipitation with 1% formic acid-acetonitrile: methanol (2:1, v/v). Using matrix-matched calibration curves and weighted least squares linear regression, a good linearity (r(2) > 0.99) was achieved with a concentration range of 0.1-200 mu g/L for gelsemine, koumine and humantenmine; estimated LOD and LOQ values were 0.10 mu g/L and 0.2 mu g/L, respectively. The mean of the recoveries was in the range of 82.68-400.35% of porcine plasma at four different levels, and the intra-day and inter-day precision (CV) were lower than 15% with a range of 2.46-8.76% and 2.73-10.83%, respectively. The proposed method has proved to be suitable for accurate, quantitative determination of gelsemine, koumine and humantenmine in porcine plasma.
引用
收藏
页码:54 / 60
页数:7
相关论文