Temperature changes during cortical bone drilling with a newly designed step drill and an internally cooled drill

被引:133
作者
Augustin, Goran [1 ,2 ]
Davila, Slavko [1 ,2 ]
Udilljak, Toma [3 ]
Staroveski, Tomislav [3 ]
Brezak, Danko [3 ,5 ]
Babic, Slaven [4 ]
机构
[1] Univ Hosp Ctr Zagreb, Dept Surg, Zagreb 10000, Croatia
[2] Univ Zagreb, Sch Med, Zagreb 10000, Croatia
[3] Univ Zagreb, Fac Mech Engn & Naval Architecture, Dept Technol, Zagreb 10000, Croatia
[4] Clin Hosp Ctr Sestre Milosrdnice, Dept Traumatol, Zagreb 10000, Croatia
[5] Univ Zagreb, Fac Mech Engn & Naval Architecture, Dept Robot & Prod Syst Automat, Zagreb 10000, Croatia
关键词
IN-VITRO; HEAT-PRODUCTION; IRRIGATION; PARAMETERS; IMPLANTS; TISSUE;
D O I
10.1007/s00264-012-1491-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Bone drilling causes an increase in bone temperature, and a temperature above 47A degrees C is critical because it causes thermal bone necrosis. Thermal osteonecrosis is common with the drill diameter of a parts per thousand yen4.5 mm without cooling. The aim of this study was to determine the increase of bone temperature during drilling using newly contructed two-step and internally cooled drills. An experiment was set up according to a central composite design. An internally cooled drill (3.4 mm and 4.5 mm) and a two-step drill (2.5/3.4 and 3.4/4.5 mm) were used in combination with feed rates of (0.02, 0.04, 0.10, 0.16 and 0.18 mm/rev) and cutting speeds (1.18, 10.68, 33.61, 56.55 and 66.05 m/min) with and without cooling with water of 24A degrees C. Bone temperatures were measured with thermocouples. Drilling was performed on pig diaphyses with a three-axis mini milling machine. Bone temperatures in all combinations of parameters with internal cooling were below the critical 47A degrees C (p = 0.05). The highest temperatures were detected using a 4.5-mm drill (40.5A degrees C). A statistically significant effect other than cooling was found with the drill diameter and feed. A drill diameter of 3.4 mm with internal cooling developed a maximum temperature of 38.5A degrees C and without cooling 46.3A degrees C. For the same conditions a drill with diameter of 4.5 mm reached temperatures of 40.5A degrees C and 55.7A degrees C, respectively. The effect of feed rate is inversely proportional to the increase in bone temperature. With the feed rate 0.16 mm/rev, temperature was below critical even using the 4.5-mm drill (46.4A degrees C, p = 0.05). Using the 3.4-mm drill all temperatures were below critical (46.2A degrees C, p = 0.05). The two-step drill compared to a standard drill with the same diameter did not show statistical differences in maximum bone temperatures for all combinations of parameters (p = 0.05). A two-step drill does not have any advantages over a standard twist drill of the same diameter. An internally cooled drill causes a significantly smaller increase of bone temperature during drilling with water of 24A degrees C. An internally cooled drill is currently the 'ideal' drill for traumatology/orthopaedics because it produces the smallest increase in bone drilling temperature. If internal cooling is used the regulation of other drilling parameters is of no importance.
引用
收藏
页码:1449 / 1456
页数:8
相关论文
共 20 条
  • [1] Interspecies differences in bone composition, density, and quality:: Potential implications for in vivo bone research
    Aerssens, J
    Boonen, S
    Lowet, G
    Dequeker, J
    [J]. ENDOCRINOLOGY, 1998, 139 (02) : 663 - 670
  • [2] Augustin G, 2011, CLIN BIOMECH BR 1107
  • [3] Thermal osteonecrosis and bone drilling parameters revisited
    Augustin, Goran
    Davila, Slavko
    Mihoci, Kristijan
    Udiljak, Toma
    Vedrina, Denis Stjepan
    Antabak, Anko
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2008, 128 (01) : 71 - 77
  • [4] The effects of drilling force on cortical temperatures and their duration: an in vitro study
    Bachus, KN
    Rondina, MT
    Hutchinson, DT
    [J]. MEDICAL ENGINEERING & PHYSICS, 2000, 22 (10) : 685 - 691
  • [5] Thermal changes observed at implant sites during internal and external irrigation
    Benington, IC
    Biagioni, PA
    Briggs, J
    Sheridan, S
    Lamey, PJ
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2002, 13 (03) : 293 - 297
  • [6] In vitro comparison of cortical bone temperature generation between traditional sequential drilling and a newly designed step drill in the equine third metacarpal bone
    Bubeck, K. A.
    Garcia-Lopez, J.
    Maranda, L. S.
    [J]. VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2009, 22 (06) : 442 - 447
  • [7] Heat production by 3 implant drill systems after repeated drilling and sterilization
    Chacon, GE
    Bower, DL
    Larsen, PE
    McGlumphy, EA
    Beck, FM
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 64 (02) : 265 - 269
  • [8] Ercoli C, 2004, INT J ORAL MAX IMPL, V19, P335
  • [9] TEMPERATURE THRESHOLD LEVELS FOR HEAT-INDUCED BONE TISSUE-INJURY - A VITAL-MICROSCOPIC STUDY IN THE RABBIT
    ERIKSSON, AR
    ALBREKTSSON, T
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1983, 50 (01) : 101 - 107
  • [10] Eriksson R.A., 1984, ACTA ORTHOP SCAND, V55, P629