Comparison of the flaps made by femtosecond laser and automated keratomes for sub-bowman keratomileusis

被引:12
作者
Zhai Chang-bin [1 ,2 ]
Tian Lei [1 ,2 ]
Zhou Yue-hua [1 ,2 ]
Zhang Qing-wei [1 ,2 ]
Zhang Jing [1 ,2 ]
机构
[1] Capital Med Univ, Beijing Tongren Hosp, Ophthalm Ctr, Beijing 100730, Peoples R China
[2] Beijing Ophthalmol & Visual Sci Key Lab, Beijing 100730, Peoples R China
关键词
sub-bowman; keratomileusis; femtosecond laser; microkeratome; flap thickness; IN-SITU KERATOMILEUSIS; OPTICAL COHERENCE TOMOGRAPHY; STROMAL THICKNESS; CONTRALATERAL EYE; LASIK FLAP; KERATECTOMY; ABLATION;
D O I
10.3760/cma.j.issn.0366-6999.20130918
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Thin-flap laser in situ keratomileusis (LASIK) is the new trend of refractive error correction surgery, the formation of corneal flap is crucial for a success of LASIK surgery. This study aimed to assess and compare the variations of LASIK flap created by the IntraLase femtosecond laser, Mona One Use-Plus SBK and Mode M2 Single-Use 90 mu m-head microkeratome using Anterior segment optical coherence tomography (Visante OCT). Methods One hundred and sixty-one eyes of 81 consecutive patients were enrolled in this prospective study and randomly divided into three groups depending on the flap creation method: flap creation with the the IntraLase femtosecond laser (IntraLase group, 59 eyes), flap creation with the Mona One Use-Plus SBK (SBK group, 44 eyes), and flap creation with the Mona M2 Single-Use 90 pm-head microkeratome (M2SU90 group, 58 eyes). The nominal flap thickness was 110 mu m for all patients and for the three devices. One month after surgery, Visante OCT was used to measure the flap thickness at 20 locations on each cornea and the results were assessed for uniformity, regularity, and accuracy. Results At 1 month after surgery, the mean central flap thickness was (111 +/- 3) mu m in the IntraLase group, (114 +/- 8) mu m in the SBK group, and (118 +/- 13) mu m in the M2SU90 group respectively. The flaps in the IntraLase group and the SBK group were more regular, showing an almost planar configuration, than the meniscus-shaped flaps in the M2SU90 group. The maximum deviation from the intended flap thickness (110 mu m) was 6 mu m in the IntraLase group, 10 mu m in the SBK group, and 20 mu m in the M2SU90 group respectively. A difference greater than 20 mu m was observed in 0.42% of measurements in the IntraLase group; 2.95% of the measurements in the SBK group and 21.12% of measurements in the M2SU90 group. Conclusions The flaps created by the IntraLase femtosecond laser and Mona One Use-Plus SBK are more uniform; more regular, and more accurate than those created by the Mona M2 Single-Use 90 mu m-head microkeratome. The first two methods can make precise flaps for Sub-Bowman Keratomileusis.
引用
收藏
页码:2440 / 2444
页数:5
相关论文
共 20 条
[1]   Very high-frequency digital ultrasound measurement of the LASIK flap thickness profile using the intralase femtosecond laser and M2 and carriazo-pendular microkeratomes [J].
Alio, Jorge L. ;
Pinero, David P. .
JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (01) :12-23
[2]   Thin flap laser in situ keratomileusis:: Analysis of contrast sensitivity, visual, and refractive outcomes [J].
Cobo-Soriano, R ;
Calvo, MA ;
Beltrán, J ;
Llovet, FL ;
Baviera, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (07) :1357-1365
[3]   Wavefront-guided excimer laser ablation using photorefractive keratectomy and sub-bowman's keratomileusis: A contralateral eye study [J].
Durrie, Daniel S. ;
Slade, Stephen G. ;
Marshall, John .
JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (01) :77-84
[4]   Femtosecond laser versus mechanical keratome flaps in wavefront-guided laser in situ keratomileusis - Prospective contralateral eye study [J].
Durrie, DS ;
Kezirian, GA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :120-126
[5]   Technique for measuring laser in situ keratomileusis flap thickness using the IntraLase laser [J].
Eisner, RA ;
Binder, PS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (04) :556-558
[6]   The effect of flap thickness on the visual and refractive outcome of myopic laser in situ keratomileusis [J].
Eleftheriadis, H ;
Prandi, B ;
Diaz-Rato, A ;
Morcillo, M ;
Sabater, JB .
EYE, 2005, 19 (12) :1290-1296
[7]  
Flanagan GW, 2003, J REFRACT SURG, V19, P113
[8]   Real-time optical coherence tomography-guided femtosecond laser sub-Bowman keratomileusis on human donor eyes [J].
Kermani, Omid ;
Fabian, Will ;
Lubatschowski, Holger .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 146 (01) :42-45
[9]   Flap thickness reproducibility in laser in situ keratomileusis with a femtosecond laser: Optical coherence tomography measurement [J].
Kim, Jin-Hyoung ;
Lee, Doh ;
Rhee, Kang-II .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (01) :132-136
[10]  
Lazaro C, 2013, ARCH SOC ESP OFTALMO, V88, P45