Validation of novel calibration scheme with traceable point-like 22Na sources on six types of PET scanners

被引:8
作者
Hasegawa, Tomoyuki [1 ]
Oda, Keiichi [2 ]
Wada, Yasuhiro [3 ]
Sasaki, Toshiaki [4 ]
Sato, Yasushi [5 ]
Yamada, Takahiro [6 ]
Matsumoto, Mikio [6 ]
Murayama, Hideo [7 ]
Kikuchi, Kei [8 ]
Miyatake, Hiroki [8 ]
Abe, Yutaka [8 ]
Miwa, Kenta [9 ]
Akimoto, Kenta [9 ]
Wagatsuma, Kei [9 ]
机构
[1] Kitasato Univ, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
[2] Tokyo Metropolitan Inst Gerontol, Positron Med Ctr, Itabashi Ku, Tokyo 1730015, Japan
[3] RIKEN Ctr Mol Imaging Sci, Chuo Ku, Kobe, Hyogo 6500047, Japan
[4] Iwate Med Univ, Cyclotron Res Ctr, Takizawa, Iwate 0200173, Japan
[5] Adv Ind Sci & Technol, Tsukuba, Ibaraki 3058568, Japan
[6] Japan Radioisotope Assoc JRIA, Bunkyo Ku, Tokyo 1138941, Japan
[7] Natl Inst Radiol Sci, Mol Imaging Ctr, Inage Ku, Chiba 2638555, Japan
[8] Kitasato Univ Hosp, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
[9] Japanese Fdn Canc Res, Canc Inst Hosp, Koto Ku, Tokyo 1358550, Japan
关键词
Calibration; Point-like source; Traceability; Quality control; POSITRON; PERFORMANCE; VARIABILITY;
D O I
10.1007/s12149-013-0692-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To improve the reliability and convenience of the calibration procedure of positron emission tomography (PET) scanners, we have been developing a novel calibration path based on traceable point-like sources. When using Na-22 sources, special care should be taken to avoid the effects of 1.275-MeV gamma rays accompanying beta (+) decays. The purpose of this study is to validate this new calibration scheme with traceable point-like Na-22 sources on various types of PET scanners. Traceable point-like Na-22 sources with a spherical absorber design that assures uniform angular distribution of the emitted annihilation photons were used. The tested PET scanners included a clinical whole-body PET scanner, four types of clinical PET/CT scanners from different manufacturers, and a small-animal PET scanner. The region of interest (ROI) diameter dependence of ROI values was represented with a fitting function, which was assumed to consist of a recovery part due to spatial resolution and a quadratic background part originating from the scattered gamma rays. The observed ROI radius dependence was well represented with the assumed fitting function (R (2) > 0.994). The calibration factors determined using the point-like sources were consistent with those by the standard cross-calibration method within an uncertainty of +/- 4 %, which was reasonable considering the uncertainty in the standard cross-calibration method. This novel calibration scheme based on the use of traceable Na-22 point-like sources was successfully validated for six types of commercial PET scanners.
引用
收藏
页码:346 / 354
页数:9
相关论文
共 41 条
  • [1] GEANT4-a simulation toolkit
    Agostinelli, S
    Allison, J
    Amako, K
    Apostolakis, J
    Araujo, H
    Arce, P
    Asai, M
    Axen, D
    Banerjee, S
    Barrand, G
    Behner, F
    Bellagamba, L
    Boudreau, J
    Broglia, L
    Brunengo, A
    Burkhardt, H
    Chauvie, S
    Chuma, J
    Chytracek, R
    Cooperman, G
    Cosmo, G
    Degtyarenko, P
    Dell'Acqua, A
    Depaola, G
    Dietrich, D
    Enami, R
    Feliciello, A
    Ferguson, C
    Fesefeldt, H
    Folger, G
    Foppiano, F
    Forti, A
    Garelli, S
    Giani, S
    Giannitrapani, R
    Gibin, D
    Cadenas, JJG
    González, I
    Abril, GG
    Greeniaus, G
    Greiner, W
    Grichine, V
    Grossheim, A
    Guatelli, S
    Gumplinger, P
    Hamatsu, R
    Hashimoto, K
    Hasui, H
    Heikkinen, A
    Howard, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) : 250 - 303
  • [2] [Anonymous], 2009, IAEA HUM HLTH SER
  • [3] [Anonymous], 1998, 616751 IEC
  • [4] [Anonymous], 2008, NEMA STAND PUBL
  • [5] Azuelos G., 1976, Atomic Data and Nuclear Data Tables, V17, P103, DOI 10.1016/0092-640X(76)90003-6
  • [6] A METHOD FOR MEASURING THE ABSOLUTE SENSITIVITY OF POSITRON EMISSION TOMOGRAPHIC SCANNERS
    BAILEY, DL
    JONES, T
    SPINKS, TJ
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1991, 18 (06): : 374 - 379
  • [7] Bailey DL, 1997, EUR J NUCL MED, V24, P660
  • [8] Bailey DL, 1998, THEORY PRACTICE 3D P
  • [9] Usefulness of specific calibration coefficients for gamma-emitting sources measured by radionuclide calibrators in nuclear medicine
    Bochud, Francois O.
    Laedermann, Jean-Pascal
    Baechler, Sebastien
    Kosinski, Marek
    Bailat, Claude J.
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 4073 - 4080
  • [10] Standards for PET Image Acquisition and Quantitative Data Analysis
    Boellaard, Ronald
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 : 11S - 20S