Undersensing by an ICD Due to Alternans of the Ventricular Electrogram

被引:4
作者
Van Heuverswyn, Frederic E. [1 ]
Timmers, Liesbeth [1 ]
Stroobandt, Roland X. [1 ]
Barold, S. Serge [2 ]
机构
[1] Univ Hosp Ghent, Ctr Heart, B-9000 Ghent, Belgium
[2] Florida Heart Rhythm Inst, Tampa, FL USA
关键词
implantable cardioverter-defibrillator (ICD); alternans; ventricular tachycardia; ventricular undersensing by ICD; ELECTRICAL ALTERNANS; TACHYCARDIA; COMPLEX;
D O I
10.1111/j.1542-474X.2012.00521.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Alternans of the ventricular electrogram (VEGM) during ventricular tachycardia (VT) is a rare cause of ventricular undersensing by an implantable cardioverter-defibrillator (ICD). This report describes a patient with a St. Jude ICD who exhibited sustained monomorphic VT associated with surface QRS alternans, alternating cycle lengths, alternans of the VEGM causing intermittent undersensing of the smaller component, and intermittent 2:1 counting of ventricular intervals during 1:1 sensing in response to the ICD detection algorithm. VEGM undersensing was corrected noninvasively simply by programming the threshold start from 62.5% to 50% which increased the sensitivity based on the amplitude of the VEGM. This maneuver did not affect the satisfactory and stable defibrillation threshold. Ann Noninvasive Electrocardiol 2013;18(1):84-89
引用
收藏
页码:84 / 89
页数:6
相关论文
共 50 条
[41]   Avoiding inappropriate ventricular tachycardia detection due to T-wave oversensing in an implantable cardioverter defibrillator: a novel application of the electrogram width criterion [J].
Duru, F ;
Bauersfeld, U ;
Candinas, R .
EUROPACE, 2001, 3 (01) :80-84
[42]   Inappropriate ICD discharges due to “triple counting” during normal sinus rhythm [J].
Ejaz Khan ;
Apostolos Voudouris ;
Stephen R. Shorofsky ;
Robert W. Peters .
Journal of Interventional Cardiac Electrophysiology, 2006, 17 :153-155
[43]   Significance of Ventricular Arrhythmia Based on Stored Electrogram Analysis in a Pacemaker Population [J].
Hosoda, Junya ;
Ishikawa, Toshiyuki ;
Matsumoto, Katsumi ;
Kiyokuni, Masayoshi ;
Taguchi, Yuka ;
Narikawa, Masatoshi ;
Hibi, Kiyoshi ;
Sugano, Teruyasu ;
Ishigami, Tomoaki ;
Tamura, Kouichi ;
Kimura, Kazuo .
INTERNATIONAL HEART JOURNAL, 2020, 61 (05) :922-926
[44]   Formation of Spatially Discordant Alternans Due to Fluctuations and Diffusion of Calcium [J].
Sato, Daisuke ;
Bers, Donald M. ;
Shiferaw, Yohannes .
PLOS ONE, 2013, 8 (12)
[45]   Spatially discordant alternans due to periodic pacing site alternation [J].
Bak, Tymoteusz ;
Sato, Daisuke .
HEART RHYTHM, 2025, 22 (02) :365-374
[46]   Inappropriate ICD discharges due to "triple counting" during normal sinus rhythm [J].
Khan, Ejaz ;
Voudouris, Apostolos ;
Shorofsky, Stephen R. ;
Peters, Robert W. .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2006, 17 (02) :153-155
[47]   Two Ventricular Tachycardias with Cycle Length and QRS Alternans: Insights into the Mechanism from Mapping and Ablation of the Tachycardias [J].
Kalbfleisch, Steven ;
Hart, David ;
Weiss, Raul .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (12) :e31-e35
[48]   A Wavelet-Based Electrogram Onset Delineator for Automatic Ventricular Activation Mapping [J].
Alcaine, Alejandro ;
Soto-Iglesias, David ;
Calvo, Mireia ;
Guiu, Esther ;
Andreu, David ;
Fernandez-Armenta, Juan ;
Berruezo, Antonio ;
Laguna, Pablo ;
Camara, Oscar ;
Pablo Martinez, Juan .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (12) :2830-2839
[49]   Heart Rate-Dependent Hysteresis of T-Wave Alternans in Primary Prevention ICD Patients [J].
Burattini, Laura ;
Man, Sumche ;
Fioretti, Sandro ;
Di Nardo, Francesco ;
Swenne, Cees A. .
ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2016, 21 (05) :460-469
[50]   Digoxin and Risk of Ventricular Tachyarrhythmia and Death in ICD Recipients [J].
Ojo, Amole ;
Mcnitt, Scott ;
Polonsky, Bronislava ;
Aktas, Mehmet K. ;
Rosero, Spencer ;
Hall, Burr ;
Kutyifa, Valentina ;
Rao, Nilesh ;
Rao, Nikhila ;
Goldenberg, Ilan .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2024, 10 (07) :1468-1476