Undersensing by an ICD Due to Alternans of the Ventricular Electrogram

被引:4
|
作者
Van Heuverswyn, Frederic E. [1 ]
Timmers, Liesbeth [1 ]
Stroobandt, Roland X. [1 ]
Barold, S. Serge [2 ]
机构
[1] Univ Hosp Ghent, Ctr Heart, B-9000 Ghent, Belgium
[2] Florida Heart Rhythm Inst, Tampa, FL USA
关键词
implantable cardioverter-defibrillator (ICD); alternans; ventricular tachycardia; ventricular undersensing by ICD; ELECTRICAL ALTERNANS; TACHYCARDIA; COMPLEX;
D O I
10.1111/j.1542-474X.2012.00521.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Alternans of the ventricular electrogram (VEGM) during ventricular tachycardia (VT) is a rare cause of ventricular undersensing by an implantable cardioverter-defibrillator (ICD). This report describes a patient with a St. Jude ICD who exhibited sustained monomorphic VT associated with surface QRS alternans, alternating cycle lengths, alternans of the VEGM causing intermittent undersensing of the smaller component, and intermittent 2:1 counting of ventricular intervals during 1:1 sensing in response to the ICD detection algorithm. VEGM undersensing was corrected noninvasively simply by programming the threshold start from 62.5% to 50% which increased the sensitivity based on the amplitude of the VEGM. This maneuver did not affect the satisfactory and stable defibrillation threshold. Ann Noninvasive Electrocardiol 2013;18(1):84-89
引用
收藏
页码:84 / 89
页数:6
相关论文
共 50 条
  • [1] Alternans of the Ventricular Electrogram in Patients with an Implanted Cardioverter-Defibrillator
    Barold, S. Serge
    Kucher, Andreas
    de Meester, Antoine
    Stroobandt, Roland X.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2015, 38 (12): : 1470 - 1480
  • [2] Defibillator Electrogram T Wave Alternans as a Predictor of Spontaneous Ventricular Tachyarrhythmias in Defibrillator Recipients
    Kim, Jin Won
    Pak, Hui-Nam
    Park, Jae Hyung
    Nam, Gi-Byoung
    Kim, Sook Kyoung
    Lee, Hyun Soo
    Jang, Jin Kun
    Choi, Jong Il
    Kim, Young-Hoon
    CIRCULATION JOURNAL, 2009, 73 (01) : 55 - 62
  • [3] Alternans-Induced ICD Therapy
    Barold, S. Serge
    Kucher, Andreas
    McEneaney, David J.
    Stroobandt, Roland X.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2015, 38 (09): : 1109 - 1113
  • [4] Fatal Undersensing of Ventricular Fibrillation due to Intermittent High-Amplitude R Waves
    Anguera, Ignasi
    Sabate, Xavier
    Sugranes, Gemma
    Cequier, Angel
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (09): : e284 - e286
  • [5] Alternans during fascicular ventricular tachycardia due to digitalis toxicity
    Barold, S. Serge
    JOURNAL OF ELECTROCARDIOLOGY, 2018, 51 (03) : 450 - 451
  • [6] QRS alternans during ventricular tachycardia
    Jastrzebski, Marek
    Gomula, Piotr
    Dabrowska, Barbara
    KARDIOLOGIA POLSKA, 2009, 67 (03) : 317 - 321
  • [7] Endocardial Electrogram Characteristics of Epicardial Ventricular Arrhythmias
    Tzou, Wendy S.
    Nguyen, Duy T.
    Aleong, Ryan G.
    Varosy, Paul D.
    Katz, David F.
    Heath, Russell R.
    Schuller, Joseph L.
    Lowery, Christopher M.
    Lewkowiez, Laurent
    Sauer, William H.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (06) : 649 - 654
  • [8] Alternans in QRS amplitude during ventricular tachycardia
    Maury, P
    Metzger, J
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (02): : 142 - 150
  • [9] Undersensing of ventricular tachycardia in a pacemaker patient: What is the mechanism?
    Monkhouse, Christopher
    Whittaker-Axon, Sarah
    Chow, Anthony W. C.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (05): : 553 - 556
  • [10] Discordant Alternans as a Mechanism for Initiation of Ventricular Fibrillation In Vitro
    Munoz, Laura M.
    Gelzer, Anna R. M.
    Fenton, Flavio H.
    Qian, Wei
    Lin, WeiYe
    Gilmour, Robert F., Jr.
    Otani, Niels F.
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2018, 7 (17):