Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies

被引:8
作者
Bazan, Ovandir [1 ]
Ortiz, Jayme Pinto [1 ]
机构
[1] EPUSP, Dept Mech Engn, Av Prof Mello de Moraes 2231, BR-05508030 Sao Paulo, SP, Brazil
关键词
Ventricular Function; Left; Heart valve prosthesis; In Vitro Techniques; Heart Rate; IMAGE-RECONSTRUCTION; FLOW; PHASES; VITRO; CT;
D O I
10.5935/1678-9741.20160036
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. Methods: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves. Results: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition. Conclusion: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.
引用
收藏
页码:171 / 173
页数:3
相关论文
共 14 条
  • [11] International Standard, 2015, 584032013E ISO
  • [12] International Standard, 2015, 584022015E ISO
  • [13] Pulsatile flow studies of a porcine bioprosthetic aortic valve in vitro: PIV measurements and shear-induced blood damage
    Lim, WL
    Chew, YT
    Chew, TC
    Low, HT
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (11) : 1417 - 1427
  • [14] US FDA, 2010, GUID IND FD IN PRESS