Insertion of distal locking screws of tibial intramedullary nails: A comparison between the free-hand technique and the SURESHOT™ Distal Targeting System

被引:29
作者
Moreschini, O. [1 ]
Petrucci, V. [1 ]
Cannata, R. [1 ]
机构
[1] Univ Rome Sapienza, Dept Anat Histol Med Legal Sci Locomotor Syst, Rome, Italy
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2014年 / 45卷 / 02期
关键词
Intramedullary nailing; Distal locking; Distal screw; Tibial fracture; Fluoroscopy; EM guided; Radiation exposure; SURESHOT (TM); RADIATION-EXPOSURE; INTERLOCKING NAILS; NAVIGATION SYSTEM; FRACTURES; DEFORMATION; SURGEON;
D O I
10.1016/j.injury.2013.09.023
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: Positioning of the distal locking screws of an intramedullary nail is often challenging and time consuming because of difficult localisation of the distal locking holes, potential screw malalignment and nail deformation during insertion. The standard free-hand technique under fluoroscopic control involves considerable radiation exposure of both the patient and the surgical team. In this study, we aimed to compare the free-hand technique with a new system that utilises electromagnetic (EM) tracking data (SURESHOT (TM) Distal Targeting System) to localise distal locking holes. Material and methods: Patients admitted from March 2010 to January 2013 for tibial fracture that required intramedullary nailing were analysed retrospectively. We compared intraoperative radiation exposure time and distal locking time in patients treated with the standard free-hand technique and distal locking using the EM field-generating device. Intraoperative radiation exposure time and distal locking time were used for comparison. Results: Data from a total of 50 patients were analysed. The standard free-hand technique and the EM field-generating device were used in 25 (group 1) and 25 (group 2) patients, respectively. Mean distal locking time was 1258.6 (450-2289) s in group 1 and 603.5 (360-1140) s in group 2. Mean radiation exposure time was 19.4 (6-33) s in group 1 and 4.6 (1-10) s in group 2. Conclusion: The EM field-generating device significantly reduces distal locking time and, more importantly, significantly decreases duration of exposure to ionising radiation. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:405 / 407
页数:3
相关论文
共 27 条
[1]  
Brumback RJ, 1996, CLIN ORTHOP RELAT R, P292
[2]   The Insertion of Intramedullary Nail Locking Screws Without Fluoroscopy: A Faster and Safer Technique [J].
Chan, Daniel S. ;
Burris, Richard B. ;
Erdogan, Murat ;
Sagi, H. Claude .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2013, 27 (07) :363-366
[3]  
GREGORY P, 1995, CLIN ORTHOP RELAT R, P48
[4]  
Gugala Z, 2001, Injury, V32 Suppl 4, pSD21
[5]   A SIMPLE JIG TO EASE THE INSERTION OF DISTAL SCREWS IN INTRAMEDULLARY LOCKING NAILS [J].
HASHEMINEJAD, A ;
GARLICK, N ;
GODDARD, NJ .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1994, 25 (06) :407-408
[6]   Dynamisation and early weight-bearing in tibial reamed intramedullary nailing: Its safety and effect on fracture union [J].
Hernandez-Vaquero, Daniel ;
Suarez-Vazquez, Abelardo ;
Iglesias-Fernandez, Susana ;
Garcia-Garcia, Jesus ;
Cervero-Suarez, Javier .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2012, 43 :S63-S67
[7]   Next generation distal locking for intramedullary nails using an electromagnetic X-ray-radiation-free real-time navigation system [J].
Hoffmann, Michael ;
Schroeder, Malte ;
Lehmann, Wolfgang ;
Kammal, Michael ;
Rueger, Johannes Maria ;
Ruecker, Andreas Herrman .
JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2012, 73 (01) :243-248
[8]   External jig-aided intramedullary interlocking nailing of diaphyseal fractures: experience from a tropical developing centre [J].
Ikpeme, Ikpeme ;
Ngim, Ngim ;
Udosen, Anthony ;
Onuba, Obiora ;
Enembe, Okokon ;
Bello, Segun .
INTERNATIONAL ORTHOPAEDICS, 2011, 35 (01) :107-111
[9]   Radiation-free distal locking of intramedullary nails: Evaluation of a new electromagnetic computer-assisted guidance system [J].
Ioannis, Stathopoulos ;
Panagiotis, Karampinas ;
Dimitrios-Stergios, Evangelopoulos ;
Kalliopi, Lampropoulou-Adamidou ;
John, Vlamis .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2013, 44 (06) :872-875
[10]  
Jankovic A, 2013, INJURY, V44, pS33, DOI 10.1016/S0020-1383(13)70195-5