Comparison of the electrochemical behavior of amorphous Zr55Cu30Ni5Al10, stainless steel (316LVM), and CoCrMo (F75) in simulated body fluid with and without addition of protein

被引:3
作者
Tabeshian, Ali [1 ]
Persson, Dan [2 ]
Arnberg, Lars [1 ,3 ]
Aune, Ragnhild [1 ]
机构
[1] Norwegian Univ Sci & Technol, Mat Sci & Engn, Alfred Getz Vei2, N-7491 Trondheim, Trondheim, Norway
[2] Swerea KIMAB AB, Corros & Surface Engn, Ringgold Stand Inst, Box 7047, SE-16407 Kista, Sweden
[3] Kungliga Tekn Hgsk, Skolan Tekn Vetenskap, Ringgold Stand Inst, Mat Sci & Engn, SE-10044 Stockholm, Sweden
来源
MATERIALS AND CORROSION-WERKSTOFFE UND KORROSION | 2019年 / 70卷 / 04期
关键词
CoCrMo; electrochemical measurement; metal ion release; metallic glass; stainless steel; BULK METALLIC GLASSES; PASSIVE FILMS; CORROSION; ALLOYS; HIP;
D O I
10.1002/maco.201810480
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Electrochemical behavior of, and metal ion release from the bulk amorphous (glassy) Zr55Cu30Ni5Al10 alloy (Zr-MG) was evaluated in simulated body fluid (phosphate buffer saline [PBS]), with and without additions of protein (albumin Fraction V) at pH 7.4 and 5.2 and at body temperature 310 K (37 degrees C). The passivation behavior and susceptibility to pitting of the Zr-MG was compared with conventional load bearing implant materials, that is, the medical grade ASTM F75 cast CoCrMo alloy (CoCrMo) and AISI 316 LVM low carbon vacuum re-melted stainless steel alloy (SS). Furthermore, the metal ion release from the main constituent elements of each alloy was measured and compared. All materials showed passive behavior in the PBS solution with and without presence of albumin, though the passive region was smaller for the Zr-MG compared to the CoCrMo and SS. Moreover, all materials experienced pitting corrosion in the PBS solution while the Zr-MG was the most susceptible and the CoCrMo was the least one. Protein additions to the CoCrMo and SS prevented the formation of stable pits at pH 7.4 and 5.2. A decrease in passive region and pitting potential was seen in the case of albumin additions for the Zr-MG at pH 7.4, while the opposite was seen at pH 5.2. Furthermore, the total metal ion release from the Zr-MG was less than for the CoCrMo.
引用
收藏
页码:652 / 660
页数:9
相关论文
共 42 条
[1]  
[Anonymous], 2018, 1099315 EN ISO
[2]   Amorphous Metals for Hard-tissue Prosthesis [J].
Demetriou, Marios D. ;
Wiest, Aaron ;
Hofmann, Douglas C. ;
Johnson, William L. ;
Han, Bo ;
Wolfson, Nikolaj ;
Wang, Gongyao ;
Liaw, Peter K. .
JOM, 2010, 62 (02) :83-91
[3]   Investigations on the electrochemical behaviour of Zr-based bulk metallic glasses [J].
Gebert, A ;
Buchholz, K ;
Leonhard, A ;
Mummert, K ;
Eckert, J ;
Schultz, L .
MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 1999, 267 (02) :294-300
[4]   Ti based biomaterials, the ultimate choice for orthopaedic implants - A review [J].
Geetha, M. ;
Singh, A. K. ;
Asokamani, R. ;
Gogia, A. K. .
PROGRESS IN MATERIALS SCIENCE, 2009, 54 (03) :397-425
[5]   Metallic glasses ... on the threshold [J].
Greer, A. Lindsay .
MATERIALS TODAY, 2009, 12 (1-2) :14-22
[6]  
Hallab N.J., 2004, Biomaterials science: An introduction to materials in medicine, V2nd
[7]  
Hansen DC, 2008, ELECTROCHEM SOC INTE, V17, P31
[8]   Effect of pH on the polarization behavior of Zr65Al7.5Ni10Cu17.5 amorphous alloy in a phosphate-buffered solution [J].
Hiromoto, S ;
Tsai, AP ;
Sumita, M ;
Hanawa, T .
CORROSION SCIENCE, 2000, 42 (12) :2193-2200
[9]   Passive and transpassive behaviour of CoCrMo in simulated biological solutions [J].
Hodgson, AWE ;
Kurz, S ;
Virtanen, S ;
Fervel, V ;
Olsson, COA ;
Mischler, S .
ELECTROCHIMICA ACTA, 2004, 49 (13) :2167-2178
[10]   Biocompatible Ni-free Zr-based bulk metallic glasses with high-Zr-content: Compositional optimization for potential biomedical applications [J].
Hua, Nengbin ;
Huang, Lu ;
Chen, Wenzhe ;
He, Wei ;
Zhang, Tao .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2014, 44 :400-410