Biomechanical evaluation of a pre-contoured clavicle plate

被引:32
作者
Goswami, Tarun [1 ]
Markert, Ronald J. [1 ]
Anderson, Christopher G. [1 ]
Sundaram, Sandhya S. [1 ]
Crosby, Lynn A. [1 ]
机构
[1] Wright State Univ, Boonshoft Sch Med, Dept Orthopaed Surg Sports Med & Rehabil, Dayton, OH 45409 USA
关键词
D O I
10.1016/j.jse.2008.02.017
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Recent attention has been focused on the operative treatment of mid-shaft fractures of the clavicle. This study compares the in-vitro biomechanical properties of a preformed titanium clavicle plate (Acumed) to a Synthes 3.5 mm limited-contact dynamic-compression (LCDC) plate using a cadaveric osteotomy model. An osteotomy was performed on 7 pairs of human clavicles and were randomly plated with either a Synthes 3.5 mm LCDC stainless steel plate or an Acumed titanium pre-contoured clavicle plate. After plating, specimens were tested on on EnduraTEC material testing apparatus for axial compression and tension strength, as well as torsional strength in compression and tension. Biomechanical test results for plated specimens are reported for the LCDC plate and the Acumed plate, and the 2 plates are compared. This exploratory study supports investigations with larger sample sizes to determine if the Acumed pre-contoured plate differs from the LCDC plate in biomechanical properties and the clinical implications of such differences.
引用
收藏
页码:815 / 818
页数:4
相关论文
共 20 条
[1]   NONUNION OF FRACTURES OF THE MIDSHAFT OF THE CLAVICLE - TREATMENT WITH A MODIFIED HAGIE INTRAMEDULLARY PIN AND AUTOGENOUS BONE-GRAFTING [J].
BOEHME, D ;
CURTIS, RJ ;
DEHAAN, JT ;
KAY, SP ;
YOUNG, DC ;
ROCKWOOD, CA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (08) :1219-1226
[2]   Complications of plate fixation of fresh displaced midclavicular fractures [J].
Bostman, O ;
Manninen, M ;
Pihlajamaki, H .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1997, 43 (05) :778-783
[3]   Closed treatment of displaced middle-third fractures of the clavicle gives poor results [J].
Hill, JM ;
McGuire, MH ;
Crosby, LA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (04) :537-539
[4]   Effects of plate location and selection on the stability of midshaft clavicle osteotomies: A biomechanical study [J].
Iannotti, MR ;
Crosby, LA ;
Stafford, P ;
Grayson, G ;
Goulet, R .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (05) :457-462
[5]  
Leupin S, 1998, Swiss Surg, V4, P89
[6]   Deficits following nonoperative treatment of displaced midshaft clavicular fractures [J].
McKee, MD ;
Pedersen, EM ;
Jones, C ;
Stephen, DJG ;
Kreder, HJ ;
Schemitsch, EH ;
Wild, LM ;
Potter, J .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (01) :35-40
[7]   Midshaft malunions of the clavicle [J].
McKee, MD ;
Wild, LM ;
Schemitsch, EH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (05) :790-797
[8]   LOW-CONTACT DYNAMIC COMPRESSION PLATING OF THE CLAVICLE [J].
MULLAJI, AB ;
JUPITER, JB .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1994, 25 (01) :41-45
[9]   NONUNION OF THE CLAVICLE [J].
NEER, CS .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1960, 172 (10) :1006-1011
[10]  
NEVIASER RJ, 1975, CLIN ORTHOP RELAT R, P103