Effect of LVAD Outlet Graft Anastomosis Angle on the Aortic Valve, Wall, and Flow

被引:30
作者
Inci, Gizem [1 ]
Sorguven, Esra [1 ]
机构
[1] Yeditepe Univ, Dept Mech Engn, TR-34755 Istanbul, Turkey
关键词
VENTRICULAR ASSIST DEVICE; BLOOD PUMP; SIMULATION; LOCATION; PATTERNS; CANNULA;
D O I
10.1097/MAT.0b013e3182578b6a
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Left ventricular assist devices (LVADs), which pump blood from the left ventricle to the aorta are an important therapy option for patients with end-stage cardiovascular diseases. Recent publications show that even with optimized LVADs fatal complications can occur because of the blood deformations around the inflow cannula or through the LVAD outlet graft-aorta anastomosis. This study investigates the effects of the anastomosis geometry on the flow through the aorta, on the pressure and wall shear stress (WSS) distributions on the aortic wall and on the total entropy generation in the anastomosis region. Anastomosis geometry is defined with two angles, one on the coronal plane and the other on the transversal plane. Turbulent flow simulations are performed for each geometry. Results indicate that 3% to 5% of the work given by the LVAD is dissipated because of the viscous losses in the anastomosis region. The entropy generation, as well as the maximum WSS, increases as the inclination angle decreases. Some portion of the blood streaming out of the LVAD conduit flows toward the aortic valve; therefore the reverse-flow region extends up to the aortic valve in some cases, which may be one of the causes of aortic-valve dysfunction. Results of this study provide insight on the importance of the anastomosis geometry on the hemodynamics in the aorta and downstream the aortic valve, stresses on the aortic wall, and viscous losses. ASAIO Journal 2012;58:373-381.
引用
收藏
页码:373 / 381
页数:9
相关论文
共 24 条
[1]   Computational flow study of the continuous flow ventricular assist device, prototype number 3 blood pump [J].
Anderson, JB ;
Wood, HG ;
Allaire, PE ;
Bearnson, G ;
Khanwilkar, P .
ARTIFICIAL ORGANS, 2000, 24 (05) :377-385
[2]  
[Anonymous], 2005, FLUENT VERS 6 2 US G
[3]   Variational multiscale residual-based turbulence modeling for large eddy simulation of incompressible flows [J].
Bazilevs, Y. ;
Calo, V. M. ;
Cottrell, J. A. ;
Hughes, T. J. R. ;
Reali, A. ;
Scovazzi, G. .
COMPUTER METHODS IN APPLIED MECHANICS AND ENGINEERING, 2007, 197 (1-4) :173-201
[4]   Patient-specific isogeometric fluid-structure interaction analysis of thoracic aortic blood flow due to implantation of the Jarvik 2000 left ventricular assist device [J].
Bazilevs, Y. ;
Gohean, J. R. ;
Hughes, T. J. R. ;
Moser, R. D. ;
Zhang, Y. .
COMPUTER METHODS IN APPLIED MECHANICS AND ENGINEERING, 2009, 198 (45-46) :3534-3550
[5]  
Cengel YunusA., 2002, THERMODYNAMICS ENG A, V5
[6]   Numerical simulation of an axial blood pump [J].
Chua, Leok Poh ;
Su, Boyang ;
Lim, Tau Meng ;
Zhou, Tongming .
ARTIFICIAL ORGANS, 2007, 31 (07) :560-570
[7]   The Development of Aortic Insufficiency in Left Ventricular Assist Device-Supported Patients [J].
Cowger, Jennifer ;
Pagani, Francis D. ;
Haft, Jonathan W. ;
Romano, Matthew A. ;
Aaronson, Keith D. ;
Kolias, Theodore J. .
CIRCULATION-HEART FAILURE, 2010, 3 (06) :668-674
[8]   APPLICATIONS OF CFD TECHNIQUE IN THE DESIGN AND FLOW ANALYSIS OF IMPLANTABLE AXIAL FLOW BLOOD PUMP [J].
Fan Hui-min ;
Hong Fang-wen ;
Zhang Guo-ping ;
Ye Liang ;
Liu Zhong-min .
JOURNAL OF HYDRODYNAMICS, 2010, 22 (04) :518-525
[9]   The use of computational fluid dynamics in the development of ventricular assist devices [J].
Fraser, Katharine H. ;
Taskin, M. Ertan ;
Griffith, Bartley P. ;
Wu, Zhongjun J. .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (03) :263-280
[10]   Multilaboratory Particle Image Velocimetry Analysis of the FDA Benchmark Nozzle Model to Support Validation of Computational Fluid Dynamics Simulations [J].
Hariharan, Prasanna ;
Giarra, Matthew ;
Reddy, Varun ;
Day, Steven W. ;
Manning, Keefe B. ;
Deutsch, Steven ;
Stewart, Sandy F. C. ;
Myers, Matthew R. ;
Berman, Michael R. ;
Burgreen, Greg W. ;
Paterson, Eric G. ;
Malinauskas, Richard A. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2011, 133 (04)