Clinical outcome of the PAS-Port® proximal anastomosis system in off-pump coronary artery bypass grafting in 201 patients

被引:0
作者
Verberkmoes, N. J. [1 ]
Mokhles, M. M. [2 ]
Bramer, S. [1 ]
Van Straten, A. H. M. [1 ]
Ter Woorst, J. F. [1 ]
Maessen, J. G. [3 ]
Berreklouw, E. [1 ]
机构
[1] Catharina Hosp, Dept Cardiothorac Surg, Eindhoven, Netherlands
[2] Erasmus MC, Dept Cardiothorac Surg, Rotterdam, Netherlands
[3] MUMC, Cardiovasc Res Inst Maastricht CARIM, Dept Cardiothorac Surg, Maastricht, Netherlands
关键词
Outcome Assessment (Health Care); Surgical instruments; Coronary artery bypass; off-pump; RANDOMIZED-TRIAL; FOLLOW-UP; SURGERY; DEVICE; PATENCY; DISEASE;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim. The PAS-Port (R) Proximal Anastomosis System (Cardica, Inc, Redwood City, CA, USA) has been used worldwide since March 2003. The objective of the present study was to evaluate the clinical outcome of the PAS-Port (R) Proximal Anastomosis System. Methods. All the patients who underwent off-pump coronary artery bypass grafting in the Catharina Hospital Eindhoven between August 2006 and April 2010 were included in a non-randomized retrospective case-control study, if they had at least one proximal vein graft anastomosis. Study end-points consisted of overall survival, coronary reintervention and postoperative stroke. Results. The study included 312 patients (201 cases, 111 controls). After 36 months of follow-up there was no difference in survival between cases and controls (92.2% vs. 93.7%, P=0.52). No significant difference could be detected between cases and controls with respect to overall coronary reintervention-free survival (93% vs. 96.4%, P=0.20) and freedom from coronary reintervention due to proximal vein graft failure (98% vs. 100% P=0.14). The use of the PAS-Port system could not be identified as an independent risk factor of coronary reintervention (p=0.21). Postoperative stroke rates of cases and controls (2% vs. 0.9%, P=0.42) were comparable. Conclusion. The clinical outcomes in patients treated with the PAS-Port (R) Proximal Anastomosis System were satisfactory compared with those treated with the conventional hand-sewing technique. The use of the PAS-Port system was not associated with higher adverse outcome in terms of overall survival, stroke, coronary reintervention-free survival and freedom from reintervention due to proximal vein graft failure.
引用
收藏
页码:389 / 395
页数:7
相关论文
共 50 条
  • [31] The future of off-pump coronary artery bypass grafting: a North American perspective
    Yanagawa, Bobby
    Nedadur, Rashmi
    Puskas, John D.
    JOURNAL OF THORACIC DISEASE, 2016, 8 : S832 - S838
  • [32] Off-Pump versus On-Pump Coronary-Artery Bypass Grafting in Elderly Patients
    Diegeler, Anno
    Boergermann, Jochen
    Kappert, Utz
    Breuer, Martin
    Boening, Andreas
    Ursulescu, Adrian
    Rastan, Ardawan
    Holzhey, David
    Treede, Hendrik
    Riess, Friedrich-Christian
    Veeckmann, Philippe
    Asfoor, Amjad
    Reents, Wilko
    Zacher, Michael
    Hilker, Michael
    NEW ENGLAND JOURNAL OF MEDICINE, 2013, 368 (13) : 1189 - 1198
  • [33] Coronary Artery Bypass in Young Patients-On or Off-Pump?
    Okano, Ryoi
    Liou, Yi-Jia
    Yu, Hsi-Yu
    Wu, I-Hui
    Chou, Nai-Kuan
    Chen, Yih-Sharng
    Chi, Nai-Hsin
    JOURNAL OF CLINICAL MEDICINE, 2019, 8 (02)
  • [34] Incidence of perioperative stroke in clampless aortic anastomosis during off-pump coronary artery bypass grafting
    Formica, Francesco
    Tata, Giuseppe
    Singh, Gurmeet
    Mariani, Serena
    D'Alessandro, Stefano
    Messina, Luigi Amerigo
    Sangalli, Fabio
    Paolini, Giovanni
    HEART AND VESSELS, 2018, 33 (06) : 595 - 604
  • [35] Redo Coronary Artery Bypass Grafting: On-Pump and Off-Pump Coronary Artery Bypass Grafting Revascularization Techniques
    Song Wu
    Feng Wan
    Zhe Zhang
    Hong Zhao
    Zhong-qi Cui
    Ji-yan Xie
    Chinese Medical Sciences Journal, 2015, 30 (01) : 28 - 33
  • [36] Long-Term Outcomes of On-Versus Off-Pump Coronary Artery Bypass Grafting
    Smart, Neil A.
    Dieberg, Gudrun
    King, Nicola
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 71 (09) : 983 - 991
  • [37] Total Arterial Off-Pump Coronary Artery Bypass Grafting: A 10-Year Experience
    Yuan, Biao
    Qi, Hong-Wei
    Zhao, Jian-Gang
    Wang, Xi-Zheng
    Chen, Nan
    Zhao, Feng-Hua
    Wang, Li-Fang
    Dong, Huan-Ping
    CHINESE MEDICAL JOURNAL, 2018, 131 (18) : 2179 - 2184
  • [38] Complete revascularization is compromised in off-pump coronary artery bypass grafting
    Robertson, Mark W.
    Buth, Karen J.
    Stewart, Keir M.
    Wood, Jeremy R.
    Sullivan, John A.
    Hirsch, Gregory M.
    Friesen, Camille L. Hancock
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 145 (04) : 992 - 998
  • [39] Passive graft perfusion in off-pump coronary artery bypass grafting
    Lu Feng
    Ji Bing-yang
    Liu Jin-ping
    Liu Ming-zheng
    Wang Gu-yan
    Hu Sheng-shou
    CHINESE MEDICAL JOURNAL, 2007, 120 (03) : 192 - 196
  • [40] Should obese patients undergo on- or off-pump coronary artery bypass grafting?
    Pullan, Mark
    Kirmani, Bilal H.
    Conley, Thomas
    Oo, Aung
    Shaw, Matthew
    McShane, James
    Poullis, Michael
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2015, 47 (02) : 309 - 315