Clinical outcome of the PAS-Port® proximal anastomosis system in off-pump coronary artery bypass grafting in 201 patients

被引:0
作者
Verberkmoes, N. J. [1 ]
Mokhles, M. M. [2 ]
Bramer, S. [1 ]
Van Straten, A. H. M. [1 ]
Ter Woorst, J. F. [1 ]
Maessen, J. G. [3 ]
Berreklouw, E. [1 ]
机构
[1] Catharina Hosp, Dept Cardiothorac Surg, Eindhoven, Netherlands
[2] Erasmus MC, Dept Cardiothorac Surg, Rotterdam, Netherlands
[3] MUMC, Cardiovasc Res Inst Maastricht CARIM, Dept Cardiothorac Surg, Maastricht, Netherlands
关键词
Outcome Assessment (Health Care); Surgical instruments; Coronary artery bypass; off-pump; RANDOMIZED-TRIAL; FOLLOW-UP; SURGERY; DEVICE; PATENCY; DISEASE;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim. The PAS-Port (R) Proximal Anastomosis System (Cardica, Inc, Redwood City, CA, USA) has been used worldwide since March 2003. The objective of the present study was to evaluate the clinical outcome of the PAS-Port (R) Proximal Anastomosis System. Methods. All the patients who underwent off-pump coronary artery bypass grafting in the Catharina Hospital Eindhoven between August 2006 and April 2010 were included in a non-randomized retrospective case-control study, if they had at least one proximal vein graft anastomosis. Study end-points consisted of overall survival, coronary reintervention and postoperative stroke. Results. The study included 312 patients (201 cases, 111 controls). After 36 months of follow-up there was no difference in survival between cases and controls (92.2% vs. 93.7%, P=0.52). No significant difference could be detected between cases and controls with respect to overall coronary reintervention-free survival (93% vs. 96.4%, P=0.20) and freedom from coronary reintervention due to proximal vein graft failure (98% vs. 100% P=0.14). The use of the PAS-Port system could not be identified as an independent risk factor of coronary reintervention (p=0.21). Postoperative stroke rates of cases and controls (2% vs. 0.9%, P=0.42) were comparable. Conclusion. The clinical outcomes in patients treated with the PAS-Port (R) Proximal Anastomosis System were satisfactory compared with those treated with the conventional hand-sewing technique. The use of the PAS-Port system was not associated with higher adverse outcome in terms of overall survival, stroke, coronary reintervention-free survival and freedom from reintervention due to proximal vein graft failure.
引用
收藏
页码:389 / 395
页数:7
相关论文
共 50 条
  • [21] Stroke in Off-Pump Coronary Artery Bypass Grafting
    Yaku, Hitoshi
    Doi, Kiyoshi
    ANNALS OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 16 (04) : 225 - 227
  • [22] Off-Pump Coronary Artery Bypass Grafting for Poorly Controlled Diabetic Patients
    Matsuura, Kaoru
    Imamaki, Mizuho
    Ishida, Atsushi
    Shimura, Hitoshi
    Niitsuma, Yuriko
    ANNALS OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 15 (01) : 18 - 22
  • [23] The influence of on-pump and off-pump coronary artery bypass grafting on hemorheological parameters
    Papp, J.
    Toth, A.
    Sandor, B.
    Kiss, R.
    Rabai, M.
    Kenyeres, P.
    Juricskay, I.
    Kesmarky, G.
    Szabados, S.
    Toth, K.
    CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2011, 49 (1-4) : 331 - 346
  • [24] Ten-year survival after off-pump coronary artery bypass grafting compared with traditional bypass grafting
    Gonzalez, Roberto L.
    Saldivia, Diego Z.
    Reyes, Rodrigo M.
    Alarcon, Felipe O.
    Seguel, Enrique S.
    Stockins, Aleck L.
    Barra, Sebastian M.
    Schaub, Andres C.
    Riquelme, Alejandra U.
    Madrid, Patricio C.
    Perez, Alejandro G.
    Alarcon, Emilio C.
    REVISTA MEDICA DE CHILE, 2023, 151 (01) : 32 - 41
  • [25] Anaesthesia for off-pump coronary artery bypass grafting
    Dryden, Leon
    Maccario, Marco
    ANAESTHESIA AND INTENSIVE CARE MEDICINE, 2018, 19 (06) : 281 - 284
  • [26] Urgent off-pump coronary artery bypass grafting
    Hitoshi Hirose
    Atushi Amano
    Akihito Takahashi
    Shuichirou Takanashi
    The Japanese Journal of Thoracic and Cardiovascular Surgery, 2002, 50 (8): : 330 - 337
  • [27] Anaesthesia for off-pump coronary artery bypass grafting
    Kilpatrick, Thomas
    Dryden, Leon
    ANAESTHESIA AND INTENSIVE CARE MEDICINE, 2024, 25 (04) : 244 - 248
  • [28] Recent data on off-pump coronary artery bypass grafting: the CORONARY and GOPCABE trials
    Pepper, John
    EUROINTERVENTION, 2013, 9 (01) : 29 - 32
  • [29] Conversion From Off-Pump Coronary Artery Bypass Grafting to On-Pump Coronary Artery Bypass Grafting
    Keeling, Brent
    Thourani, Vinod
    Aliawadi, Gorav
    Kim, Sunghee
    Cyr, Derek
    Badhwar, Vinay
    Jacobs, Jeff
    Brennan, J. Matthew
    Meza, James
    Matsouaka, Roland
    Halkos, Michael E.
    ANNALS OF THORACIC SURGERY, 2017, 104 (04) : 1267 - 1273
  • [30] The future of off-pump coronary artery bypass grafting: a North American perspective
    Yanagawa, Bobby
    Nedadur, Rashmi
    Puskas, John D.
    JOURNAL OF THORACIC DISEASE, 2016, 8 : S832 - S838