Performance evaluation of microPET: A high-resolution lutetium oxyorthosilicate PET scanner for animal imaging

被引:0
作者
Chatziioannou, AF
Cherry, SR
Shao, YP
Silverman, RW
Meadors, K
Farquhar, TH
Pedarsani, M
Phelps, ME
机构
[1] Univ Calif Los Angeles, Sch Med, Crump Inst Biol Imaging, Dept Mol & Med Pharmacol, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Sch Med, US DOE, Lab Struct Biol & Mol Med, Los Angeles, CA 90095 USA
关键词
PET; instrumentation; performance evaluation; animal imaging;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new dedicated PET scanner, microPET, was designed and developed at the University of California, Los Angeles, for imaging small laboratory animals. The goal was to provide a compact system with superior spatial resolution at a fraction of the cost of a clinical PFT scanner. Methods: The system uses fiberoptic readout of individually cut lutetium oxyorthosilicate (LSO) crystals to achieve high spatial resolution. Each microPET detector consists of an 8 X 8 array of 2 x 2 x 10-mm LSO scintillation crystals that are coupled to a 64-channel photomultiplier tube by optical fibers. The tomograph consists of 30 detectors in a continuous ring with a 17.2-cm diameter and fields of view (FOVs) of 11.25 cm in the transaxial direction and 1.8 cm in the axial direction. The system has eight crystal rings and no interplane septa. It operates exclusively in the three-dimensional mode and has an electronically controlled bed that is capable of wobbling with a radius of 300 mu m. We describe the performance of the tomograph in terms of its spatial, energy and timing resolution, as well as its sensitivity and counting-rate performance. We also illustrate its overall imaging performance with phantom and animal studies that demonstrate the potential applications of this device to biomedical research. Results: Images reconstructed with three-dimensional filtered backprojection show a spatial resolution of 1.8 mm at the center of the FOV (CFOV), which remains <2.5 mm for the central 5 cm of the transaxial FOV. The resulting volumetric resolution of the system is <8 mu L. The absolute system sensitivity measured with a 0.74 MBq (20 mu Ci) Ge-68 point source at the CFOV is 5.62 Hz/kBq. The maximum noise equivalent counting rate obtained with a 6.4-cm diameter cylinder spanning the central 56% of the FOV is 10 kcps, whereas the scatter fraction is 37% at the CFOV for an energy window of 250-650 keV and the same diameter cylinder. Conclusion: This is the first PET scanner to use the new scintillator LSO and uses a novel detector design to achieve high volumetric spatial resolution. The combination of imaging characteristics of this prototype system (resolution, sensitivity, counting rate performance and scatter fraction) opens up new possibilities in the study of animal models with PET.
引用
收藏
页码:1164 / 1175
页数:12
相关论文
共 40 条
[1]   A METHOD FOR MEASURING THE ABSOLUTE SENSITIVITY OF POSITRON EMISSION TOMOGRAPHIC SCANNERS [J].
BAILEY, DL ;
JONES, T ;
SPINKS, TJ .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1991, 18 (06) :374-379
[2]   DETERMINATION OF OBJECT CONTOUR FROM PROJECTIONS FOR ATTENUATION CORRECTION IN CRANIAL POSITRON EMISSION TOMOGRAPHY [J].
BERGSTROM, M ;
LITTON, J ;
ERIKSSON, L ;
BOHM, C ;
BLOMQVIST, G .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1982, 6 (02) :365-372
[3]   THE DESIGN AND PHYSICAL CHARACTERISTICS OF A SMALL ANIMAL POSITRON EMISSION TOMOGRAPH [J].
BLOOMFIELD, PM ;
RAJESWARAN, S ;
SPINKS, TJ ;
HUME, SP ;
MYERS, R ;
ASHWORTH, S ;
CLIFFORD, KM ;
JONES, WF ;
BYARS, LG ;
YOUNG, J ;
ANDREACO, M ;
WILLIAMS, CW ;
LAMMERTSMA, AA ;
JONES, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (06) :1105-1126
[4]  
Brix G, 1997, J NUCL MED, V38, P1614
[5]   Performance study of a 3D small animal PET scanner based on BaF2 crystals and a photo sensitive wire chamber [J].
Bruyndonckx, P ;
Liu, XA ;
Tavernier, S ;
Zhang, SP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 392 (1-3) :407-413
[6]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[7]   Investigation of LSO crystals for high spatial resolution positron emission tomography [J].
Casey, ME ;
Eriksson, L ;
Schmand, M ;
Andreaco, MS ;
Paulus, M ;
Dahlbom, M ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1109-1113
[8]   STUDY ON THE USE OF TRANSMISSION SCANS FOR WHOLE-BODY PET ATTENUATION CORRECTION [J].
CHATZIIOANNOU, A ;
DAHLBOM, M ;
HOH, CJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1545-1550
[9]   EFFECTS OF SCATTER ON MODEL PARAMETER ESTIMATES IN 3D PET STUDIES OF THE HUMAN BRAIN [J].
CHERRY, SR ;
HUANG, SC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1174-1179
[10]   Optical fiber readout of scintillator arrays using a multi-channel PMT: A high resolution PET detector for animal imaging [J].
Cherry, SR ;
Shao, YP ;
Siegel, S ;
Silverman, RW ;
Mumcuoglu, E ;
Meadors, K ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1932-1937