Vibration isolation pad for medical purposes: modeling and simulation

被引:1
作者
Abu Hanieh, Ahmed [1 ]
机构
[1] Birzeit Univ, Birzeit, Israel
关键词
Finite element model; integral force feedback; magnetic resonance imaging; MRI; piezoelectric actuator; vibration isolation; COIL;
D O I
10.1177/1077546311429340
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This project aims at establishing new research about protecting sensitive human beings from vibration syndrome. The main application targeted here is the protection of patients undergoing the Magnetic Resonance Imaging test where noise and vibration can cause severe damage to the test results. Stability and accuracy are very important issues in this medical test. An active vibration isolation pad will be designed and manufactured in this research where the force of the vibrating human body is measured and compensated to be fed back to a linear piezoelectric actuator to isolate the disturbing source of vibrations from propagating into the patient's body. This paper presents analytical and numerical models for the vibration isolation system where the Integral Force Feedback control technique is applied. The results prove high authority and robustness for this control technique especially on the first vibration mode.
引用
收藏
页码:2031 / 2038
页数:8
相关论文
共 15 条
  • [1] Abu Hanieh A., 2008, INT REV MECH ENG IRE, V2, P290
  • [2] Abu Hanieh A, 2009, ACTIVE ISOLATION CON
  • [3] Claeyssen F, 1998, P ACT 98 C, P170
  • [4] Evans J, 2003, P 10 INT C SOUND VIB, P112
  • [5] Evans J, 2004, JOURNAL ACOUSTICS RP
  • [6] Evans JB, 2000, J PERINATOL, V20, P105
  • [7] Hao Ko Ying, 2008, THESIS U SAINS MALAY
  • [8] Development of the technical capabilities needed to build and position a prepolarization coil for a magnetic resonance imaging magnet
    Jaensch, M.
    Young, I. R.
    Besant, C.
    Atkins, A.
    Lamperth, M.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H2) : 185 - 194
  • [9] Vibration and acoustic noise characterization of a gradient coil insert in a 4 T MRI
    Mechefske, CK
    Yao, G
    Wang, FL
    [J]. JOURNAL OF VIBRATION AND CONTROL, 2004, 10 (06) : 861 - 880
  • [10] AN ACTIVE VIBRATION ISOLATION SYSTEM FOR INERTIAL REFERENCE AND PRECISION-MEASUREMENT
    NELSON, PG