High frequency single element and annular array transducers incorporating PVDF

被引:6
作者
Alves, CHF
Snook, K
Ritter, T
Shung, KK
机构
来源
MEDICAL IMAGING 2000: ULTRASONIC IMAGING AND SIGNAL PROCESSING | 2000年 / 3982卷
关键词
high frequency ultrasonic transducers; annular arrays; electrical matching; ultrasound backscatter microscopy;
D O I
10.1117/12.382219
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Polyvinylidene fluoride (PVDF) transducers have been developed for ultrasonic backscatter microscopy (UBM) applications. Single element devices encompassed the frequency range of 40 to 65 MHz and incorporated a variety of backing materials and tuning circuits. Pulse-echo testing was performed using the 1 mu J energy setting on a Panametrics 5900 pulser/receiver, configured with +40dB gain and 20dB attenuation. A broad spectrum of performance was observed depending on the choice of backing and tuning, with -6dB bandwidths of 55% to 116% and echo amplitudes from 100mVpp to 940mVpp. It was observed that electrical impedance matching increased sensitivity and tuned the center frequency at the expense of reduced bandwidth. For imaging applications the tuned devices provided superior performance. Although single element transducers in this frequency range provide axial resolution on the order of 50 Gun, a tradeoff exists between the lateral resolution and the depth of held. To obtain high lateral resolution and a long depth of field an annular array design was adopted. Laser dicing was used to fabricate the elements of the array fi om a sheet of PVDF. Interconnect to each element was achieved using pin contacts molded into the backing. The design incorparated a 5mm aperture with six equal area elements focused at 10mm. At a center frequency of 50 MHz this design will achieve a lateral resolution of 58 mu m, comparable to the axial resolution, over the entire 7.5mm depth of field.
引用
收藏
页码:116 / 121
页数:6
相关论文
共 12 条
[1]   AN ANNULAR ARRAY SYSTEM FOR HIGH-RESOLUTION BREAST ECHOGRAPHY [J].
ARDITI, M ;
TAYLOR, WB ;
FOSTER, FS ;
HUNT, JW .
ULTRASONIC IMAGING, 1982, 4 (01) :1-31
[2]   TRANSIENT FIELDS OF CONCAVE ANNULAR ARRAYS [J].
ARDITI, M ;
FOSTER, FS ;
HUNT, JW .
ULTRASONIC IMAGING, 1981, 3 (01) :37-61
[3]  
BROWN LF, 1993, P 1992 IEEE ULTR S, V1, P539
[4]   A parametric study of the concentric-ring transducer design for MRI guided ultrasound surgery [J].
Fjield, T ;
Fan, XB ;
Hynynen, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1996, 100 (02) :1220-1230
[5]   PRINCIPLES AND APPLICATIONS OF ULTRASOUND BACKSCATTER MICROSCOPY [J].
FOSTER, FS ;
PAVLIN, CJ ;
LOCKWOOD, GR ;
RYAN, LK ;
HARASIEWICZ, KA ;
BERUBE, L ;
RAUTH, AM .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1993, 40 (05) :608-617
[6]   AN INTEGRATED MULTIELEMENT ARRAY TRANSDUCER FOR ULTRASOUND IMAGING [J].
HATFIELD, JV ;
SCALES, NR ;
ARMITAGE, AD ;
HICKS, PJ ;
CHEN, QX ;
PAYNE, PA .
SENSORS AND ACTUATORS A-PHYSICAL, 1994, 41 (1-3) :167-173
[7]  
*KTECH CORP, 1998, PIEZ SOL PVDF POL FL
[8]  
PAYNE PA, 1994, ULTRASON, P1523, DOI 10.1109/ULTSYM.1994.401880
[9]   THE DESIGN AND FABRICATION OF HIGH-FREQUENCY POLY(VINYLIDENE FLUORIDE) TRANSDUCERS [J].
SHERAR, MD ;
FOSTER, FS .
ULTRASONIC IMAGING, 1989, 11 (02) :75-94
[10]   A 100 MHZ B-SCAN ULTRASOUND BACKSCATTER MICROSCOPE [J].
SHERAR, MD ;
STARKOSKI, BG ;
TAYLOR, WB ;
FOSTER, FS .
ULTRASONIC IMAGING, 1989, 11 (02) :95-105