Performance characteristics of a tandem spray chamber arrangement in double focusing sector field ICP-MS

被引:27
|
作者
Olofsson, RS
Rodushkin, I [1 ]
Axelsson, MD
机构
[1] Lulea Univ Technol, SGAB Analyt, S-97187 Lulea, Sweden
[2] Lulea Univ Technol, Div Appl Geol, S-97187 Lulea, Sweden
关键词
D O I
10.1039/b002609n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A tandem cyclone-Scott type spray chamber configuration combined with infrared heating and a low-flow Micromist nebulizer was evaluated for the use in a ICP-SFMS sample introduction system. The system provides about a 3-fold sensitivity gain and lower oxide formation compared to the standard introduction system, although matrix-induced signal suppression is more pronounced, especially for elements with high ionization potential. The technique was evaluated by accurate analysis of riverine water reference material SLRS-3 with generally good agreement with certified and literature values. Long-term precision at a 20 mu g l(-1) concentration level was 0.7% RSD in low resolution and 1.6% RSD in medium resolution. Precision in the range 0.03-0.09% RSD for lead isotope ratios at 1 mu g l(-1) was obtained.