Monte Carlo calculations and experimental measurements of dosimetry parameters of a new 103Pd source

被引:34
作者
Li, ZF
Palta, JR
Fan, JJ
机构
[1] Univ Florida, Ctr Hlth Sci, Dept Radiat Oncol, Gainesville, FL 32610 USA
[2] N Amer Sci Inc, Calibrat Lab, Chatsworth, CA 91311 USA
关键词
prostate neoplasms; brachytherapy dosimetry; palladium; 103; Monte Carlo calculations;
D O I
10.1118/1.598975
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Permanent prostate implantation using I-125 (iodine) or Pd-103 (palladium) sources is a popular treatment option in the management of early prostate cancer. As sources of new designs are developed and marketed for application in permanent prostate implantations, their dosimetric characteristics must be carefully determined in order to maintain the accuracy of patient treatment. This report presents the results of experimental measurements and Monte Carlo calculations of the dosimetric parameters performed for a newly available Pd-103 Seed source. The measurements were performed in a large scanning water phantom using a diode detector. The positioning of the source and detector was achieved by a computer-controlled positioning mechanism in the scanning water phantom. The dose rate constant in water for the new Pd-103 source was determined from measurements with the diode detector calibrated with I-125 sources of known air-kerma strength. The radial dose function values for the source were measured using the diode detector. Monte Carlo photon transport calculations were then used to calculate the dosimetric parameters of dose rate constant, radial dose function, and anisotropy function using an accurate geometric model of the source. The measured dose rate constant of 0.693 cGy/U-hr compares well with the Monte Carlo calculated value of 0.677 cCy/U-hr. These results are further compared with data on existing Pd-103 sources. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)03605-1].
引用
收藏
页码:1108 / 1112
页数:5
相关论文
共 31 条
  • [1] DIODE DOSIMETRY OF MODEL-6711 AND MODEL-6712 I-125 SEEDS IN A WATER PHANTOM
    AHMAD, M
    FONTENLA, DP
    TSAO, STC
    CHUI, CS
    REIFF, JE
    ANDERSON, LL
    HUANG, DYC
    SCHELL, MC
    [J]. MEDICAL PHYSICS, 1992, 19 (02) : 391 - 399
  • [2] DOSE-RATE DETERMINATION FOR I-125 SEEDS
    CHIUTSAO, ST
    ANDERSON, LL
    OBRIEN, K
    SANNA, R
    [J]. MEDICAL PHYSICS, 1990, 17 (05) : 815 - 825
  • [3] THERMOLUMINESCENT DOSIMETRY FOR PD-103 SEEDS (MODEL 200) IN SOLID WATER PHANTOM
    CHIUTSAO, ST
    ANDERSON, LL
    [J]. MEDICAL PHYSICS, 1991, 18 (03) : 449 - 452
  • [4] Accuracy of Monte Carlo photon transport simulation in characterizing brachytherapy dosimeter energy-response artefacts
    Das, RK
    Li, Z
    Perera, H
    Williamson, JF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (06) : 995 - 1006
  • [5] DIODE DOSIMETRY OF PD-103 MODEL 200 SEED IN WATER PHANTOM
    FONTENLA, DP
    AHMAD, M
    CHIUTSAO, ST
    ANDERSON, LL
    [J]. MEDICAL PHYSICS, 1994, 21 (06) : 817 - 820
  • [6] DOSE DISTRIBUTION OF I-125 SOURCES IN DIFFERENT TISSUES
    HUANG, DYC
    SCHELL, MC
    WEAVER, KA
    LING, CC
    [J]. MEDICAL PHYSICS, 1990, 17 (05) : 826 - 832
  • [7] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12) : 2015 - 2036
  • [8] LI Z, IN PRESS MED PHYS
  • [9] MONTE-CARLO CALCULATION OF KERMA TO A POINT IN THE VICINITY OF MEDIA INTERFACES
    LI, ZF
    WILLIAMSON, JF
    PERERA, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (12) : 1825 - 1840
  • [10] VOLUME-BASED GEOMETRIC MODELING FOR RADIATION TRANSPORT CALCULATIONS
    LI, ZF
    WILLIAMSON, JF
    [J]. MEDICAL PHYSICS, 1992, 19 (03) : 667 - 677