Intraocular pressure measured by dynamic contour tonometer and ocular response analyzer in normal tension glaucoma

被引:18
作者
Morita, Tetsuya [1 ,2 ]
Shoji, Nobuyuki [1 ,3 ]
Kamiya, Kazutaka [1 ,2 ]
Hagishima, Mana [1 ]
Fujimura, Fusako [1 ,3 ]
Shimizu, Kimiya [1 ,2 ]
机构
[1] Kitasato Univ Hosp, Dept Ophthalmol, Kanagawa, Japan
[2] Kitasato Univ, Sch Med, Dept Ophthalmol, Kanagawa 2288555, Japan
[3] Kitasato Univ, Sch Allied Hlth Sci, Dept Orthopt & Visual Sci, Kanagawa 2288555, Japan
关键词
Dynamic contour tonometer; Ocular response analyzer; Normal tension glaucoma; Intraocular pressure; Corneal biochemical properties; OPEN-ANGLE GLAUCOMA; CENTRAL CORNEAL THICKNESS; APPLANATION TONOMETRY; BIOMECHANICAL PROPERTIES; JAPANESE;
D O I
10.1007/s00417-009-1169-4
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To investigate intraocular pressure (IOP) measurement values in normal tension glaucoma (NTG) eyes using two different types of tonometer that are supposed to be little affected by corneal biochemical properties. This study included 30 normal eyes of 16 healthy subjects and 30 eyes of 16 patients with NTG. IOP was measured with a Goldmann applanation tonometer (GAT), a Pascal dynamic contour tonometer (DCT), and a Reichert ocular response analyzer (ORA) three times each for normal and NTG eyes. The main measures were GAT-IOP, DCT-IOP, corneal-compensated IOP (IOPcc), Goldmann-correlated IOP (IOPg), and central corneal thickness (CCT). In normal eyes, GAT-IOP was 13.2 +/- 1.4 mmHg; DCT-IOP, 13.0 +/- 1.6 mmHg; IOPcc, 13.6 +/- 2.0 mmHg; and IOPg, 12.4 +/- 2.0 mmHg. Multivariate analysis revealed no significant differences between the four measurements (p = 0.08). CCT was 524.6 +/- 27.3 microns. In NTG eyes, GAT-IOP was 13.1 +/- 1.3 mmHg; DCT-IOP, 13.7 +/- 1.3 mmHg; IOPcc, 15.2 +/- 2.0 mmHg; and IOPg, 12.7 +/- 2.0 mmHg. Multivariate analysis showed significant differences between the four measurements (p < 0.01). Sheff,'s test showed that IOPcc was significantly higher than GAT-IOP, DCT-IOP, and IOPg (GAT-IOP vs IOPcc: p < 0.0001; DCT-IOP vs IOPcc: p = 0.01; IOPcc vs IOPg: p < 0.0001). CCT was 515.4 +/- 32.9 microns, with no significant difference between normal and NTG eyes (p = 0.15). We investigated the values of IOP in NTG eyes as measured by the DCT and ORA. IOPcc was significantly greater than GAT-IOP, DCT-IOP and IOPg in NTG eyes, suggesting the possibility that IOP values may be underestimated.
引用
收藏
页码:73 / 77
页数:5
相关论文
共 21 条
[1]   Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, Ocular Response Analyzer, and Goldmann tonometry in glaucomatous eyes [J].
Annette, Hager ;
Kristina, Loge ;
Bernd, Schroeder ;
Mark-Oliver, Fuellhas ;
Wolfgang, Wiegand .
JOURNAL OF GLAUCOMA, 2008, 17 (05) :361-365
[2]  
Copt RP, 1999, ARCH OPHTHALMOL-CHIC, V117, P14
[3]   Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness [J].
Doyle, A ;
Lachkar, Y .
JOURNAL OF GLAUCOMA, 2005, 14 (04) :288-292
[4]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[5]   The prevalence of primary open-angle glaucoma in Japanese - The Tajimi study [J].
Iwase, A ;
Suzuki, Y ;
Araie, M ;
Yamamoto, T ;
Abe, H ;
Shirato, S ;
Kuwayama, Y ;
Mishima, HK ;
Shimizu, H ;
Tomita, G ;
Inoue, Y ;
Kitazawa, Y .
OPHTHALMOLOGY, 2004, 111 (09) :1641-1648
[6]   Comparison of dynamic contour tonometry with Goldmann applanation tonometry [J].
Kaufmann, C ;
Bachmann, LM ;
Thiel, MA .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (09) :3118-3121
[7]   Clinical comparison of contour and applanation tonometry and their relationship to pachymetry [J].
Kniestedt, C ;
Lin, S ;
Choe, J ;
Bostrom, A ;
Nee, M ;
Stamper, RL .
ARCHIVES OF OPHTHALMOLOGY, 2005, 123 (11) :1532-1537
[8]   Repeatability of intraocular pressure and corneal biomechanical properties measurements by the ocular response analyser [J].
Kynlgopoulos, M. ;
Schlote, T. ;
Kotecha, A. ;
Tzamalis, A. ;
Pajic, B. ;
Haefliger, I. .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2008, 225 (05) :357-360
[9]  
Leske M Cristina, 2004, Curr Opin Ophthalmol, V15, P102
[10]   Determining in vivo biomechanical properties of the cornea with an ocular response analyzer [J].
Luce, DA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :156-162