Four-dimensional dose reconstruction through in vivo phase matching of cine images of electronic portal imaging device

被引:4
|
作者
Yoon, Jihyung [1 ]
Jung, Jae Won [1 ]
Kim, Jong Oh [2 ]
Yi, Byong Yong [3 ]
Yeo, Inhwan [4 ]
机构
[1] EAST Carolina Univ, Dept Phys, Greenville, NC 27858 USA
[2] Univ Pittsburgh, Inst Canc, Dept Radiat Oncol, Pittsburgh, PA 15232 USA
[3] Univ Maryland, Sch Med, Dept Radiat Oncol, Baltimore, MD 21201 USA
[4] Loma Linda Univ, Med Ctr, Dept Radiat Med, Loma Linda, CA 92354 USA
关键词
4D; inverse dose reconstruction; phase matching; EPID; Monte Carlo; VMAT QA; DOSIMETRY; MOTION; IMRT;
D O I
10.1118/1.4954317
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A method is proposed to reconstruct a four-dimensional (4D) dose distribution using phase matching of measured cine images to precalculated images of electronic portal imaging device (EPID). Methods: (1) A phantom, designed to simulate a tumor in lung (a polystyrene block with a 3 cm diameter embedded in cork), was placed on a sinusoidally moving platform with an amplitude of 1 cm and a period of 4 s. Ten-phase 4D computed tomography (CT) images of the phantom were acquired. A planning target volume (PTV) was created by adding a margin of 1 cm around the internal target volume of the tumor. (2) Three beams were designed, which included a static beam, a theoretical dynamic beam, and a planning-optimized dynamic beam (PODB). While the theoretical beam was made by manually programming a simplistic sliding leaf motion, the planning-optimized beam was obtained from treatment planning. From the three beams, three-dimensional (3D) doses on the phantom were calculated; 4D dose was calculated by means of the ten phase images (integrated over phases afterward); serving as "reference" images, phase-specific EPID dose images under the lung phantom were also calculated for each of the ten phases. (3) Cine EPID images were acquired while the beams were irradiated to the moving phantom. (4) Each cine image was phase-matched to a phase-specific CT image at which common irradiation occurred by intercomparing the cine image with the reference images. (5) Each cine image was used to reconstruct dose in the phase-matched CT image, and the reconstructed doses were summed over all phases. (6) The summation was compared with forwardly calculated 4D and 3D dose distributions. Accounting for realistic situations, intratreatment breathing irregularity was simulated by assuming an amplitude of 0.5 cm for the phantom during a portion of breathing trace in which the phase matching could not be performed. Intertreatment breathing irregularity between the time of treatment and the time of planning CT was considered by utilizing the same reduced amplitude when the phantom was irradiated. To examine the phase matching in a humanoid environment, the matching was also performed in a digital phantom (4D XCAT phantom). Results: For the static, the theoretical, and the planning-optimized dynamic beams, the 4D reconstructed doses showed agreement with the forwardly calculated 4D doses within the gamma pass rates of 92.7%, 100%, and 98.1%, respectively, at the isocenter plane given by 3%/3 mm criteria. Excellent agreement in dose volume histogram of PTV and lung-PTV was also found between the two 4D doses, while substantial differences were found between the 3D and the 4D doses. The significant breathing irregularities modeled in this study were found not to be noticeably affecting the reconstructed dose. The phase matching was performed equally well in a digital phantom. Conclusions: The method of retrospective phase determination of a moving object under irradiation provided successful 4D dose reconstruction. This method will provide accurate quality assurance and facilitate adaptive therapy when distinguishable objects such as well-defined tumors, diaphragm, and organs with markers (pancreas and liver) are covered by treatment beam apertures. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:4420 / 4430
页数:11
相关论文
共 4 条
  • [1] Analysis of dose distribution reproducibility based on a fluence map of in vivo transit dose using an electronic portal imaging device
    Tardi, Didin
    Fitriandini, Aninda
    Fauziah, Annisa Rahma
    Wibowo, Wahyu Edy
    Siswantining, Titin
    Pawiro, Supriyanto Ardjo
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (01):
  • [2] Evaluation of two-dimensional electronic portal imaging device using integrated images during volumetric modulated arc therapy for prostate cancer
    Inui, Shoki
    Ueda, Yoshihiro
    Ono, Shunsuke
    Ohira, Shingo
    Isono, Masaru
    Nitta, Yuya
    Ueda, Hikari
    Miyazaki, Masayoshi
    Koizumi, Masahiko
    Teshima, Teruki
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2021, 26 (02) : 281 - 290
  • [3] Deep learning-based 3D in vivo dose reconstruction with an electronic portal imaging device for magnetic resonance-linear accelerators: a proof of concept study
    Li, Yongbao
    Xiao, Fan
    Liu, Biaoshui
    Qi, Mengke
    Lu, Xingyu
    Cai, Jiajun
    Zhou, Linghong
    Song, Ting
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (23):
  • [4] In vivo dose estimations through transit signal measured with thimble chamber positioned along the central axis at electronic portal imaging device level in medical linear accelerator in carcinoma of the middle-third esophagus patients undergoing three-dimensional conformal radiotherapy
    Kumar, Putha Suman
    Banerjee, S.
    Kumar, E. S. Arun
    Srinivas, Challapalli
    Vadhiraja, B. M.
    Saxena, P. U.
    Ravichandran, Ramamoorthy
    Kasturi, Dinesh Pai
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2018, 14 (02) : 300 - 307