The learning curve for intraoperative neuromonitoring of the recurrent laryngeal nerve in thyroid surgery

被引:37
作者
Wojtczak, Beata [1 ]
Kaliszewski, Krzysztof [1 ]
Sutkowski, Krzysztof [1 ]
Glod, Mateusz [1 ]
Barczynski, Marcin [2 ]
机构
[1] Wroclaw Med Univ, Dept & Clin Gen Gastroenterol & Endocrine Surg, Wroclaw, Poland
[2] Jagiellonian Univ, Coll Med, Chair Gen Surg 3, Dept Endocrine Surg, Krakow, Poland
关键词
Learning curve; Thyroid surgery; IONM; Recurrent laryngeal nerve; IDENTIFICATION; OPERATIONS; GOITER; PALSY;
D O I
10.1007/s00423-016-1438-8
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose Intraoperative neuromonitoring (IONM) of the recurrent laryngeal nerve (RLN) is often used in thyroid surgery. However, this procedure is complex and requires a learning period to master the technique. The aim of the study was to evaluate the learning curve for IONM. Methods A 3-year period (2012-2014) of working with IONM (NIM3.0, Medtronic) was prospectively analyzed with a special emphasis on comparing the initial implementation phase in 2012 (101 patients, 190 RLNs at risk) with subsequent years of IONM use in 2013 (70 patients, 124 RLNs at risk) and 2014 (65 patients, 120 RLNs at risk). Results The rate of successful IONM-assisted RLN identification increased gradually over the 3-year study period (92.11 % in 2012 vs. 95.16 % in 2013 vs. 99.16 % in 2014; p = 0.022), with a corresponding decrease in the rate of technical problems (12.87, 4.3, and 4.6 %, respectively; p = 0.039). The rate of RLN injuries tended to decrease over time: 3.68, 1.55, and 0.83 %, respectively (p = 0.220). Between 2012 and 2014, increases in the sensitivity (71.4 vs. 100 %), specificity (98 vs. 99 %), positive predictive value (62.5 vs. 75 %), negative predictive value (98 vs. 100 %), and overall accuracy of IONM (97.4 vs. 99.6 %) were observed (p = 0.049). Increasing experience with IONM resulted in more frequent utilization of total thyroidectomy (92 % in 2012 vs. 100 % in 2013-2014; p= 0.004). Conclusions There was a sharp decrease in the number of technical problems involving equipment setup from 2012 to 2014.
引用
收藏
页码:701 / 708
页数:8
相关论文
共 28 条
[11]   Intraoperative monitoring of the recurrent laryngeal nerve in thyroid surgery [J].
Dralle, H. ;
Sekulla, C. ;
Lorenz, K. ;
Brauckhoff, M. ;
Machens, A. .
WORLD JOURNAL OF SURGERY, 2008, 32 (07) :1358-1366
[12]  
Dralle H, 2014, CHIRURG, V85, P236, DOI 10.1007/s00104-013-2705-7
[13]   Influence of Intraoperative Neuromonitoring on Surgeons' Technique During Thyroidectomy [J].
Duclos, Antoine ;
Lifante, Jean-Christophe ;
Ducarroz, Simon ;
Soardo, Pietro ;
Colin, Cyrille ;
Peix, Jean-Louis .
WORLD JOURNAL OF SURGERY, 2011, 35 (04) :773-778
[14]   Intraoperative electrophysiologic monitoring of the recurrent laryngeal nerve [J].
Eisele, DW .
LARYNGOSCOPE, 1996, 106 (04) :443-449
[15]   Laryngeal recurrent nerve injury in surgery for benign thyroid diseases - Effect of nerve dissection and impact of individual surgeon in more than 27,000 nerves at risk [J].
Hermann, M ;
Alk, G ;
Roka, R ;
Glaser, K ;
Freissmuth, M .
ANNALS OF SURGERY, 2002, 235 (02) :261-268
[16]  
JATZKO GR, 1994, SURGERY, V115, P139
[17]   Intraoperative neuromonitoring of the recurrent laryngeal nerve results and learning curve [J].
Jonas, J. ;
Baehr, R. .
ZENTRALBLATT FUR CHIRURGIE, 2006, 131 (06) :443-448
[18]  
Kielar M, 2011, S U BA ZDROWIA, V68, P70
[19]   Results of Intraoperative Neuromonitoring in Thyroid Surgery and Preoperative Vocal Cord Paralysis [J].
Lorenz, Kerstin ;
Abuazab, Mohammed ;
Sekulla, Carsten ;
Schneider, Rick ;
Phuong Nguyen Thanh ;
Dralle, Henning .
WORLD JOURNAL OF SURGERY, 2014, 38 (03) :582-591
[20]   IONM-guided goiter surgery leading to two-stage thyroidectomy-indication and results [J].
Melin, Magnus ;
Schwarz, Katharina ;
Lammers, Bernhard J. ;
Goretzki, Peter E. .
LANGENBECKS ARCHIVES OF SURGERY, 2013, 398 (03) :411-418