Effect of the Enhancement Threshold on the Computer-Aided Detection of Breast Cancer using MRI

被引:18
作者
Levman, Jacob E. D. [1 ]
Causer, Petrina [1 ]
Warner, Ellen [1 ]
Martel, Anne L. [1 ]
机构
[1] Univ Toronto, Sunnybrook Res Inst, Dept Med Biophys, Toronto, ON M4N 3M5, Canada
关键词
Breast imaging; computer-aided detection; enhancement threshold; magnetic resonance imaging; CONTRALATERAL BREAST; NEOADJUVANT CHEMOTHERAPY; DISCRIMINATING BENIGN; MALIGNANT LESIONS; CLASSIFICATION; IMAGES; MAMMOGRAPHY; SYSTEM; CAD;
D O I
10.1016/j.acra.2009.03.018
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. To evaluate the effect that variations in the enhancement threshold have on the diagnostic accuracy of two computer-aided detection (CAD) systems for magnetic resonance based breast cancer screening. Materials and Methods. Informed consent was obtained from all patients participating in cancer screening and this study was approved by the participating institution's review board. This retrospective study was nested in a prospective, single-institution, high-risk, breast screening study involving dynamic contrast-enhanced magnetic resonance imaging. Only those screening examinations (n = 223) for which a histopathological diagnosis was available were included. Two CAD methods were performed: the signal enhancement ratio (SER) and support vector machines (SVMs). Statistical analysis was performed by tracking changes in each CAD test's diagnostic accuracy (eg, receiver-operating characteristic [ROC] curve area, maximum possible sensitivity) with changes in the enhancement threshold. Results. The enhancement threshold plays a significant role in affecting a CAD test's potential sensitivity, ROC curve area, and number of assumed true and false-positive predictions per cancerous examination. A high threshold can also limit the CAD-based detection of the full size of a lesion. Conclusions. Enhancement thresholds can limit a CAD test's ability to diagnose a lesion's full size and as such should not be raised above 60%. The clinically used SER method exhibits a high rate of false positives at low enhancement thresholds and as such the threshold should not be set lower than 50%. The SVM method yielded better results in our study than the SER method at clinically realistic enhancement thresholds.
引用
收藏
页码:1064 / 1069
页数:6
相关论文
共 27 条
[1]  
ALMARTEL, 2007, PHYS MED BIOL, V52, P3803
[2]   MR imaging - guided breast localization system with medial or lateral access [J].
Causer, Petrina A. ;
Piron, Cameron A. ;
Jong, Roberta A. ;
Curpen, Belinda N. ;
Luginbuhl, Chris A. ;
Glazier, Joan E. ;
Warner, Ellen ;
Hill, Kimberley ;
Muldoon, Joanne ;
Taylor, Glen ;
Wong, John W. ;
Plewes, Donald B. .
RADIOLOGY, 2006, 240 (02) :369-379
[3]  
CHANG CC, LIBSVM MDASHA LIB FO
[4]   Computer-aided detection applied to breast MRI: Assessment of CAD-generated enhancement and tumor sizes in breast cancers before and after neoadjuvant chemotherapy [J].
DeMartini, WB ;
Lehman, CD ;
Peacock, S ;
Russell, MT .
ACADEMIC RADIOLOGY, 2005, 12 (07) :806-814
[5]   Receiver operating characteristic analysis: A primer [J].
Eng, J .
ACADEMIC RADIOLOGY, 2005, 12 (07) :909-916
[6]  
Fischer D R, 2004, Eur Radiol, V14, P394
[7]   Bilateral imaging using separate interleaved 3D volumes and dynamically switched multiple receive coil arrays [J].
Greenman, RL ;
Lenkinski, RE ;
Schnall, MD .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (01) :108-115
[8]   Dynamic contrast-enhanced magnetic resonance imaging as an imaging biomarker [J].
Hylton, Nola .
JOURNAL OF CLINICAL ONCOLOGY, 2006, 24 (20) :3293-3298
[9]  
JOE BN, P INT SOC MAG RES ME, V14, P1796
[10]   Healthy premenopausal breast parenchyma in dynamic contrast-enhanced MR imaging of the breast: Normal contrast medium enhancement and cyclical-phase dependency [J].
Kuhl, CK ;
Bieling, HB ;
Gieseke, J ;
Kreft, BP ;
Sommer, T ;
Lutterbey, G ;
Schild, HH .